Paper
4 April 1994 Time-correlated single-photon counting using a confocal scanning laser microscope
Hjalmar Brismar
Author Affiliations +
Proceedings Volume 2184, Three-Dimensional Microscopy: Image Acquisition and Processing; (1994) https://doi.org/10.1117/12.172081
Event: IS&T/SPIE 1994 International Symposium on Electronic Imaging: Science and Technology, 1994, San Jose, CA, United States
Abstract
A system for 3D lifetime measurements of fluorophores has been developed and applied to multiple stained specimens. The system is built around a picosecond detector system for time correlated single photon counting and a confocal scanning laser microscope. It is capable of measuring lifetimes in the interval 1 - 20 ns, with a resolution of approximately 10 ps. By studying the lifetime of the fluorescence in a specimen it is possible to identify different fluorophores even if their emission spectra overlap. In addition, information regarding the chemical environment may be obtained due to effects from the environment on the fluorescence lifetime of the fluorophore.
© (1994) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Hjalmar Brismar "Time-correlated single-photon counting using a confocal scanning laser microscope", Proc. SPIE 2184, Three-Dimensional Microscopy: Image Acquisition and Processing, (4 April 1994); https://doi.org/10.1117/12.172081
Lens.org Logo
CITATIONS
Cited by 1 scholarly publication.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Luminescence

Data modeling

Picosecond phenomena

Glasses

Microscopes

Rhodamine

Microscopy

Back to Top