22 May 2020 Linear Stokes measurement of thermal targets using compact LWIR spectropolarimeter
Author Affiliations +
Abstract
The trade-off between spectral resolution and instrument throughput is analyzed for a compact, uncooled, longwave infrared (LWIR) channeled spectropolarimeter (IRCSP). The IRCSP was developed as a part of the Sub-mm Wave and InfraRed Polarimeters (SWIRP) project out of NASA's Goddard Spaceflight Center. The IRCSP scientific objective targets measurements of AOLP and DOLP with 1-µm spectral resolution from 8.5 - 12.5µm in a single snapshot. The geometry of the field stop determines the field of view (FOV) of the IRCSP. This work relates the spectral resolution, instrument throughput, and polarimetric accuracy of a spectro-polarimeter to the FOV. The accuracy of linear Stokes retrievals for low temperature thermal targets are predicted for varying FOV and measurement noise conditions. This work presents a method to quantify the achievable accuracy in AOLP and DOLP as a function of field stop dimensions and signal-to-noise ratio (SNR). While smaller field stops are shown to improve accuracy as the spectral resolution is increased, low SNR is the dominant source of error for the IRCSP prototype. For the IRCSP, a SNR of at least 80 is required to produce DOLP measurements with < 5% error for targets with DOLP < 0.2.
Conference Presentation
© (2020) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Kira A. Hart, Meredith Kupinski, Dong Wu, and Russell A. Chipman "Linear Stokes measurement of thermal targets using compact LWIR spectropolarimeter", Proc. SPIE 11412, Polarization: Measurement, Analysis, and Remote Sensing XIV, 114120H (22 May 2020); https://doi.org/10.1117/12.2557887
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Polarimetry

Modulation

Long wavelength infrared

Polarization

Spectral resolution

Sensors

Calibration

Back to Top