Shifted apparent diffusion coefficient (sADC) is an alternative to magnetic resonance elastography for calculating liver stiffness virtually, without the use of special elasticity software/hardware. The aim of this work was to investigate the effects of changing repetition time (TR), echo time (TE) and number of excitations (NEX) on sADC quantification using an abdominal phantom. Diffusion weighted images with b-values of 0, 200, and 1500 s/mm2 were acquired at 1.5T using a fat-suppressed spin echo sequence with varying TR (1 to 16s), TE (75 to 90ms) and NEX (4 to 12). ADC maps were generated by the scanner and sADC maps were calculated from the b=200 and b=1500 diffusion weighted images. Regions of interest were selected in the phantom liver parenchyma and lesion. Percent difference was calculated for each variable as [(Maximum-Minimum)/Maximum]*100. The session was performed 5 times and the mean ADC/sADC for each variable across each session was used to calculate the between-session percent difference. At TR>5000ms the percent difference in ADC and sADC was generally less within a session than between sessions. At TR<5000ms, ADC increased initially with TR. For NEX and TE, percent difference between sessions was generally greater than percent difference within sessions. This phantom study showed little obvious effect of varying TR (when TR>5000ms), NEX or TE on ADC or sADC compared to the differences between sessions with the MR protocol used.
|