Paper
1 May 1998 Design and development of a sensor for the direct and continuous measurement of inhaled nitric oxide: factors affecting sensitivity
Bhairavi R. Parikh, Babs R. Soller
Author Affiliations +
Proceedings Volume 3253, Biomedical Sensing and Imaging Technologies; (1998) https://doi.org/10.1117/12.308049
Event: BiOS '98 International Biomedical Optics Symposium, 1998, San Jose, CA, United States
Abstract
Nitric oxide (NO), in concentrations between 0 and 20 ppm, is currently being used as an inhaled agent to treat patients with post surgical complications and respiratory disorders. Because excessive levels of NO can be detrimental to the patient, NO must be monitored accurately and continuously. Currently available instruments have problems that limit their usefulness for this application. This paper discusses the development of an inexpensive, direct and continuous sensor for the measurement of inhaled nitric oxide. The sensor incorporates a 0.05 inch, gas permeable, flow-through liquid cell into a probe, which can be incorporated into a ventilator circuit. Sensor operation is based on the complexation reaction of NO with cytochrome-c, a biologically derived heme. The complex is monitored spectrophotometrically by measuring the absorbance in the visible region of the spectrum at 563 nm. The sensor is specific to NO in the presence of oxygen. This paper will address experiments to optimize sensitivity of the sensor. Increasing the flow rate and pressure of NO into the sensing chamber increased the optical absorbance at a high concentration of NO. Increasing the concentration of cytochrome-c increased the sensitivity of the sensor. The sensor is currently sensitive to a minimum concentration of 5 ppm and linear in the range of 5 to 175 ppm.
© (1998) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Bhairavi R. Parikh and Babs R. Soller "Design and development of a sensor for the direct and continuous measurement of inhaled nitric oxide: factors affecting sensitivity", Proc. SPIE 3253, Biomedical Sensing and Imaging Technologies, (1 May 1998); https://doi.org/10.1117/12.308049
Lens.org Logo
CITATIONS
Cited by 1 scholarly publication.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Sensors

Absorbance

NOx

Molecules

Calibration

Oxygen

Chemical elements

Back to Top