9 December 1999 Fast-response sensors for dissolved oxygen
Sheila Smith, Alan Hamill, Mahesh Uttamlal, Michael Campbell, David Bailly
Author Affiliations +
Abstract
In this work we have been developing fast response optical sensors for dissolved oxygen determination based on measuring the luminescence lifetime of a metalloporphyrin polymer. The sensor is produced by electropolymerization of the metalloporphyrin monomer units of platinum tetraphenylporphyrin (Pt-TPP), platinum octaethylporphyrin (Pt-OEP), palladium tetraphenylporphyrin (Pd-TPP) or palladium octaethylporphyrin (Pd-OEP). The polymerization process results in films which are in the region of micrometers thick. The Stern-Volmer quenching constants as determined from luminescence lifetime measurements for these sensors range from 0.90 (mg 1-1)-1) for Pt-TPP to 1.83 (mg 1-1)-1) for Pd-OEP. The response time of these sensors to a step change from an oxygen free to an oxygen saturated solution is in the millisecond region.
© (1999) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Sheila Smith, Alan Hamill, Mahesh Uttamlal, Michael Campbell, and David Bailly "Fast-response sensors for dissolved oxygen", Proc. SPIE 3860, Fiber Optic Sensor Technology and Applications, (9 December 1999); https://doi.org/10.1117/12.372962