Paper
6 August 2009 Improved spectral radiance responsivity calibration of charge-coupled-device (CCD) imaging spectrometer with an internally illuminated integrating sphere
Shu-rong Wang, Zhen-duo Zhang, Fu-tian Li, Xiao-hu Yang
Author Affiliations +
Abstract
A technique for calibrating the spectral radiance responsivity of the CCD imaging spectrometer with an internally illuminated integrating sphere is described. The spectral radiance of the integrating sphere is obtained by two steps. Firstly, a Spectralon panel diffuser and an ultraviolet spectrometer are combined into a new spectral radiometer which transfers the spectral irradiance of a NIST standard of spectral irradiance to that of the receiving aperture of the integrating sphere. Subsequently, the spectral radiance of the integrating sphere is derived from heat transfer theory for Lambertian radiators. The overall uncertainty of determining the spectral radiance of the integrating sphere is ±2.3%. On the basis of known spectral radiance, the radiance calibration of an available Czerny-Turner imaging spectrometer in our laboratory has been completed in 200-400nm with an uncertainty of about ±2.7%.
© (2009) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Shu-rong Wang, Zhen-duo Zhang, Fu-tian Li, and Xiao-hu Yang "Improved spectral radiance responsivity calibration of charge-coupled-device (CCD) imaging spectrometer with an internally illuminated integrating sphere", Proc. SPIE 7384, International Symposium on Photoelectronic Detection and Imaging 2009: Advances in Imaging Detectors and Applications, 73841J (6 August 2009); https://doi.org/10.1117/12.835573
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Spectroscopy

Integrating spheres

Calibration

Charge-coupled devices

Diffusers

Lamps

Ultraviolet radiation

RELATED CONTENT


Back to Top