1.INTRODUCTIONThere is a large performance gap between conventional, electron-impact X-ray sources and synchrotron radiation sources. Electron-impact sources are compact with low to moderate cost, making them accessible for a wide range of X-ray applications and laboratories. On the other hand, their flux and brightness are limited, limiting their measurement speed and/or data quality on many types of samples. Furthermore, their spectrum, consisting of fixed-energy emission lines (depending on the choice of target material) on top of a continuum of bremsstrahlung (with a cutoff energy given by the acceleration voltage), makes them less-well suited for applications requiring energy tunability and a high degree of monochromaticity. Such applications are well-served by synchrotron radiation facilities, which provide high-flux, high-brightness beams that are tunable and can be monochromatized to a high degree. However, synchrotrons are large and expensive to build and operate, which is why they are typically operated as government-funded facilities and can provide only a limited amount of beamtime. An Inverse Compton Scattering (ICS) X-ray source1,2 can bridge this gap by providing a narrow-band, high-flux and tunable X-ray source that fits into a laboratory at a cost of a few percent of a large synchrotron facility. It works by colliding a high-power laser beam with a relativistic electron beam, in which case the energy of the backscattered photons is in the X-ray (or gamma-ray) regime. So far, the only ICS source in regular user operation is the Munich Compact Light Source (MuCLS),3,4 a combination of Lyncean Technologies’ commercially available Compact Light Source (CLS)5 and a beamline with two endstations built by researchers at the Technical University of Munich.6 The application focus of the MuCLS is biomedical imaging of centimeter-sized samples in the 15-35 keV energy range, well-matched to the beam properties of the first generation Lyncean CLS with ~4 mrad divergence and spectral bandwidth of 3-5%. For a summary of application examples from the MuCLS incl. further references, please see Ref. 6. Following design principles developed for a narrow-band gamma-ray source for the ELI-NP project (www.eli-np.ro), we have recently developed a concept for a next generation CLS that will be more than two orders of magnitude brighter than the source in operation at the MuCLS. With two different configurations (laser wavelengths), it can cover an X-ray energy range of about 30-90 keV, or 60-180 keV (the lower end of the energy range might be reduced further depending on optimization of the design). This shift towards higher energies changes the application focus of the source towards studies of larger samples and denser materials, enabling use cases in geo- and materials science as well as biomedical imaging of larger samples. Furthermore, the energy range and higher flux of this source provides a pathway for transferring radiotherapy applications currently performed at synchrotrons into a clinical setting. The performance (flux and brightness) of synchrotron beamlines typically drops towards higher energies. By contrast, ICS sources perform better at higher energies because the opening angle of the photon beam scales inversely with the electron beam energy : θ0 = 1/y, where θ0 is the opening angle, containing 1/2 of the total photon flux, and γ is the electron beam energy in units of electron rest mass. Consequently, the next generation CLS is expected to reach the performance level of some synchrotron beamlines at higher energies. This is qualitatively illustrated in Fig. 1, and a more quantitative comparison to select beamlines is provided in Sec. 3. In the following sections, we first briefly review the principles of inverse Compton scattering. We then present the design and target performance parameters of the next generation CLS, followed by a quantitative comparison to some laboratory sources and synchrotron beamlines. Finally, we discuss some applications well-suited for this X-ray source. 2.INVERSE COMPTON SCATTERINGWe have previously described the principles of inverse Compton scattering,5 which we briefly summarize here. In the wave picture, inverse Compton scattering can be explained as analogous to undulator radiation in a synchrotron facility, where the electromagnetic field of a laser beam generates X-rays in the same way as the magnetic field of a permanent magnet undulator. In the particle picture, it can be explained by backscattering of a laser photon off a relativistic electron as illustrated in Fig. 2a. The energy of the backscattered photon is upshifted into the X-ray regime by energy transfer from the electron. The peak energy (for head-on collision with θi = 180° and θf = 0°, referring to the angles denoted in Fig. 2a), the energy Ex of the backscattered photon is given by where γ is the Lorentz factor (electron energy in units of electron rest mass) and EL is the energy of the incident laser photon. The energy of photons scattered at larger angles (|θf|> 0) is lower as explained in Ref. 5, contributing to the finite bandwidth of an inverse Compton scattering X-ray source. Fig. 2b shows the peak X-ray energy for the next generation CLS design presented in Sec. 3. 3.NOVEL DESIGN FOR AN INVERSE COMPTON SCATTERING SOURCEOur work on developing a state-of-the-art ICS-based gamma-ray source with narrow bandwidth and high spectral density for the ELI-NP project (www.eli-np.ro) has led to new approaches for optimizing an ICS source’s brightness: narrow bandwidth, high spectral density, and narrow beam divergence. Here we present a conceptual design for a next generation CLS that is more than two orders of magnitude brighter than the existing CLS design in operation at the MuCLS. The principle is primarily based upon an increase in electron beam energy from 45 MeV to a maximum of 100 MeV, which yields a lower emittance electron beam, decreased damping time, and higher storage ring current through charge accumulation (top-up mode). The lower emittance, combined with more parallel electron and laser beams, leads to a lower divergence and narrower bandwidth X-ray beam. The ICS X-rays have higher energy due to the higher e-beam energy but can be shifted back down in energy by utilizing an optical cavity with a longer wavelength laser (2 μm instead of 1 μm in the existing CLS). Optical enhancement cavities in the near-infrared region (~1.5 μm) and short-wave infrared (3-5 μm) spectral ranges have been extensively used in spectroscopic applications like Cavity Ring-Down Spectroscopy (CRDS).7,8 These examples demonstrate that technologies of low-loss optical coatings and materials exist and can be used for the 2 μm cavity. Moreover, short-pulse sources with the wavelengths near 2 μm have seen considerable progress in research and productization during the last decade.9,10 3.1Beam parametersThe target beam parameters of the next generation CLS are presented in Table 1 in comparison to the source in operation at the MuCLS. The increase in electron-beam energy to 100 MeV shifts the X-ray photon energy up significantly, even when utilizing a longer-wavelength (2 μm) laser. This design is expected to achieve a brightness of more than 2.5 orders of magnitude higher than the MuCLS at the respective peak energy. While we currently expect to achieve an X-ray energy as low as 30 keV with a 2 μm laser, this limit might be reduced somewhat through further optimization of the design. Table 1.Target beam parameters of the next generation CLS. (*) The lower end of the energy range might be reduced somewhat through further optimization of the design.
With this increase in brightness, it is now practical to utilize the tradeoff of beam divergence, flux and bandwidth that is possible with ICS sources. This is illustrated in Fig. 3 and quantified in Table 1. ICS generates a cone of X-rays, the usable portion of which is defined by an aperture. If the highest flux or large beam size are desired, a large aperture is used (typical values are several milliradians). This is useful, for example, for direct-beam imaging of large samples. However, because ICS photons scattered at larger angles have a lower energy, a large beam divergence increases the bandwidth of the X-ray beam. If narrow bandwidth is desired, a smaller beam divergence can be used. Smaller beam size is also preferred for focused beam applications, since the achievable aperture with X-ray optics is limited in particular at higher energies. 3.2Conceptual designA conceptual design of the instrument is shown in Fig. 4. Electron bunches are generated in the photocathode illuminator and accelerated to their target energy in the electron beam injector (linear accelerator). The electron bunches are then injected into the storage ring where they circulate at a rate of approximately 25 MHz. Charge losses are compensated by frequent top-up injection. One leg of the storage ring is shared with a low-loss, power-enhancement optical cavity which is fed by a mode-locked laser that is frequency-matched to the storage ring cycle frequency. With a gain of approximately 10,000, several hundred kilowatts of laser power can be stored in the optical cavity. Laser pulses and electron bunches collide at the interaction point and generate X-rays via inverse Compton scattering. 3.3Comparison with synchrotron and laboratory sources: Flux densityFlux density of the X-ray beam is relevant for techniques such as direct-beam imaging or radiotherapy. Here we compare the flux density achievable with the next generation CLS to select synchrotron beamlines and laboratory sources. For micro-CT imaging, synchrotron beamlines and the CLS operate in a quasi-parallel-beam geometry, where distance from the source is generally fixed and spatial resolution is determined by the detector. This is limited to slightly below 1 μm for state-of-the-art X-ray cameras based on scintillators with optical magnification. Most laboratory-based micro-CT systems operate in a divergent beam geometry with small source spot size, large geometric magnification, and large detector pixels. There are also some systems operating closer to a parallel beam geometry with moderate to large source spot size, low geometric magnification, and small detector pixels. For a divergent beam, flux density scales inversely with the square of the distance from the source, and distance of sample and/or detector from the source can be varied within some constraints in most instruments. Generally, the experimental geometry including source-to-sample and sample-to-detector distances is optimized to provide the highest throughput while accommodating the size of the sample and achieving the desired imaging resolution, which is a function of source size, detector resolution and geometry. To provide a realistic flux density comparison, for laboratory sources we have therefore estimated the geometry and corresponding source-to-sample distance to provide imaging resolution around 1 μm with a realistic and optimized detector setup. Fig. 5 shows the flux density for the next generation CLS, select synchrotron beamlines and estimates of a liquid metal jet source and a transmission microfocus source. With laboratory sources, for direct beam imaging the entire broadband spectrum of the source is used (sometimes filtered with a transmission filter). The data point in the plot represents the approximate mean energy of the spectrum, while the flux density shown includes all the photons in the entire spectrum. Many synchrotron beamlines can operate with different types of monochromators (multilayer monochromators typically providing a bandwidth of order 10-2, or crystal monochromators providing bandwidth of order 10-4), or even “pink” beams (raw source spectrum filtered by upstream beamline optics). Because for most direct beam applications, bandwidth is not critical, we have used pink beam or multilayer monochromator data where available. The figure quantitatively shows the performance drop of synchrotron beamlines, and performance increase of the CLS, towards higher energies, so that the performance of the next generation CLS is comparable to some synchrotron beamlines at those energies. A detailed list of the instruments used for comparison, incl. references, is given in Table 2 in the Appendix. 3.4Comparison with synchrotron and laboratory sources: Focused fluxHere we compare the focused flux achievable with the next generation CLS with select synchrotron beamlines and laboratory systems. Based on the CLS beam properties with source size on the order 100 μm FWHM, divergence of a few milliradians and a minimum source-to-optic distance of about 2 m, focusing optics with several millimeters acceptance are needed to make best use of the available flux. Therefore, we consider a multilayer-coated KB-mirror pair in 1:1 focusing geometry that was custom-designed for and has been demonstrated with the first generation CLS from 12-20 keV with ~4 mrad beam acceptance (unpublished). We have analyzed the design for use with the next generation CLS and concluded that it will work with a beam acceptance of ~1 mrad up to 90 keV X-ray energy. X-ray optics such as compound refractive lenses or zone plates have significantly smaller apertures would therefore utilize only a small fraction of the available flux. Therefore we consider focal spot sizes on the order of 100 to a few hundred microns, relevant for applications such as diffraction or moderate spot size scanning probes. This is also comparable to the spot size achievable with standard laboratory sources combined with polycapillary or Montel optics. Focal spot sizes in the micron or submicron range are generally not achievable with meaningful flux with either ICS or standard laboratory sources so are excluded from the comparison here. Focusing beamlines and laboratory source / focusing optics combinations with spot sizes from sub-100 micron to several hundred microns were researched online. It should be noted that for laboratory sources, the highest energy practically available for focusing is 24 keV, the Kα emission line of Indium used in a liquid metal jet source. The Compact Light Source therefore can provide focused beam capabilities at higher energies up to at least 90 keV in a laboratory setting, which would otherwise only be available at synchrotron beamlines. Fig. 6 plots focused flux, normalized to 100 x 100 μm2, as a function of energy for the next generation CLS, MuCLS, and select synchrotron beamlines and laboratory systems. Focal spot sizes have been categorized into the 50-150 μm range and >150 μm (round equivalent) and are represented by different-size data points in the plot. As for flux density discussed above, the performance of the next generation CLS is comparable to some synchrotron beamlines in particular at high energies. Details and references are provided in Table 3. 4.APPLICATIONSThe next generation Compact Light Source design described in this paper can provide capabilities in particular at higher X-ray energies that are found at only a small number of synchrotron beamlines. The high flux density at higher X-ray energies is ideal for micro-CT imaging of denser samples, for example found in materials or geosciences. It also enables biomedical imaging of larger (several centimeter scale) samples. The high flux density should enable dynamic imaging on time scales not accessible with standard laboratory instruments, on the order of milliseconds for 2D radiography, or seconds for 3D tomography. Fast imaging is also instrumental for high-throughput sample screening, necessary for some industrial applications. In imaging, energy tunability allows optimizing the contrast for specific sample sizes or compositions. It also enables spectroscopic (K-edge subtraction) imaging, for example using Gadolinium contrast agents (K-edge at 50 keV). Also, monochromatic beams provide superior image contrast and avoid beam-hardening artifacts typically present with standard laboratory sources, and are more dose-efficient for radiation-sensitive samples. Utilizing lower beam divergence, the use of X-ray optics becomes feasible even at higher energies, which enables focused beam applications otherwise not possible in a laboratory setting. Examples include diffuse scattering or pair distribution function, which are best performed at energies of 50-100 keV. The substantial increase in data acquisition speed compared to standard laboratory sources enables dynamic diffraction studies otherwise only available at a synchrotron. Lastly, the high flux density and energy up to 180 keV enables radiotherapy applications with monochromatic beams that are currently only possible at synchrotron radiation facilities. Examples include microbeam radiation therapy (MRT) or metal nanoparticle-based radiotherapy. A compact source with appropriate beam parameters provides a pathway to translating such applications into a clinical setting. 5.CONCLUSIONSThe gap between conventional laboratory and synchrotron X-ray sources (see also Sect. 2-4 [15]) can be bridged by an inverse Compton scattering source, which can provide synchrotron-like beam properties such as high flux, monochromaticity, tunability and partial coherence in a laboratory setting. While synchrotron performance typically drops towards higher energies, ICS sources improve, so that at high energies ICS performance can be comparable to some synchrotron beamlines. An ICS source can thus form the cornerstone of a local, multidisciplinary user facility, serving a wide range of applications and X-ray techniques. This enables spreading synchrotron-like capabilities geographically and building local communities in X-ray instrumentation and applications. Local control of experimental time and access allows overcoming some of the limitations of the standard synchrotron access model, characterized by competitive beamtime proposals and assignment of 2-3-day slots of beamtime with weeks or months of waiting time. For example, it enables iterative and longitudinal studies that require short amounts of measurement time at regular intervals over periods of weeks or months, which is typically not possible at synchrotron beamlines. Difficult-to-transport samples (e.g., in vivo, precious, or hazardous samples) can be accommodated more easily at a local facility. Sample pre-screening at a local facility could greatly increase the efficiency of expensive synchrotron or free electron laser facilities. Furthermore, ICS sources provide a pathway to translating some synchrotron applications to clinical or possibly industrial settings. AppendicesAPPENDIXTables 2 and 3 below list the synchrotron beamlines and laboratory sources used for the performance comparisons in Secs. 3.3 and 3.4. Table 2.Select synchrotron beamlines and laboratory source estimates used for flux density comparison in Sec. 3.3.
Table 3.Select synchrotron beamlines and laboratory source estimates used for focused flux comparison in Sec. 3.4.
REFERENCESHajima, R.,
“Status and Perspectives of Compton Sources,”
Physics Procedia, 84 35
–39
(2016). https://doi.org/10.1016/j.phpro.2016.11.007 Google Scholar
Jacquet, M.,
“Potential of compact Compton sources in the medical field,”
Physica Medica, 32
(12), 1790
–1794
(2016). https://doi.org/10.1016/j.ejmp.2016.11.003 Google Scholar
Eggl, E., Dierolf, M., Achterhold, K., Jud, C., Günther, B., Braig, E., Gleich, B. and Pfeiffer, F.,
“The Munich Compact Light Source: initial performance measures,”
Journal of Synchrotron Radiation, 23
(5), 1137
–1142
(2016). https://doi.org/10.1107/S160057751600967X Google Scholar
“Munich Compact Light Source (MuCLS),”
(2021). Google Scholar
Hornberger, B., Kasahara, J., Gifford, M., Ruth, R. and Loewen, R.,
“A compact light source providing high-flux, quasi-monochromatic, tunable X-rays in the laboratory,”
Advances in Laboratory-based X-Ray Sources, 2SPIE, San Diego, United States
(2019). https://doi.org/10.1117/12.2527356 Google Scholar
Günther, B., Gradl, R., Jud, C., Eggl, E., Huang, J., Kulpe, S., Achterhold, K., Gleich, B., Dierolf, M. and Pfeiffer, F.,
“The versatile X-ray beamline of the Munich Compact Light Source: design, instrumentation and applications,”
Journal of Synchrotron Radiation, 27
(5),
(2020). https://doi.org/10.1107/S1600577520008309 Google Scholar
Thorpe, M. J., Balslev-Clausen, D., Kirchner, M. S. and Ye, J.,
“Cavity-enhanced optical frequency comb spectroscopy: application to human breath analysis,”
Opt. Express, 16
(4), 2387
(2008). https://doi.org/10.1364/OE.16.002387 Google Scholar
Long, D. A., Fleisher, A. J., Liu, Q. and Hodges, J. T.,
“Ultra-sensitive cavity ring-down spectroscopy in the mid-infrared spectral region,”
Opt. Lett., 41
(7), 1612
(2016). https://doi.org/10.1364/OL.41.001612 Google Scholar
Mirov, S. B., Fedorov, V. V., Martyshkin, D., Moskalev, I. S., Mirov, M. and Vasilyev, S.,
“Progress in Mid-IR Lasers Based on Cr and Fe-Doped II–VI Chalcogenides,” IEEE J. Select. Topics,”
Quantum Electron., 21
(1), 292310
(2015). Google Scholar
Ma, J., Qin, Z., Xie, G., Qian, L. and Tang, D.,
“Review of mid-infrared mode-locked laser sources in the 2.0 μm–3.5 μm spectral region,”
Applied Physics Reviews, 6
(2), 021317
(2019). https://doi.org/10.1063/1.5037274 Google Scholar
Weitkamp, T., Tafforeau, P., Boller, E., Cloetens, P., Valade, J.-P., Bernard, P., Peyrin, F., Ludwig, W., Helfen, L., Baruchel, J., Denecke, M. and Walker, C. T.,
“Status and evolution of the ESRF beamline ID19,”
in Proceedings of the 20th International Congress,
(2010). https://doi.org/10.1063/1.3399253 Google Scholar
Stevenson, A. W., Crosbie, J. C., Hall, C. J., Häusermann, D., Livingstone, J. and Lye, J. E.,
“Quantitative characterization of the X-ray beam at the Australian Synchrotron Imaging and Medical Beamline (IMBL),”
J Synchrotron Rad, 24
(1), 110
–141
(2017). https://doi.org/10.1107/S1600577516015563 Google Scholar
Drakopoulos, M., Connolley, T., Reinhard, C., Atwood, R., Magdysyuk, O., Vo, N., Hart, M., Connor, L., Humphreys, B., Howell, G., Davies, S., Hill, T., Wilkin, G., Pedersen, U., Foster, A., De Maio, N., Basham, M., Yuan, F. and Wanelik, K.,
“I12: the Joint Engineering, Environment and Processing (JEEP) beamline at Diamond Light Source,”
Journal of Synchrotron Radiation, 22
(3),
(2015). https://doi.org/10.1107/S1600577515003513 Google Scholar
Larsson, D. H., Vågberg, W., Yaroshenko, A., Yildirim, A. Ö. and Hertz, H. M.,
“High-resolution short-exposure small-animal laboratory x-ray phase-contrast tomography,”
Scientific Reports, 6
(2016). https://doi.org/10.1038/srep39074 Google Scholar
D. Bleiner,
“The Science and Technology of X-ray Lasers: A 2020 Update Proc,”
(2021). Google Scholar
|