Abstract
Radiochromic film has been applied to mapping of dose distributions in radiation therapy. However, this particular RCF product suffers from intrinsic nonuniform response. In this study, we have carefully measured the nonuniformity of unirradiated RCF, and the response of irradiated RCF in pixel-to-pixel as function of various absorbed doses, in order to create a method of predicting the error of irradiated, unirradiated RCF.
© (2006) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Hideki Aoyama, Yoshiharu Azuma, Sachiko Goto, and Tetsuhiro Sumimoto "High-accuracy measuring method of absorbed dose using radiochromic film and a microdensitometer for high-energy x rays", Proc. SPIE 6357, Sixth International Symposium on Instrumentation and Control Technology: Signal Analysis, Measurement Theory, Photo-Electronic Technology, and Artificial Intelligence, 63571L (24 October 2006); https://doi.org/10.1117/12.716968