Paper
23 July 2003 Instrumental requirements for blood glucose sensing based on diffuse reflectance NIR spectroscopy
Hidenobu Arimoto, Tomohiro Ota, Masatoshi Tarumi, Yukio Yamada
Author Affiliations +
Abstract
The study on the magnitude of spectral change in blood glucose measurements with diffuse reflectance NIR spectroscopy is presented. Spectral change is estimated by a Monte Carlo simulation and measurements of absorbance spectra of aqueous glucose solution. Required sensitivity of spectrophotometers for monitoring the changein the blood glucose concentration as small as 10mg/dL has been obtained using the estimated change in the absorbance spectrum and mean pathlength of light in tissue.
© (2003) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Hidenobu Arimoto, Tomohiro Ota, Masatoshi Tarumi, and Yukio Yamada "Instrumental requirements for blood glucose sensing based on diffuse reflectance NIR spectroscopy", Proc. SPIE 4965, Optical Diagnostics and Sensing in Biomedicine III, (23 July 2003); https://doi.org/10.1117/12.478381
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Glucose

Absorbance

Monte Carlo methods

Absorption

Blood

Scattering

Diffuse reflectance spectroscopy

Back to Top