9 March 2017 Ultra-high spatial resolution multi-energy CT using photon counting detector technology
S. Leng, R. Gutjahr, A. Ferrero, S. Kappler, A. Henning, A. Halaweish, W. Zhou, J. Montoya, C. McCollough
Author Affiliations +
Abstract
Two ultra-high-resolution (UHR) imaging modes, each with two energy thresholds, were implemented on a research, whole-body photon-counting-detector (PCD) CT scanner, referred to as sharp and UHR, respectively. The UHR mode has a pixel size of 0.25 mm at iso-center for both energy thresholds, with a collimation of 32 × 0.25 mm. The sharp mode has a 0.25 mm pixel for the low-energy threshold and 0.5 mm for the high-energy threshold, with a collimation of 48 × 0.25 mm. Kidney stones with mixed mineral composition and lung nodules with different shapes were scanned using both modes, and with the standard imaging mode, referred to as macro mode (0.5 mm pixel and 32 × 0.5 mm collimation). Evaluation and comparison of the three modes focused on the ability to accurately delineate anatomic structures using the high-spatial resolution capability and the ability to quantify stone composition using the multi-energy capability. The low-energy threshold images of the sharp and UHR modes showed better shape and texture information due to the achieved higher spatial resolution, although noise was also higher. No noticeable benefit was shown in multi-energy analysis using UHR compared to standard resolution (macro mode) when standard doses were used. This was due to excessive noise in the higher resolution images. However, UHR scans at higher dose showed improvement in multi-energy analysis over macro mode with regular dose. To fully take advantage of the higher spatial resolution in multi-energy analysis, either increased radiation dose, or application of noise reduction techniques, is needed.
© (2017) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
S. Leng, R. Gutjahr, A. Ferrero, S. Kappler, A. Henning, A. Halaweish, W. Zhou, J. Montoya, and C. McCollough "Ultra-high spatial resolution multi-energy CT using photon counting detector technology", Proc. SPIE 10132, Medical Imaging 2017: Physics of Medical Imaging, 101320Y (9 March 2017); https://doi.org/10.1117/12.2255589
Lens.org Logo
CITATIONS
Cited by 19 scholarly publications and 2 patents.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Spatial resolution

Sensors

Collimation

Collimators

Dual energy imaging

Computed tomography

Lung

Back to Top