An explanation for the extremely low, but variable, radiation dosages measured in a linear slit scanning radiography system

[+] Author Affiliations
J. H. Potgieter, Mattieu de Villiers, Martin Scheelke

Lodox Systems (Pty) Ltd. (South Africa)

Gerhard de Jager

Univ. of Cape Town (South Africa)

Proc. SPIE 5745, Medical Imaging 2005: Physics of Medical Imaging, 1138 (August 30, 2005); doi:10.1117/12.595216
Text Size: A A A
From Conference Volume 5745

  • Medical Imaging 2005: Physics of Medical Imaging
  • Michael J. Flynn
  • San Diego, CA | February 12, 2005

abstract

Clinical trials performed for the FDA"s Section 510k compliance submission of the Statscan digital, full-body, linear slit scanning diagnostic radiography system revealed that comparable diagnostic results with a commercial full-field screen film device were obtained with the Statscan using much lower radiation doses. For certain imaging procedures the doses for Statscan were as much as twenty to thirty times lower. However the results varied by a large amount and in particular the results for chest radiographs were anomalous in that the Statscan dose was less reduced. Whilst it is well known that slit scanning radiography has considerably lower radiation exposure than full-field devices due to its much lower scatter to primary ratio and also that digital radiography has the potential for lower radiation dosages, it was thought that that this alone did not fully account for the dose differences. This paper suggests that these dose differences, including the anomaly mentioned above, can be explained by considering the unique way that slit scanning is undertaken by Statscan i.e. by scanning the tube, detector, slit and collimators together along a linear path. The effect on measured skin entrance doses is explained and the dosage differences as affected by digital technology, higher DQE, slit scanning (low scatter to primary ratio) and linear slit scanning methods are quantified. Furthermore it is explained how the Statscan geometry leads an improved "skin sparing" effect.

© (2005) COPYRIGHT SPIE--The International Society for Optical Engineering. Downloading of the abstract is permitted for personal use only.
Citation

J. H. Potgieter ; Mattieu de Villiers ; Martin Scheelke and Gerhard de Jager
"An explanation for the extremely low, but variable, radiation dosages measured in a linear slit scanning radiography system", Proc. SPIE 5745, Medical Imaging 2005: Physics of Medical Imaging, 1138 (August 30, 2005); doi:10.1117/12.595216; http://dx.doi.org/10.1117/12.595216


Access This Proceeding
Sign in or Create a personal account to Buy this proceeding ($15 for members, $18 for non-members).

Figures

Tables

NOTE:
Citing articles are presented as examples only. In non-demo SCM6 implementation, integration with CrossRef’s "Cited By" API will populate this tab (http://www.crossref.org/citedby.html).

Some tools below are only available to our subscribers or users with an online account.

Related Content

Customize your page view by dragging & repositioning the boxes below.

Related Book Chapters

Topic Collections

Advertisement
  • Don't have an account?
  • Subscribe to the SPIE Digital Library
  • Create a FREE account to sign up for Digital Library content alerts and gain access to institutional subscriptions remotely.
Access This Proceeding
Sign in or Create a personal account to Buy this proceeding ($15 for members, $18 for non-members).
Access This Proceeding
Sign in or Create a personal account to Buy this article ($15 for members, $18 for non-members).
Access This Chapter

Access to SPIE eBooks is limited to subscribing institutions and is not available as part of a personal subscription. Print or electronic versions of individual SPIE books may be purchased via SPIE.org.