15 October 2003 Performance assessment of two time-domain-scanning optical mammographs
Author Affiliations +
Abstract
Two time-domain scanning optical mammographs are presently tested in clinical trials within the EU project "OPTIMAMM" supported by the EC. To assess their performance (e.g., accuracy, sensitivity, stability), systematic measurements were performed on breast-like phantoms. Both, estimation of optical properties and acquisition of good contrast images for diagnostic purposes were considered. The proposed assessment procedure can be applied to characterize and improve other novel and existing instruments for photon migration imaging and spectroscopy.
© (2003) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Heidrun Wabnitz, Paola Taroni, Dirk Grosenick, Antonio Pifferi, Alessandro Torricelli, Adam Liebert, Michael Moeller, Rinaldo Cubeddu, and Herbert Rinneberg "Performance assessment of two time-domain-scanning optical mammographs", Proc. SPIE 5138, Photon Migration and Diffuse-Light Imaging, (15 October 2003); https://doi.org/10.1117/12.500533