13 March 2013 A one step vs. a multi step geometric calibration of an optical coherence tomography
Jesús Díaz Díaz, Maik Rahlves, Omid Majdani, Eduard Reithmeier, Tobias Ortmaier
Author Affiliations +
Abstract
This contribution compares two approaches for the geometric calibration of an optical coherence tomography (OCT) which forms part of a medical navigation system. For this purpose, a one step and a multi step calibration is performed with a self-produced 3D reference structure and a high-accurate 6 degrees of freedom (DoF) parallel robot, respectively. These 3D landmark-based geometric calibrations are based on the identification of a parameterized grey-box OCT model. We show in experimental results that both methods reduce systematic errors by more than one order of magnitude.
© (2013) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Jesús Díaz Díaz, Maik Rahlves, Omid Majdani, Eduard Reithmeier, and Tobias Ortmaier "A one step vs. a multi step geometric calibration of an optical coherence tomography", Proc. SPIE 8573, Design and Quality for Biomedical Technologies VI, 85730D (13 March 2013); https://doi.org/10.1117/12.2005264