Full Content is available to subscribers

Subscribe/Learn More  >
Proceedings Article

The adaptive BioCD: interferometric immunoassay on a spinning disk

[+] Author Affiliations
Leilei Peng, Manoj M. Varma, Fred E. Regnier, David D. Nolte

Purdue Univ. (USA)

Proc. SPIE 5692, Advanced Biomedical and Clinical Diagnostic Systems III, 224 (April 11, 2005); doi:10.1117/12.589217
Text Size: A A A
From Conference Volume 5692

  • Advanced Biomedical and Clinical Diagnostic Systems III
  • Tuan Vo-Dinh; Warren S. Grundfest; David A. Benaron; Gerald E. Cohn
  • San Jose, CA | January 22, 2005

abstract

The speed of interferometric detection is at least 1000 times faster than the fluorometric detection used in the vast majority of clinical diagnostic systems. This opens the possibility to perform thousands of assays in the time it takes fluorescence to perform only one. Molecules immobilized on a spinning disk, like a CD, present the fastest and simplest means of interrogating thousands of micron-scale interferometer elements per second. However, the challenge of interferometry on a spinning disk is to maintain stable phase in the presence of mechanical vibration. In this paper, we demonstrate the first use of adaptive optics in an adaptive optical homodyne mixer to perform interferometry on a multi-analyte BioCD. The BioCD is a 4" diameter glass disk printed with a spoke pattern of protein. When the disk spins, the periodic protein pattern is transferred into a high-speed optical phase modulation by spinning the disk at 3000 rpm in the path of a probe laser. A nonlinear optical film mixes the signal beam with a stable reference beam in a two-wave mixing configuration that adaptively phase-locks the two beams to create stable phase in spite of mechanical vibration. Specific binding of antibody to printed antigen is detected as an increased homodyne signal. Multi-analyte detection on Anti Mouse and Anti Rabbit IgG is performed in which Mouse IgG and Rabbit IgG act as the non-specific reagent to each other. Detection is made on circular tracks. The technique has the potential of fast screening for large numbers of protein interactions.

© (2005) COPYRIGHT SPIE--The International Society for Optical Engineering. Downloading of the abstract is permitted for personal use only.
Citation

Leilei Peng ; Manoj M. Varma ; Fred E. Regnier and David D. Nolte
"The adaptive BioCD: interferometric immunoassay on a spinning disk", Proc. SPIE 5692, Advanced Biomedical and Clinical Diagnostic Systems III, 224 (April 11, 2005); doi:10.1117/12.589217; http://dx.doi.org/10.1117/12.589217


Access This Proceeding
Sign in or Create a personal account to Buy this proceeding ($15 for members, $18 for non-members).

Figures

Tables

NOTE:
Citing articles are presented as examples only. In non-demo SCM6 implementation, integration with CrossRef’s "Cited By" API will populate this tab (http://www.crossref.org/citedby.html).

Some tools below are only available to our subscribers or users with an online account.

Related Content

Customize your page view by dragging & repositioning the boxes below.

Related Book Chapters

Topic Collections

Advertisement
  • Don't have an account?
  • Subscribe to the SPIE Digital Library
  • Create a FREE account to sign up for Digital Library content alerts and gain access to institutional subscriptions remotely.
Access This Proceeding
Sign in or Create a personal account to Buy this proceeding ($15 for members, $18 for non-members).
Access This Proceeding
Sign in or Create a personal account to Buy this article ($15 for members, $18 for non-members).
Access This Chapter

Access to SPIE eBooks is limited to subscribing institutions and is not available as part of a personal subscription. Print or electronic versions of individual SPIE books may be purchased via SPIE.org.