Paper
20 September 1989 Linear HgCdTe Radiometer
G. Eppeldauer, L. Novak
Author Affiliations +
Abstract
The basic modes of electrical operation of photoconductive optical radiation detectors are analyzed. The nonlinearity inherent in "voltage mode" measurements can be eliminated by using "current mode" measurements. A HgCdTe photoconductive radiometer has been designed, based on this analysis, which measures the current through a biased detector. A built-in calibrating capability in the circuits of the radiometer makes it possible to eliminate the effects of a long-term drift in the bias voltage, thereby achieving a higher precision.
© (1989) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
G. Eppeldauer and L. Novak "Linear HgCdTe Radiometer", Proc. SPIE 1110, Imaging Infrared: Scene Simulation, Modeling, and Real Image Tracking, (20 September 1989); https://doi.org/10.1117/12.960756