11 May 2017 UV scale calibration transfer from an improved pyroelectric detector standard to field UV-A meters and 365 nm excitation sources
G. P. Eppeldauer, V. B. Podobedov, C. C. Cooksey
Author Affiliations +
Abstract
Calibration of the emitted radiation from UV sources peaking at 365 nm, is necessary to perform the ASTM required 1 mW/cm2 minimum irradiance in certain military material (ships, airplanes etc) tests. These UV “black lights” are applied for crack-recognition using fluorescent liquid penetrant inspection. At present, these nondestructive tests are performed using Hg-lamps. Lack of a proper standard and the different spectral responsivities of the available UV meters cause significant measurement errors even if the same UV-365 source is measured. A pyroelectric radiometer standard with spectrally flat (constant) response in the UV-VIS range has been developed to solve the problem. The response curve of this standard determined from spectral reflectance measurement, is converted into spectral irradiance responsivity with <0.5% (k=2) uncertainty as a result of using an absolute tie point from a Si-trap detector traceable to the primary standard cryogenic radiometer. The flat pyroelectric radiometer standard can be used to perform uniform integrated irradiance measurements from all kinds of UV sources (with different peaks and distributions) without using any source standard. Using this broadband calibration method, yearly spectral calibrations for the reference UV (LED) sources and irradiance meters is not needed. Field UV sources and meters can be calibrated against the pyroelectric radiometer standard for broadband (integrated) irradiance and integrated responsivity. Using the broadband measurement procedure, the UV measurements give uniform results with significantly decreased uncertainties.
Conference Presentation
© (2017) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
G. P. Eppeldauer, V. B. Podobedov, and C. C. Cooksey "UV scale calibration transfer from an improved pyroelectric detector standard to field UV-A meters and 365 nm excitation sources", Proc. SPIE 10181, Advanced Optics for Defense Applications: UV through LWIR II, 101810W (11 May 2017); https://doi.org/10.1117/12.2263093
Lens.org Logo
CITATIONS
Cited by 1 scholarly publication.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Ultraviolet radiation

Calibration

Pyroelectric detectors

Radiometry

Sensors

Standards development

Ultraviolet sources

RELATED CONTENT

Standardization of UV LED measurements
Proceedings of SPIE (September 08 2015)
Broadband radiometric LED measurements
Proceedings of SPIE (September 07 2016)
Absolute Intensity Measurements In The Vacuum Ultraviolet
Proceedings of SPIE (August 12 1986)
Traceability methods comparison of broadband UV radiometers
Proceedings of SPIE (November 30 2011)
Uniform calibration of night vision goggles and test sets
Proceedings of SPIE (October 08 2007)

Back to Top