8 April 2009 Improvements in medical CT image reconstruction accuracy in the presence of metal objects by using x-rays up to 1 MeV
James Clayton, Gary Virshup, Ming Yang, Radhe Mohan, Lei Dong
Author Affiliations +
Abstract
The use of flat panels based on amorphous silicon technology (a-Si) for digital radiography has been accepted by the medical community as having advantages over film-based systems. Radiation treatment planning employs computed tomographic (CT) data sets and projection images to delineate tumor targets and normal structures that are to be spared from radiation treatment. The accuracy of CT numbers is crucial for radiotherapy dose calculations in general but is even more important for charged particle therapy. Conventional CT scanners operating at kilovoltage X-ray energies typically exhibit significant image reconstruction artifacts in the presence of metal implants in human body. We demonstrate a significant improvement in metal artifact reductions and electron density measurements using an amorphous silicon a-Si imager obtained with an X-ray source that can operate at energies up to 1 MeV. The data collected with the higher energy system will be compared and contrasted to CT results obtained at standard kilovoltage energies.
© (2009) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
James Clayton, Gary Virshup, Ming Yang, Radhe Mohan, and Lei Dong "Improvements in medical CT image reconstruction accuracy in the presence of metal objects by using x-rays up to 1 MeV", Proc. SPIE 7295, Health Monitoring of Structural and Biological Systems 2009, 72951P (8 April 2009); https://doi.org/10.1117/12.817401
Lens.org Logo
CITATIONS
Cited by 1 scholarly publication.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
X-rays

Metals

Radiotherapy

X-ray imaging

X-ray computed tomography

Amorphous silicon

Gold

Back to Top