Sources of error associated with the measurement of four performance parameters are discussed in this paper. These are the measurement of half-value layer (HVL), light field edge location, x-ray and light field congruence, and x-ray field size. HVL measurement parameters discussed are field size, the detector energy-dependent response, and x-ray beam filtration. Light field edge location by both visual and photometric methods are compared and differences quantified for one x-ray beam collimator. The impact of the different methods of light field edge location on x-ray and light field congruence are described. X-ray field size measurements for x-ray systems having large focal spots are discussed. The effect of the large focal spot as it impacts on small x-ray field size measurements is described.
© (1976) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Tommie J. Morgan, Libby B. Smith, and Leland R. Kirkland "Some Sources of Errors in the Performance Testing Of Diagnostic X-Ray Systems", Proc. SPIE 0096, Application of Optical Instrumentation in Medicine V, (23 December 1976); https://doi.org/10.1117/12.965443
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
X-rays

Aluminum

Diagnostics

Sensors

Visualization

Diagnostic tests

Mirrors

RELATED CONTENT

Information Detection In Diagnostic Radiology
Proceedings of SPIE (June 15 1984)
A Comparison Of Some Models Of X-Ray Beam Hardening
Proceedings of SPIE (December 23 1976)
Determination Of High-Intensity X-Ray Spectra
Proceedings of SPIE (July 06 1979)
Digital Radiography
Proceedings of SPIE (August 18 1980)

Back to Top