Paper
28 April 2004 Relationship of 2D ultrasonic spectral parameters to the physical properties of soft tissue scatterers
Tian Liu, Frederic L Lizzi, Jeffrey A. Ketterling, Paul Lee, Andrew Kalisz, Ronald H Silverman, Gerald J Kutcher
Author Affiliations +
Abstract
We have conducted a general study that relates calibrated 2-D ultrasonic spectral parameters to the physical properties of sub-resolution tissue scatterers. Our 2-D spectra are computed form digital radio-frequency echo data obtained as the transducer linearly scans along the cross-range (scan direction) with increments smaller than the half beam width. Acquired data are Fourier transformed with respect to range (beam) and cross-range (scan) directions. To quantitatively measure and classify the physical properties of tissues, we have defined two spectral functions and four spectral parameters. The 2-D spectral functions are: radially integrated spectral power (RISP) and angularly integrated spectral power (AISP). The summary parameters are: peak value and 3-dB width of the RISP, slope and intercept of the AISP. These parameter are understood in terms of the beam properties, transducer parameters and the physical properties of the tissue microstructures including size, shape, orientation, concentration and acoustic impedance. Our theoretical model indicates that 1) the 3-dB width of the RISP is predominantly determined by the scatterer size along the beam direction; 2) the slope of the linear fit of the AISP is predominantly determined by the scatterer size along range direction; 3) the concentration and the relative acoustic impedance fluctuation of the scatterers change the overall spectrum magnitude. The predictions of the theoretical model have been verified using beef muscle fibers examined with 40 MHz center frequency.
© (2004) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Tian Liu, Frederic L Lizzi, Jeffrey A. Ketterling, Paul Lee, Andrew Kalisz, Ronald H Silverman, and Gerald J Kutcher "Relationship of 2D ultrasonic spectral parameters to the physical properties of soft tissue scatterers", Proc. SPIE 5373, Medical Imaging 2004: Ultrasonic Imaging and Signal Processing, (28 April 2004); https://doi.org/10.1117/12.535836
Lens.org Logo
CITATIONS
Cited by 2 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Transducers

Tissues

Spatial frequencies

Spectrum analysis

Ultrasonics

Ultrasonography

Acoustics

Back to Top