6 March 2008 Detection of low contrast test patterns on an LCD with different luminance and illuminance settings
Author Affiliations +
Abstract
The DICOM part 14 grayscale standard display function provides one way of harmonizing image appearance under different monitor luminance settings. This function is based on ideal observer conditions where the eye is always adapted to the target luminance and thereby also at peak contrast sensitivity. Clinical workstations are however often exposed to variations in ambient light due to a sub-optimal reading room light environment. Also, clinical images are inhomogeneous and low-contrast patterns must be detected even at luminance levels that differ from the eye adaptation level. All deviations from ideal luminance conditions cause the observer to detect patterns with reduced eye sensitivity but the magnitude of this reduction is unclear. The purpose of this paper was to quantify the effect different luminance settings have on the contrast threshold. A method to display well-defined sinusoidal low-contrast test patterns on an LCD has previously been developed and was used in this study. The observers were exposed to light from three different areas: 1) A small sinusoidal test pattern. 2) The remaining of the display surface. 3) Ambient light from outside the display area covering most of the observer's field of view. By adjusting the luminance from each of these three areas, two major effects could be quantified. The first effect was similar to Barten's f-factor where the target luminance differs from the observer's adaptation level while the second effect concerned the influence of areas outside the display surface. When a luminance range of 1-350 cd/m2 was used, the contrast needed to detect a dark object in a gray surrounding was almost doubled compared to a dark object in a dark surrounding. Ambient light from outside the display area has a moderate effect on the contrast threshold, except for the combination of high ambient light and dark objects where the contrast threshold increased considerably.
© (2008) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Patrik Sund, Magnus Båth, and Lars Gunnar Månsson "Detection of low contrast test patterns on an LCD with different luminance and illuminance settings", Proc. SPIE 6917, Medical Imaging 2008: Image Perception, Observer Performance, and Technology Assessment, 69170N (6 March 2008); https://doi.org/10.1117/12.770259
Lens.org Logo
CITATIONS
Cited by 3 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Target detection

LCDs

Eye

Light sources and illumination

Contrast sensitivity

Medical imaging

Visualization

Back to Top