19 March 2014 Imaging informatics based on method of MR temperature measurement in high-intensity focused ultrasound
Author Affiliations +
Abstract
MR imaging has been used to perform imaging guided high-intensity focused ultrasound (HIFU) and meanwhile can also be used precisely to measure tissue temperature in theory. But in practice, the temperature environment and target are complex. Therefore, it is difficult to measure targeted temperature just by simply using the theory of numerical calculation based on MR image information. In this presentation, we presented new MR temperature measurement, based on imaging informatics, to measure the targeted tissue temperature in MR imaging guided HIFU therapeutic procedure. By heating up the water phantom experiments under HIFU, the new algorithm gives a satisfactory result compared with existing algorithm. Based on experimental data, we can see the accuracy increase 37.5% from 0.4048℃ up to 0.2530℃ when we choose new algorithms.
© (2014) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Xiangjiao Chen and Jianguo Zhang "Imaging informatics based on method of MR temperature measurement in high-intensity focused ultrasound", Proc. SPIE 9039, Medical Imaging 2014: PACS and Imaging Informatics: Next Generation and Innovations, 90390S (19 March 2014); https://doi.org/10.1117/12.2043410