22 July 2016 Multi-resolution waveguide image slicer for the PEPSI instrument
Author Affiliations +
Abstract
A waveguide image slicer with resolutions up to 270.000 (planned: 300.000) for the fiber fed PEPSI echelle spectrograph at the LBT and single waveguide thicknesses of down to 70 μm has been manufactured and tested. The waveguides were macroscopically prepared, stacked up to an order of seven and thinned back to square stack cross sections. A high filling ratio was achieved by realizing homogenous adhesive gaps of 3.6 μm, using index matching adhesives for TIR within the waveguides. The image slicer stacks are used in immersion mode and are miniaturized to enable implementation in a set of 2x8. The overall efficiency is between 92 % and 96 %.
© (2016) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Erik Beckert, Klaus G. Strassmeier, Manfred Woche, Gerd Harnisch, Marcel Hornaff, Michael Weber, and Stuart Barnes "Multi-resolution waveguide image slicer for the PEPSI instrument", Proc. SPIE 9912, Advances in Optical and Mechanical Technologies for Telescopes and Instrumentation II, 99125O (22 July 2016); https://doi.org/10.1117/12.2231096