19 July 1999 Improved accuracy tristimulus colorimeters calibrated with standard detectors
Author Affiliations +
Proceedings Volume 3749, 18th Congress of the International Commission for Optics; (1999) https://doi.org/10.1117/12.354857
Event: ICO XVIII 18th Congress of the International Commission for Optics, 1999, San Francisco, CA, United States
Abstract
A new method has been developed to calibrate tristimulus colorimeters for high accuracy color measurements. Instead of traditional lamp standards, modern, high accuracy detector standards are suggested for calibration. After high accuracy absolute spectral response determination of the tristimulus receivers, color (spectral) correction and peak (amplitude) normalization can minimize uncertainties caused by imperfect realizations of the Commission Internationale de l'Eclairage color matching functions. As a result of the corrections, stable light sources of different spectral power distributions can be measured with an accuracy dominated by the sub 0.1% uncertainty of novel spectral response determinations.
© (1999) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
George P. Eppeldauer "Improved accuracy tristimulus colorimeters calibrated with standard detectors", Proc. SPIE 3749, 18th Congress of the International Commission for Optics, (19 July 1999); https://doi.org/10.1117/12.354857