20 January 2004 Fully differential readout electronics for coplanar-grid detectors
James E Toney, Christopher M Baer, Nader Kalkhoran
Author Affiliations +
Abstract
We have demonstrated a new amplifier topology for coplanar-grid detectors that provides true differential readout with a single, fully-differential, charge-sensitive preamplifier. A prototype, multi-detector system with adjustable gain for each detector has been demonstrated using this fully differential approach. In its initial implementation using general-purpose amplifier chips, this system produces comparable noise performance to the traditional two-amplifier readout with specialized, charge-sensitive preamplifiers. In lieu of a differential gain to correct for electron trapping, the circuit uses a differential sampling scheme. This method enables a symmetrical photopeak to be obtained, but introduces some undesired filtering that limits energy resolution.
© (2004) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
James E Toney, Christopher M Baer, and Nader Kalkhoran "Fully differential readout electronics for coplanar-grid detectors", Proc. SPIE 5198, Hard X-Ray and Gamma-Ray Detector Physics V, (20 January 2004); https://doi.org/10.1117/12.504882