29 July 2016 Sensitivity estimates for the SPICA Mid-Infrared Instrument (SMI)
Author Affiliations +
Abstract
We present the latest results of the sensitivity estimate for spectrometers of the SPICA Mid-Infrared Instrument (SMI). SMI has three spectroscopic channels; low resolution spectrometer (LRS), medium resolution spectrometer (MRS) and high resolution spectrometer (HRS). Taking account of the results of optical design of each spectrometer and the latest information of the expected performance of detector arrays, the continuum sensitivity for a point source, the continuum sensitivity for an extended source, the line sensitivity for a point source, the line sensitivity for an extended source, and the saturation limit are calculated for LRS, MRS and HRS and are provided in this paper.
© (2016) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Itsuki Sakon, Hidehiro Kaneda, Shinki Oyabu, Daisuke Ishihara, Takehiko Wada, and Naofumi Fujishiro "Sensitivity estimates for the SPICA Mid-Infrared Instrument (SMI)", Proc. SPIE 9904, Space Telescopes and Instrumentation 2016: Optical, Infrared, and Millimeter Wave, 99043V (29 July 2016); https://doi.org/10.1117/12.2232402