Abstract
Star simulator is necessary equipment for calibration and performance testing of star sensor in laboratory. A novel method for adjustment of star simulator has been proposed in this paper. This method is based on ordinary Penta prism, guide rail, Charge Coupled Devices (CCD) camera and sub-pixel centroid algorithm. The CCD camera is used to image the star light replicated by the Penta prism instead of the autocollimator compared with the traditional adjustment of collimator. The analysis shows that the limiting parallelism error of the single star simulator is less than 0.49" after adjustment using this proposed method. It can satisfy the high accuracy adjustment of single star simulator.
© (2010) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
De-zhi Su, Hai-bo Liu, and Wen-liang Wang "Novel method for adjusting of single star simulator using Penta prism", Proc. SPIE 7656, 5th International Symposium on Advanced Optical Manufacturing and Testing Technologies: Optical Test and Measurement Technology and Equipment, 76561X (11 October 2010); https://doi.org/10.1117/12.865186