Paper
2 November 2001 Ultrahigh resolution OCT using continuum generation in an air-silica microstructure optical fiber
Ingmar Hartl, Xing De Li, Christian C. Chudoba, Tony H. Ko, James G. Fujimoto, Robert S. Windeler
Author Affiliations +
Abstract
We demonstrate ultrahigh resolution optical coherence tomography using the continuum generation in an air-silica microstructure fiber. A broadband OCT system was developed, supporting a bandwidth of 370 nm at 1.3 micrometers center wavelength. We achieved longitudinal resolutions of 2.5 micrometers in air, or approximately 2 micrometers in tissue. This is to our knowledge the highest longitudinal OCT resolution demonstrated at this wavelength range and the first application of this new light source for OCT. We will also describe the application of this technique for imaging biological tissue in vivo.
© (2001) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Ingmar Hartl, Xing De Li, Christian C. Chudoba, Tony H. Ko, James G. Fujimoto, and Robert S. Windeler "Ultrahigh resolution OCT using continuum generation in an air-silica microstructure optical fiber", Proc. SPIE 4431, Photon Migration, Optical Coherence Tomography, and Microscopy, (2 November 2001); https://doi.org/10.1117/12.447396