22 March 2010 A research prototype system for quantum-counting clinical CT
S. Kappler, F. Glasser, S. Janssen, E. Kraft, M. Reinwand
Author Affiliations +
Abstract
Recent publications emphasize the benefits of quantum-counting applied to the field of Computed Tomography (CT). We present a research prototype scanner with a CdTe-based quantum-counting detector and 20 cm field-of-view (FOV). As of today there is no direct converter material on the market able to operate reliably in the harsh high-flux regime of clinical CT scanners. Nevertheless, we investigate the CT imaging performance that could be expected with high-flux capable material. Therefore we chose pixel sizes of 0.05 mm2, a good compromise between high-flux counting ability and energy resolution. Every pixel is equipped with two energy threshold counters, enabling contrast-optimization and dual-energy scans. We present a first quantitative analysis of contrast measurements, in which we limit ourselves to a low-flux scenario. Using an Iodine-based contrast agent, we find 17% contrast enhancement at 120 kVp, compared to energy-integrating CT. In addition, the general dual-energy capability was confirmed in first measurements. We conclude our work by demonstrating good agreement of measurement results and detailed CT-system simulations.
© (2010) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
S. Kappler, F. Glasser, S. Janssen, E. Kraft, and M. Reinwand "A research prototype system for quantum-counting clinical CT", Proc. SPIE 7622, Medical Imaging 2010: Physics of Medical Imaging, 76221Z (22 March 2010); https://doi.org/10.1117/12.844238
Lens.org Logo
CITATIONS
Cited by 54 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Sensors

X-ray computed tomography

Scanners

X-rays

Prototyping

Signal attenuation

Computed tomography

Back to Top