23 March 2010 Angular dependence of mammographic dosimeters in digital breast tomosynthesis
Author Affiliations +
Abstract
Digital Breast Tomosynthesis (DBT) is an emerging imaging modality that combines tomography with conventional digital mammography. In developing DBT dosimetry, a direct application of mammographic dosimetry has appeal. However, DBT introduces rotation of the x-ray tube relative to the dosimeter, thus raising questions about the angular dependence of mammographic dosimeters. To measure this dependence, two ionization chambers, two solid-stated detectors, and one photodiode were rotated relative to an incident Mo/Mo x-ray beam. In this isocentric DBT simulation, the signal of each dosimeter was studied over an angular range of 180° for tube voltages of 26 to 34 kV. One ionization chamber was then modeled numerically to study the response to various monoenergetic beams. The results show that all dosimeters underestimate dose to varying degrees; solid-state detectors show the greatest angular dependence while ionization chambers show the least. Correction factors were computed from the data for isocentric DBT images using projection angles up to ±25°; these factors ranged from 1.0014 to 1.1380. The magnitude of the angular dependence generally decreased with increasing energy, as shown with both the measured and modeled data. As a result, the error arising in measuring DBT dose with a mammographic dosimeter varies significantly; it cannot always be disregarded. The use of correction factors may be possible but is largely impractical, as they are specific to the dosimeter, x-ray beam, and DBT geometry. Instead, an angle-independent dosimeter may be more suitable for DBT.
© (2010) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Lena R. Bradley, Ann-Katherine Carton, and Andrew D. A. Maidment "Angular dependence of mammographic dosimeters in digital breast tomosynthesis", Proc. SPIE 7622, Medical Imaging 2010: Physics of Medical Imaging, 76225L (23 March 2010); https://doi.org/10.1117/12.845369
Lens.org Logo
CITATIONS
Cited by 3 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Digital breast tomosynthesis

X-rays

Ionization

Sensors

Solid state electronics

Photodiodes

Signal attenuation

Back to Top