13 March 2009 Dual-energy contrast enhanced digital mammography: theoretical and experimental study of optimal monoenergetic beam parameters using synchrotron radiation
Author Affiliations +
Proceedings Volume 7258, Medical Imaging 2009: Physics of Medical Imaging; 72583U (2009) https://doi.org/10.1117/12.811596
Event: SPIE Medical Imaging, 2009, Lake Buena Vista (Orlando Area), Florida, United States
Abstract
Dual-energy imaging with the injection of an iodinated contrast medium has the potential to depict cancers in the breast, by the demonstration of tumour angiogenesis and the suppression of the breast structure noise. The present study investigates the optimum monoenergetic beam parameters for this application. First, a theoretical study of the beam parameters was performed to find the best compromise between the quality of the dualenergy recombined image and the patient dose. The result of this analysis was then validated by phantom experiments using synchrotron monoenergetic radiation at the European Synchrotron Radiation Facility (ESRF, Grenoble, France). For an average breast of 5cm thickness and 50% glandularity, the theoretical simulations show an optimum at 20 keV for the low energy and 34 keV for the high energy, for a CsI detector of a standard mammography system. The SDNR variations with the low energy, the high energy or the dose repartition are very similar between the measurements on images acquired with synchrotron radiation and the simulated values. This ensures the accuracy of our theoretical optimization and the validity of the optimal beam parameters found in this study. The aim of this work is to demonstrate the potential of Dual-Energy CEDM (Contrast Enhanced Digital Mammography) with ideal monoenergetic sources, in order to provide an indicator of how to shape the polyenergetic spectra produced with classical X-ray sources for this application.
© (2009) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Sylvie Puong, Razvan Iordache, Xavier Bouchevreau, and Serge Muller "Dual-energy contrast enhanced digital mammography: theoretical and experimental study of optimal monoenergetic beam parameters using synchrotron radiation", Proc. SPIE 7258, Medical Imaging 2009: Physics of Medical Imaging, 72583U (13 March 2009); https://doi.org/10.1117/12.811596
Lens.org Logo
CITATIONS
Cited by 3 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Image quality

Iodine

Breast

Sensors

Tissues

Synchrotron radiation

Digital mammography

Back to Top