4 March 2016 Analysis of the variation in OCT measurements of a structural bottle neck for eye-brain transfer of visual information from 3D-volumes of the optic nerve head, PIMD-Average [0;2π]
Per G. Söderberg, Filip Malmberg, Camilla Sandberg-Melin
Author Affiliations +
Proceedings Volume 9693, Ophthalmic Technologies XXVI; 96930O (2016) https://doi.org/10.1117/12.2218208
Event: SPIE BiOS, 2016, San Francisco, California, United States
Abstract
The present study aimed to analyze the clinical usefulness of the thinnest cross section of the nerve fibers in the optic nerve head averaged over the circumference of the optic nerve head. 3D volumes of the optic nerve head of the same eye was captured at two different visits spaced in time by 1-4 weeks, in 13 subjects diagnosed with early to moderate glaucoma. At each visit 3 volumes containing the optic nerve head were captured independently with a Topcon OCT- 2000 system. In each volume, the average shortest distance between the inner surface of the retina and the central limit of the pigment epithelium around the optic nerve head circumference, PIMD-Average [0;2π], was determined semiautomatically. The measurements were analyzed with an analysis of variance for estimation of the variance components for subjects, visits, volumes and semi-automatic measurements of PIMD-Average [0;2π]. It was found that the variance for subjects was on the order of five times the variance for visits, and the variance for visits was on the order of 5 times higher than the variance for volumes. The variance for semi-automatic measurements of PIMD-Average [0;2π] was 3 orders of magnitude lower than the variance for volumes. A 95 % confidence interval for mean PIMD-Average [0;2π] was estimated to 1.00 ±0.13 mm (D.f. = 12). The variance estimates indicate that PIMD-Average [0;2π] is not suitable for comparison between a onetime estimate in a subject and a population reference interval. Cross-sectional independent group comparisons of PIMD-Average [0;2π] averaged over subjects will require inconveniently large sample sizes. However, cross-sectional independent group comparison of averages of within subject difference between baseline and follow-up can be made with reasonable sample sizes. Assuming a loss rate of 0.1 PIMD-Average [0;2π] per year and 4 visits per year it was found that approximately 18 months follow up is required before a significant change of PIMDAverage [0;2π] can be observed with a power of 0.8. This is shorter than what has been observed both for HRT measurements and automated perimetry measurements with a similar observation rate. It is concluded that PIMDAverage [0;2π] has the potential to detect deterioration of glaucoma quicker than currently available primary diagnostic instruments. To increase the efficiency of PIMD-Average [0;2π] further, the variation among visits within subject has to be reduced.
Conference Presentation
© (2016) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Per G. Söderberg, Filip Malmberg, and Camilla Sandberg-Melin "Analysis of the variation in OCT measurements of a structural bottle neck for eye-brain transfer of visual information from 3D-volumes of the optic nerve head, PIMD-Average [0;2π]", Proc. SPIE 9693, Ophthalmic Technologies XXVI, 96930O (4 March 2016); https://doi.org/10.1117/12.2218208
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Head

Optic nerve

Optical coherence tomography

Eye

Visualization

Retina

Information visualization

Back to Top