Paper
14 July 2004 Evaluation of a scattering liquid flow velocity profile using Doppler optical coherence tomography and dynamic stochastic interference fringe processing
Author Affiliations +
Abstract
Flow velocity of scattering intralipid mixture within the glass capillary is measured using Doppler Optical Coherence Tomography (DOCT) technique. In the DOCT system, interference fringe frequency shift contains information about velocity of mixture. Due to scattering, interference fringe parameters are disturbed and stochastic fringe processing method should be used to evaluate fringe frequency. Interferometric signal frequency and other parameters have been estimated dynamically using discrete nonlinear Kalman filtering method. The stochastic filtering methodology and nonlinear Kalman filtering method are considered with application to experimental DOCT data processing.
© (2004) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Erkki Alarousu, Igor P. Gurov, Jukka T. Hast, Risto A. Myllyla, Tuukka Prykari, and Alexey Zakharov "Evaluation of a scattering liquid flow velocity profile using Doppler optical coherence tomography and dynamic stochastic interference fringe processing", Proc. SPIE 5475, Saratov Fall Meeting 2003: Coherent Optics of Ordered and Random Media IV, (14 July 2004); https://doi.org/10.1117/12.568575
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Filtering (signal processing)

Nonlinear filtering

Electronic filtering

Capillaries

Scattering

Signal processing

Doppler tomography