Paper
15 June 2020 Numerical jitter estimation for swept source optical coherence tomography
Author Affiliations +
Abstract
Due to asynchronization between the acquisition trigger and K-clock trigger in a swept source optical coherence tomography (SS-OCT) system, trigger jitter causes the spectrum a temporal shift in the spectral domain and thus corrupts the measurement. We study ternary distribution of the jitter signal by measuring TiO2 phantom using a SS-OCT system, and it shows one-pixel spectral shift in the spectral domain.
© (2020) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Lida Zhu, Arata Miyazawa, Pradipta Mukherjee, Ibrahim Abd El-Sadek, Kensuke Oikawa, Daisuke Oida, and Yoshiaki Yasuno "Numerical jitter estimation for swept source optical coherence tomography", Proc. SPIE 11521, Biomedical Imaging and Sensing Conference 2020, 115210T (15 June 2020); https://doi.org/10.1117/12.2573281