Paper
15 October 2012 Measurement of groove density variation of varied-line-space grating for high-resolution soft x-ray monochromator
Y. Senba, H. Kishimoto, T. Miura, H. Ohashi, S. Goto, T. Ishikawa
Author Affiliations +
Abstract
A varied-line-space grating is used in modern soft X-ray monochromators with a high resolving power. The grating parameters require high accuracy because errors of the parameters lead to the degradation of the resolving power. The parameter tolerances required to maintain a high resolving power were estimated by analytical calculation. The groove density variations of three gratings were measured with a long trace profiler. The measured errors in the parameters were found to be sufficiently small.
© (2012) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Y. Senba, H. Kishimoto, T. Miura, H. Ohashi, S. Goto, and T. Ishikawa "Measurement of groove density variation of varied-line-space grating for high-resolution soft x-ray monochromator", Proc. SPIE 8501, Advances in Metrology for X-Ray and EUV Optics IV, 850104 (15 October 2012); https://doi.org/10.1117/12.936518