7 March 2011 Verification of the QUBYX perfectlum calibration software using a PR-670 spectro radiometer and associated verification facility
Hans Roehrig, Syed F. Hashmi
Author Affiliations +
Abstract
At the University of Arizona a research project is underway which addresses consistent color and consistent gray-scale reproduction for digital color displays used in medical image interpretation, specifically for Pathology. Now the University of Arizona can enter the field of ICC Profiling and Color Management. Verification of PerfectLum Software was successful. FIT and LUM tests were performed to verify the conformance and the deviation was quantified. The maximum GSDF Error is about 5.968 %. With respect to the results, all three objectives were met and the PerfectLum calibrated display confirmed to the AAPM TG18 standards.
© (2011) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Hans Roehrig and Syed F. Hashmi "Verification of the QUBYX perfectlum calibration software using a PR-670 spectro radiometer and associated verification facility", Proc. SPIE 7966, Medical Imaging 2011: Image Perception, Observer Performance, and Technology Assessment, 79660W (7 March 2011); https://doi.org/10.1117/12.878108