Abstract
In this paper, the development of a 35 MHz 64-channel Piezoelectric Composite based Micromachined Ultrasound Transducer (PCMUT) phased array for NDE ultrasound application is presented. A 35 MHz PMN-PT single crystal 1-3 composite based PC-MUT phased array was designed with extensive acoustic field and 1D modeling. The initial modeling results demonstrated that the focused detection resolution (10% of -3 dB beam width) could be as small as 30 μm in the azimuth direction. The maximum imaging depth for ceramic samples is around 20 mm. The PC-MUT array being developed will extend the state-of-art NDE phased array technology from approximately 20 MHz to 35 MHz, which will greatly enhance the imaging resolution for a broad range of NDE ultrasound applications.
© (2009) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Kevin Snook, Xiaoning Jiang, Changhong Hu, Xuecang Geng, Ruibin Liu, John Welter, Kirk Shung, and Wesley S. Hackenberger "A 35 MHz PCMUT phased array for NDE ultrasound", Proc. SPIE 7294, Nondestructive Characterization for Composite Materials, Aerospace Engineering, Civil Infrastructure, and Homeland Security 2009, 729403 (8 April 2009); https://doi.org/10.1117/12.815621
Lens.org Logo
CITATIONS
Cited by 5 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Phased arrays

Nondestructive evaluation

Acoustics

Ceramics

Composites

Transducers

Ultrasonography

Back to Top