Paper
11 February 2011 In-vivo human retina imaging with 5μm axial resolution, at 92000 A-scans/s with 1μm spectral domain OCT system
Author Affiliations +
Abstract
We have outfitted a 1060nm Spectral Domain Optical Coherence Tomography system with a prototype, high speed infrared linear array camera and a custom spectrally reshaped superluminescent diode to achieve 5μm axial resolution at 91,911 A-scans/s image acquisition rate in-vivo in the human retina. 4dB loss of sensitivity was observed as a result of the reduced integration time (7μs) of the fast camera as compared to similar commercially available cameras with 14μs integration time and 47kHz readout rate. Fewer motion artefacts were observed in the retinal images acquired with the fast camera, while the higher axial resolution along with deeper penetration allowed for improved visualization of fine morphological details such as retinal and choroidal capillaries and the deep choroidal structure.
© (2011) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Sepideh Hariri, Patrick Ju Lee, Alireza A. Moayed, and Kostadinka Bizheva "In-vivo human retina imaging with 5μm axial resolution, at 92000 A-scans/s with 1μm spectral domain OCT system", Proc. SPIE 7889, Optical Coherence Tomography and Coherence Domain Optical Methods in Biomedicine XV, 78890K (11 February 2011); https://doi.org/10.1117/12.875746
Lens.org Logo
CITATIONS
Cited by 3 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Image resolution

Optical coherence tomography

Imaging systems

Retina

Cameras

Capillaries

Image quality

Back to Top