Paper
2 October 2003 High-resolution three-dimensional imaging inside biological media using white-light interference microscopy
Author Affiliations +
Abstract
We present a white-light interference microscope designed to produce high-resolution three-dimensional images of biological media. This technique is an alternative to conventional optical coherence tomography (OCT). The experimental setup is based on a Linnik interferometer illuminated with a tungsten halogen lamp. En face tomographic images are obtained in real-time without scanning by computing the difference of two phase-opposed interferometric images recorded by a high-resolution CCD camera. The short coherence length of the source and the compensation of dispersion mismatch in the interferometer arms yield an optical sectioning ability with 0.8 μm resolution in water. Transverse resolution of 1.0 μm is achieved by using high numerical aperture microscope objectives. A shot-noise limited detection sensitivity of 86 dB can be reached with 2 s acquisition time. High-resolution images of the Xenopus Laevis tadpole are shown.
© (2003) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Arnaud Dubois, Gael Moneron, Kate Grieve, and Albert Claude Boccara "High-resolution three-dimensional imaging inside biological media using white-light interference microscopy", Proc. SPIE 5140, Optical Coherence Tomography and Coherence Techniques, (2 October 2003); https://doi.org/10.1117/12.497723
Lens.org Logo
CITATIONS
Cited by 2 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Optical coherence tomography

Microscopes

Image resolution

Tomography

Mirrors

Tissues

Interferometers

RELATED CONTENT


Back to Top