1 February 1994 Monochromatic x-ray imaging with streak and framing cameras
Author Affiliations +
Abstract
This article describes the first experiment of our groups to combine monochromatic x-ray imaging with a time-resolving detector i.e., a streak camera and a 120 ps gated framing camera. The aim of setting such a time-resolved diagnostic is to image the x-ray emission from colliding plasmas with high spatial resolution in a very narrow spectral window. Both camera types were tested and the adjustment procedure for the crystal was tested with film as a detector. The obtained spectral and spatial resolution of the x-ray microscope was measured.
© (1994) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Eckhart Foerster, Ingo Uschmann, M. Dirksmoeller, Claude J. Chenais-Popovics, P. Renaudin, Ovidiu N. Rancu, and Jean-Claude J. Gauthier "Monochromatic x-ray imaging with streak and framing cameras", Proc. SPIE 2015, Applications of Laser Plasma Radiation, (1 February 1994); https://doi.org/10.1117/12.168012