20 June 2024 Ultrabroadband VIS-NIR spectrometer for mouse retinal OCT
Shau Poh Chong, Peter Török
Author Affiliations +
Abstract
Hyperspectral OCT imaging of retina within the visible-near-infrared wavelengths could further improve the achievable axial resolution and harness the spectroscopic information for biomarker of retinal diseases . Previous efforts to develop OCT covering both visible and near-infrared wavelengths were hampered by the challenges in designing broadband spectrometer with adequate performance. Here, we described a design and implementation of a broadband spectrometer that could achieve spectral range of more than 450 nm with central wavelength around 680 nm. With it, we further demonstrate in vivo imaging of mouse retina using a fiber-based OCT system.
Conference Presentation
(2024) Published by SPIE. Downloading of the abstract is permitted for personal use only.
Shau Poh Chong and Peter Török "Ultrabroadband VIS-NIR spectrometer for mouse retinal OCT", Proc. SPIE 13006, Biomedical Spectroscopy, Microscopy, and Imaging III, 130060M (20 June 2024); https://doi.org/10.1117/12.3022267