Abstract
Automated Breast Ultrasound systems introduce a new ultrasound-imaging technology to address some of the drawbacks of the traditional handheld ultrasound. The detection performance of this new technology is studied. The lesion detectability performance is measured using Monte-Carlo simulation techniques for the observers of the synthesized images generated by the automated breast ultrasound systems. This performance is compared to detection performance of the observer when handheld ultrasound images are available. It is shown that resolution degradation in the synthesized images results in lower detection performance as compared to the handheld images scanning the same plane.
© (2014) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Sara Bahramian and Keith Wear "Lesion detectability in automated breast ultrasound", Proc. SPIE 9040, Medical Imaging 2014: Ultrasonic Imaging and Tomography, 904008 (20 March 2014); https://doi.org/10.1117/12.2043978