Paper
9 November 2005 High-performance field-portable imaging radiometric spectrometer technology for hyperspectral imaging applications
Martin Chamberland, Vincent Farley, Alexandre Vallières, André Villemaire, Louis Belhumeur, Jean Giroux, Jean-François Legault
Author Affiliations +
Abstract
Standoff detection, identification and quantification of chemicals in the gaseous state are fundamental needs in several fields of applications. Additional required sensor characteristics include high sensitivity, low false alarms and high-speed (ideally real-time) operation, all in a compact and robust package. The thermal infrared portion of the electromagnetic spectrum has been utilized to implement such chemical sensors, either with spectrometers (with none or moderate imaging capability) or with imagers (with moderate spectral capability). Only with the recent emergence of high-speed, large format infrared imaging arrays, has it been possible to design chemical sensors offering uncompromising performance in the spectral, spatial, as well as the temporal domain. Telops has developed a novel instrument that can not only provide an early warning for chemical agents and toxic chemicals, but also one that provides a "Chemical Map" of the field of view and is man portable. To provide to best field imaging spectroscopy instrument, Telops has developed the FIRST, Field-portable Imaging Radiometric Spectrometer Technology, instrument. This instrument is based on a modular design that includes: a high performance infrared FPA and data acquisition electronics, onboard data processing electronics, a high performance Fourier transform modulator, dual integrated radiometric calibration targets, a visible boresight camera. These modules, assembled together in an environmentally robust structure, used in combination with Telops' proven radiometric and spectral calibration algorithms make this instrument a world-class passive standoff detection system for chemical imaging.
© (2005) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Martin Chamberland, Vincent Farley, Alexandre Vallières, André Villemaire, Louis Belhumeur, Jean Giroux, and Jean-François Legault "High-performance field-portable imaging radiometric spectrometer technology for hyperspectral imaging applications", Proc. SPIE 5994, Chemical and Biological Sensors for Industrial and Environmental Security, 59940N (9 November 2005); https://doi.org/10.1117/12.632104
Lens.org Logo
CITATIONS
Cited by 22 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Sensors

Calibration

Infrared imaging

Electronics

Interferometers

Staring arrays

Black bodies