Paper
3 March 2017 Optimising probe holder design for sentinel lymph node imaging using clinical photoacoustic system with Monte Carlo simulation
Author Affiliations +
Abstract
Photoacoustic tomography is a hybrid imaging modality that combines optical and ultrasound imaging. It is rapidly gaining attention in the field of medical imaging. The challenge is to translate it into a clinical setup. In this work, we report the development of a handheld clinical photoacoustic imaging system. A clinical ultrasound imaging system is modified to integrate photoacoustic imaging with the ultrasound imaging. Hence, light delivery has been integrated with the ultrasound probe. The angle of light delivery is optimized in this work with respect to the depth of imaging. Optimization was performed based on Monte Carlo simulation for light transport in tissues. Based on the simulation results, the probe holders were fabricated using 3D printing. Similar results were obtained experimentally using phantoms. Phantoms were developed to mimic sentinel lymph node imaging scenario. Also, in vivo sentinel lymph node imaging was done using the same system with contrast agent methylene blue up to a depth of 1.5 cm. The results validate that one can use Monte Carlo simulation as a tool to optimize the probe holder design depending on the imaging needs. This eliminates a trial and error approach generally used for designing a probe holder.
© (2017) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Kathyayini Sivasubramanian, Vijitha Periyasamy, Kew Kok Wen, and Manojit Pramanik "Optimising probe holder design for sentinel lymph node imaging using clinical photoacoustic system with Monte Carlo simulation", Proc. SPIE 10064, Photons Plus Ultrasound: Imaging and Sensing 2017, 100645N (3 March 2017); https://doi.org/10.1117/12.2250478
Lens.org Logo
CITATIONS
Cited by 1 scholarly publication.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Monte Carlo methods

Imaging systems

Ultrasonography

Tissues

Lymphatic system

Optical spheres

Signal to noise ratio