27 July 2023 Polarization properties of X-ray tubes used for Imaging X-ray Polarimetry Explorer calibration
Author Affiliations +
Abstract

In this work, we measured the polarization properties of the X-rays emitted from the X-ray tubes, which were used during the calibration of the instrument onboard Imaging X-ray Polarimetry Explorer (IXPE). X-ray tubes are used as a source of unpolarized X-rays to calibrate the response of the gas pixel detectors to unpolarized radiation. However, even though the characteristic fluorescent emission lines are unpolarized, continuum bremsstrahlung emission can be polarized based on the geometry of the accelerated electrons and emitted photons. Hence, characterizing the contribution of polarized X-rays from bremsstrahlung emission is of interest, also for future measurements. We find that, when accelerated electrons are parallel to the emitted photons, the bremsstrahlung emission is unpolarized, and when they are perpendicular, the polarization increases with energy, as expected from the theoretical predictions. A comparison with the theoretical predictions is also shown.

1.

Introduction

X-ray tubes are widely used in various scientific fields as a source of X-rays of known energy, depending on the material. When an electron beam strikes the anode, the characteristic X-ray lines generated in the X-ray tube can be used to calibrate X-ray detectors. Such X-ray tubes were used to calibrate the response of the gas pixel detector (GPD) onboard the Imaging X-ray Polarimetry Explorer (IXPE) to unpolarized photons.13 Even though IXPE is equipped with radioactive sources for onboard in-orbit calibration purposes,4 a comprehensive and extensive calibration had to be conducted on the ground at different energies using X-ray tubes.3,5 For this reason, even though the spectral properties of the X-rays arising from the X-ray tube are well known, their polarization properties are poorly studied because of a lack of suitable polarimeters, which is, in turn, necessary for the calibration of X-ray polarimeters such as the GPD onboard IXPE. Hence, it is crucial to constrain the polarization from these X-ray tubes while using them to calibrate polarimeters such as the GPD onboard IXPE.

Measuring the response of the detector to unpolarized radiation is an essential procedure in the calibration of an X-ray polarimeter because they are not exempt from spurious modulation that has to be carefully studied, calibrated, and fitered out.6 The spurious modulation observed in a detector can be attributed to systematic effects inherent to the detector. Hence, the source of X-rays needs to be unpolarized to measure the spurious modulation independent of the source modulation. Even though fluorescent K-shell emission from X-ray tubes is unpolarized, the continuum bremsstrahlung emission can be polarized based on the geometry of the X-ray tube.7 The contribution of the partially polarized continuum to the unpolarized characteristic line must be addressed due to the finite energy resolution of the detector. Therefore, it is crucial to understand the contribution of polarization from the continuum emission while using these X-ray tubes for measuring the response to unpolarized radiation of any X-ray polarimeter. The manufacturer does not provide the polarization properties of the X-ray tubes used for IXPE calibration because theoretically calculating the polarization of X-ray tubes from the first principle is difficult as it depends on the details of the geometry of the emission. Even if a procedure to decouple the intrinsic response of the instrument and the signal generated by the genuine partial polarization of the X-ray tube has been used for the calibration of detectors onboard IXPE,6 it is important to measure its intrinsic polarization as a cross-check and future reference. Moreover, in general, we notice that, in X-ray astronomy, after 60 years and around 50 space missions, a good reference of cross-calibrations has been achieved8,9 so that any new mission can benefit from ground facilities and a large sample of celestial sources sometimes observed simultaneously with more than one satellite. In the domain of polarimetry, IXPE is measuring the polarization of tens (potentially hundreds) of X-ray sources and is building the first catalog that will be the reference for any future experiment. It is, therefore, imperative to make a substantial effort to the best knowledge of the absolute values of the published results, including the level of systematics.

The experimental verification of X-ray tube polarization properties was only done now, as measuring the polarization in wide X-ray bands was difficult until the recent development of highly sensitive and wide-band photoelectric X-ray polarimeters. Moreover, measurements involving polarization for material science are currently performed in synchrotron facilities. However, acquiring sufficient allocation of time from synchrotron facilities for polarimetric calibrations of GPD-like detectors, which demand prolonged exposure measurements, is impractical. For example, the calibration of a detector unit (DU) of IXPE took 40 days of measurements.3 On the other hand, the constant availability of X-ray tubes renders them a convenient option for prolonged usage and enables adherence to the IXPE mission launch timeline.

In this work, we outline the analysis of the measurements performed to understand the polarization properties of X-ray tubes. Section 2 describes the method of measuring X-ray polarization with photoelectric polarimeters, and Sec. 3 gives a theoretical background of X-ray tubes and expected polarization from bremsstrahlung. Section 4 shows the measurements and results, and in Sec. 5, we conclude by discussing the observations.

2.

Measuring Polarization with IXPE

IXPE is the first dedicated mission with focusing optics in space to measure the polarization of X-rays.1,2 IXPE is a NASA astrophysics small explorer mission developed in collaboration with the Italian Space Agency (ASI) and was launched on December 9, 2021. The IXPE focal plane instrument, comprising three flight DUs, each hosting a GPD and a spare unit, were developed by Istituto Nazionale di Astrofisica/Istituto di Astrofisica e Planetologia Spaziali (INAF-IAPS) in Rome and Istituto Nazionale di Fisica Nucleare (INFN) in Pisa. The GPD is a photoelectric polarimeter that can image the photoelectron track and reconstruct the photoelectron emission direction and the absorption point.

IXPE opens a new window in X-ray astronomy by providing additional information on the polarization degree (PD) and polarization angle (PA) of X-rays along with the energy, time of arrival, and interaction point in the detector. With the addition of polarimetry, our understanding of the geometry and emission mechanism of various X-ray sources can be substantially improved. Specifically, polarization can help break the degeneracy of some geometrical and physical models developed based on spectroscopy alone.

The sensitivity of any X-ray polarimeter can be estimated by a quantity known as the minimum detectable polarization at 99% confidence (MDP99). MDP99 is defined as10,11

Eq. (1)

MDP99=4.29μ×1RS×RS+RBT,
where μ is the modulation factor (measured without background), which is the response of a polarimeter to fully polarized X-rays; T is the exposure time; and RS and RB are the source and background count rate, respectively. When the MDP99 is smaller, the sensitivity of a polarimeter is higher. Hence when μ is larger, the sensitivity of the polarimeter is larger.

In photoelectric absorption, the photoelectron’s emission direction is parallel to the electric field of the incoming photon. Hence, the azimuthal distribution would be sinusoidally modulated depending on the degree and angle of polarization. Because the GPD is a photoelectric polarimeter, given the cross-section for K-shell electrons, the azimuthal angle distribution for photoelectrons (modulation curve) has a cosine dependence, described by the following function:11

Eq. (2)

S(ϕ)=A+Bcos2(ϕϕ0),
where ϕ0 is the PA. The amplitude of modulation is then described as11

Eq. (3)

m=SmaxSminSmax+Smin=B2A+B.
However, the amplitude of modulation is not the PD of the observation and has to be normalized by the modulation factor of the detector to obtain the source PD11

Eq. (4)

p=mμ.

The PD and angle can also be expressed in terms of the Stokes parameters. Stokes parameters are generally used to express the polarization of electromagnetic radiation (I, Q, U, and V). The parameter I represents the intensity of the beam, Q represents the intensity in the +90 and 0 directions, U represents the intensity in the +45 and 45 directions, and V represents clockwise and anticlockwise circular polarization. The circular polarization arises when the electric field vector of X-rays oscillates in a circular pattern as it travels through space. However, in astronomical X-ray polarimetry, we do not consider circular polarization due to the difficulty in measurement. If an X-ray polarimeter is designed to be sensitive only to linear polarization, the presence of circular polarization can potentially reduce the measured degree of linear polarization. An example of a modulation curve of polarized flux is shown in Fig. 1. In terms of the modulation curve parameters, the Stokes parameters are expressed as11

Eq. (5)

I=A+B/2,

Eq. (6)

Q=(B/2)cos(2ϕ0),

Eq. (7)

U=(B/2)sin(2ϕ0).
The modulation curve is expressed in terms of Stokes parameters11 as

Eq. (8)

S(ϕ)=I+Qcos(2ϕ)+Usin(2ϕ).
Now the modulation amplitude (m) and angle (ϕ0) of modulation are expressed in terms of the Stokes parameters as

Eq. (9)

m=(Q2+U2)0.5I,

Eq. (10)

ϕ0=1/2tan1(U/Q).
However, the approach followed in this work and, in general, in the case of IXPE is based on measuring the stokes parameters in an event-by-event approach to account for the subtraction of systematic effects.6,12 For each photon, the Stokes parameters are calculated as

Eq. (11)

qi=2cos(2ϕi),

Eq. (12)

ui=2sin(2ϕi),
where ϕi is the initial photo-electron direction, estimated by the photo-electron track reconstruction algorithm.13,14 For a total number of N events

Eq. (13)

q=ΣiqiN,

Eq. (14)

u=ΣiuiN,
and the modulation and angle is computed as

Eq. (15)

m=(q2+u2)0.5,

Eq. (16)

ϕ0=1/2tan1(u/q).

Fig. 1

Example modulation curve for linearly polarized light, decomposed into stokes parameters as labeled (for modulation amplitude, m=0.5 and PA, ϕ0=60°).

JATIS_9_3_038002_f001.png

3.

X-ray Tubes and Expected Polarization

X-ray tubes emit radiation with the following principle. Electrons are generated from a heated cathode and are accelerated toward an anode by a high voltage. The electrons hit the target anode, get decelerated, and produce X-rays due to bremsstrahlung emission. Bremsstrahlung emission is due to the deflection of the electrons by an atomic nucleus, and the energy lost by the electron is emitted as a photon. The maximum energy of the bremsstrahlung emission depends on the peak accelerating potential of the X-ray tube, and the bremsstrahlung peaks at 1/3rd of the maximum energy.15 The intensity of the bremsstrahlung emission is directly proportional to the atomic number of the target material and the charge of the particle and is inversely proportional to the mass of the particle, in our case, the electron. The spectrum of bremsstrahlung emission in an X-ray tube is described by Kramers law as16

Eq. (17)

dI(λ)=K(λλmin1)1λ2dλ,
where K is a constant depending on the electron beam current, λ is the wavelength of the emitted photons, and λmin is the minimum wavelength of the emitted photons from an X-ray tube for a given applied voltage (V), given by the Duane-Hunt law17

Eq. (18)

λmin=hceV.
Other than the bremsstrahlung emission, the X-ray tubes also emit characteristic fluorescent X-rays when the accelerated electrons hit the target material, kick an electron from the K-shell, and get refilled by an outer shell electron. The energy of the discrete fluorescent lines depends on the binding energy of the electron in the target material. About one percent of the energy of the accelerated electrons is radiated through X-rays. The high voltage applied to the X-ray tube will change the spectral shape, whereas the applied current will change the intensity or normalization.

3.1.

Theoretical Prediction of Polarization from Bremsstrahlung Emission

Theoretical calculations of the cross-section and polarization from the bremsstrahlung emission were already published in the middle of the 20’th century.7,1821 It was found that the bremsstrahlung emission is supposed to be partially polarized. The primary step in determining the polarization is to determine the cross-section. The bremsstrahlung cross-section gives the probability of an electron transiting from one state to another state with the emission of a photon. The electron-nucleus collisions can also be elastic; hence the probability that a photon will be emitted depends on the cross-section of the bremsstrahlung, which is 137 times smaller than the cross-section of electron-nucleus elastic scattering. The differential cross-section (dσ), as a function of the photon energy for an electron emerging through the solid angle dΩ in the non-relativistic limit, which is the case for electron kinetic energies smaller than 10 keV, is expressed as

Eq. (19)

dσ=(Z2e6π2)(pp0)(dkkq4)dΩdΩ0(plp0l)2,
where Z is the atomic number or nuclear charge, e is the elementary charge;, k is the energy of the emitted photon; q is the momentum transferred to the nucleus; p and p0 are the final and initial momentum of the electron, respectively; and pl and p0l are the components of momentum of the electron in the direction of the polarization, respectively. Hence the differential cross section for a specified photon energy further depends on the atomic number of the material, the direction of the initial and emerging electrons, and the momentum of the initial and emerging electrons.

Given the energy and momentum of the initial and final electrons, along with the photon energy and momentum, linear polarization is calculated as

Eq. (20)

P=dσIIIdσIIdσIII+dσII,
where dσIII is the differential cross-section of the polarization vector perpendicular to the plane of scattering, containing the initial electron momentum and photon momentum, and dσII is the differential cross-section of the polarization vector in the plane of scattering and parallel to p.7 It is to be noted that the notations used here are the same as that of Gluckstern et al..7 The expressions for dσII and dσIII

Eq. (21)

dσII=Z2e68πdkkpp0dΩ0{8m2sin2θ0(2E02+m2)p02Δ025E02+2EE0+5  m2p02Δ02p02k2T2Δ02+2(E+E0)p02Δ0+(Lpp0)[4E0m2sin2θ0(3km2p02E)p02Δ04+2E02(E02+E2)m2(9E024EE0+E2)+2m4p02Δ02+kE02+EE0p02Δ0](ϵTpT)[4m2Δ027  kΔ0kp02k2T2Δ04]4ϵpΔ0+(1p02sin2θ02)[(2Lpp0)(2E02EE0m2m2kΔ0)4ϵT(Δ0E)2pT2ϵ(Δ0E)p]},

Eq. (22)

dσIII=Z2e68πdkkpp0dΩ0{5E02+2EE0+m2p02Δ02p02k2T2Δ022kp02Δ0+(Lpp0)[2E02(E02+E2)m2(5E022EE0+E2)p02Δ02+k(E02+EE02m2)p02Δ0]+(ϵTpT)[kΔ0k(p02k2)T2Δ0+4](1p02sin2θ02)[(2Lpp0)(2E02EE0m2m2kΔ0)4ϵT(Δ0E)2pT2ϵ(Δ0E)p]},
where

Eq. (23)

Δ0=E0p0cosθ0,

Eq. (24)

T=p02+k22p0kcosθ0,

Eq. (25)

L=ln(EE0m2+pp0EE0m2pp0),

Eq. (26)

ϵ=ln(E+pEp),

Eq. (27)

ϵT=ln(T+pTp),
and θ0 is the angle between p0 and k. E and E0 represents the initial and final energy of the incoming electron, respectively. The constants and c are taken to be 1. The effects of shielding are complicated. However, they are approximated by replacing k2 by k2+α2p02Δ02 in the logarithmically divergent terms L and ϵT. Hence

Eq. (28)

L=ln((EE0m2+pp0)2m2k2+m2α2p02Δ02),

Eq. (29)

ϵT=12ln((T+p)44  k2Δ02+4α2p02),
where

Eq. (30)

α=Z13m108.

Figures 2 and 3 show the degree of polarization and cross-section with respect to photon energy for a beam of electron of energy 0.1 MeV in aluminum.7 It is seen that the polarization is high around the lower and upper ends of the photon spectrum. A jump in the PA by 90 deg is seen as a change in the sign of the PD at medium energies of 20  keV. The shielding effect decreases the PD, especially at lower energies.

Fig. 2

PD of the outgoing photons as a function of photon energy for incident electron energies of 0.1 MeV for Z=13 (Aluminium) for different values of the angle between incoming electrons and outgoing photons (θ0) as well as with and without shielding effects.

JATIS_9_3_038002_f002.png

Fig. 3

Cross-section of the outgoing photons as a function of photon energy for incident electron energies of 0.1 MeV for Z=13 (Aluminium) for different values of the angle between incoming electrons and outgoing photons (θ0) as well as with and without shielding effects.

JATIS_9_3_038002_f003.png

4.

Measurements and Results

We now present the polarization measured for the X-ray tubes used to calibrate the GPDs onboard IXPE. There are different X-ray tubes used for the calibration of IXPE,3 and there are primarily two kinds, right angle X-ray tubes from Oxford series 5000 (Fig. 4) and head-on X-ray tubes such as the calcium and tungsten anode ones from Hamamatsu, model N1335 (Fig. 5). The experimental setup of the measurements are the same as that given in Refs. 3 and 5. The source and detector were mounted to the instrument calibration equipment (ICE), an equipment specifically designed for calibrating IXPE DUs.3 This equipment encompasses the mechanical frameworks employed to secure the detector and sources, calibration sources, stages for aligning the detector and source beam, test detectors, and their corresponding mechanical assembly (Fig. 3 in Ref. 3). The alignment process involves two stages. The first stage, named ALIGN, aims to synchronize the source to the detector. The second stage, referred to as MEAS, facilitates movement of the detector along the detector plane, orthogonal shifts, azimuthal rotations, and tilting of the detector to align the beam with the detector plane. In the case of right-angle X-ray tubes, the generated X-ray photons are perpendicular to the accelerated beam of electrons. By contrast, in the case of head-on X-ray tubes, the X-ray photons are parallel to the accelerated beam of electrons. During the calibration of the IXPE Instrument, X-ray filters were used to increase the ratio between the fluorescence and the continuum emission as only the former component is unpolarized and of interest for calibration. For our study, such filters were not used as our primary aim is to measure the polarization of the bremsstrahlung continuum. For measuring the polarization, we used an IXPE flight DU called DU FM1 (Detector Unit Flight Model 1), which is the spare unit of the IXPE Instrument and was fully calibrated as the other flight units. The measurements were undertaken in the ICE after mounting DU FM1 and the different X-ray tubes on the ICE. Figure 6 shows the sketch of the geometrical configuration of the X-ray tubes and the detector.

Fig. 4

Photograph of the Iron X-ray tube used as a perpendicular X-ray tube in this work.

JATIS_9_3_038002_f004.png

Fig. 5

Photograph of the tungsten X-ray tube used as a parallel X-ray tube in this work.

JATIS_9_3_038002_f005.png

Fig. 6

(a) A rough sketch of the experimental geometrical configuration of perpendicular and (b) parallel X-ray tubes with respect to the GPD. In the former, the photon beam is perpendicular to the electron beam, whereas in the latter, both are parallel.

JATIS_9_3_038002_f006.png

Measurements were conducted at two arbitrary angles (ϵ1 and ϵ2) that are mutually perpendicular to each other in the detector plane [(ASIC X-Y Axis)] and with respect to the laboratory frame of reference. The measurements at two angles are utilized to ensure that the polarization that is measured originates from the source itself as the polarization vector of the source will be rotated with respect to the reference frame. If the polarization indeed arises from the source and not from any detector systematics, then the measured PA must undergo a 90 deg rotation upon rotating the source. Table 1 outlines the details of the X-ray tubes and measurements used for this work. In the case of the Fe and W X-ray tubes, the high voltage was chosen deliberately to be below the K-shell electron binding energy to avoid the bright K fluorescence line from the material of the tube target.

Table 1

Table outlining the details of the measurement configurations used in this work.

X-ray tubeAnodeFluorescent line energySource configurationAngle
Hamamatsu, model N1335Ca3.695.4 kV, 0.003 mAϵ1, ϵ2
HamamatsuW8.348.0 kV, <0.001 mAϵ1
Oxford seriesFe6.406.1 kV, 0.003 mAϵ1, ϵ2
Oxford series 5000Rh2.706.2 kV, 0.001 mAϵ1, ϵ2

For the analysis of the measurements, the spurious modulation due to the detector systematics has to be taken care of and subtracted carefully as in the case of celestial sources. A method to subtract the contribution from the detector on an event-by-event basis was developed for the IXPE flight detectors.6 We followed this method for subtracting the spurious modulation. A spurious modulation database developed for calibrating IXPE,6 which contains the spurious modulation as a function of spatial and energy bins, was used to subtract the detector contribution while extracting the polarization from the source.

The data calibrated for spurious modulation were then used for further analysis. We used all events within the circular spatial region of radius 2 mm from the center of the detector as this is within the regions of the detector that is calibrated with the highest sensitivity. The data from the selected spatial region were then grouped into four different energy bins, with larger bin sizes at energy ranges with lower count-rates, to allow for sufficient statistics for measuring the polarization. The normalized and spurious modulation corrected Stokes parameters from all events in each bin are then added to get the Stokes parameters and modulation amplitude of the total measurement. The total Stokes parameters and the modulation degree are then divided by the modulation factor of DU FM1 in that particular energy bin to get the final Stokes parameters and the polarization from the X-ray tubes.5 The modulation factor in each energy bin is weighted by the spectral counts in that bin.

In the case of the measurements with the Rh and Ca X-ray tube, we could correct the gain variation during the measurement caused by charging of the gas electron multiplier (GEM), which multiplies the primary charges before collection,22 using the fluorescent line at the 2.7 and 3.69 keV. All of the events were re-scaled by a correction factor: the ratio of the fluorescent line energy and the observed Gaussian line peak energy at different times. For the case of Fe and W X-ray tubes, there were no fluorescent lines as the applied high voltage was lower than the line energy; hence, these measurements are slightly affected by charging. This means that the measured energy in those measurements can have uncertainties of up to 10%.

Figure 7 shows the spectra, and Figs. 8Fig. 9Fig. 1011 show the contours of PD and angle. The corresponding data for these measurements are noted in Tables 2Table 3Table 45. For plotting the spectrum, the 2 to 8 keV is divided into 100 energy bins and normalized by the peak of the spectrum.

Fig. 7

Spectrum of all X-ray tubes used in this work as seen in the GPD for ϵ1. The spectrum of each X-ray tube was normalized by the corresponding peak of the spectrum. The vertical lines indicate the maximum possible energy of the X-rays from the tube corresponding to the applied voltage. For W and Rh tubes, the spectrum has only continuum bremsstrahlung, as the applied voltage is below the characteristic line energy. The spectrum of ϵ2 is identical to that of ϵ1, and thus it has not been presented to avoid redundancy.

JATIS_9_3_038002_f007.png

Fig. 8

(a) and (b) Polar plots of the PD and PA for the measurements of Ca X-ray tube at angles of ϵ1 and ϵ2. Contours indicate 99.7% confidence levels. Colors black, red, blue, and magenta indicate energy ranges 2.0 to 3.0 keV, 3.0 to 3.7 keV, 3.7 to 4.2 keV, and 4.2 to 8.0 keV, respectively. These contours indicate that there are no significant polarization measurements but only constraints of upper limits.

JATIS_9_3_038002_f008.png

Fig. 9

Same as Fig. 8, but for W X-ray tube and only for the angle of ϵ1. Contours indicate 99.7% confidence levels. Colors black, red, blue, and magenta indicate energy ranges 2.0 to 3.25 keV, 3.25 to 4.25 keV, 4.25 to 5.25 keV, and 5.25 to 8.0 keV, respectively.

JATIS_9_3_038002_f009.png

Fig. 10

(a) and (b) Polar plots of the PD and PA for the measurements of Rh X-ray tube at angles of ϵ1 and ϵ2. Contours indicate 99.7% confidence levels. Black, red, blue, and magenta indicate energy ranges of 2.0 to 2.6 keV, 2.6 to 3.2 keV, 3.2 to 4.0 keV, and 4.0 to 7.0 keV, respectively. The increase in polarization with respect to energy can be seen clearly. The 90 deg phase shift observed in the PA between ϵ1 and ϵ2 confirms that the PA measurement undergoes a 90 deg rotation when the source is rotated by 90 deg and hence confirms the source origin of polarization than detector systematics.

JATIS_9_3_038002_f010.png

Fig. 11

(a) and (b) Same as Fig. 10, but for Fe X-ray tube. Contours indicate 99.7% confidence levels. Black, red, blue, and magenta indicate energy ranges of 2.0 to 3.0 keV, 3.0 to 4.0 keV, 4.0 to 5.0 keV, and 5.0 to 8.0 keV, respectively.

JATIS_9_3_038002_f011.png

Table 2

Polarization results from the Ca X-ray tube for ϵ1 and ϵ2.

Energy (keV)Q/I (%)U/I (%)PD (%)PAMDP (%)
ϵ1
2.0 to 3.00.51±0.70.32±0.7<1.315.99  deg±33.44  deg2.11
3.0 to 3.70.03±0.270.19±0.27<0.3849.05  deg±40.95  deg0.83
3.7 to 4.20.25±0.240.49±0.24<0.7931.3  deg±12.63  deg0.74
4.2 to 8.00.7±0.310.38±0.31<1.1175.74  deg±11.08  deg0.94
ϵ2
2.0 to 3.00.34±0.420.54±0.42<1.0528.89  deg±18.97  deg1.27
3.0 to 3.70.58±0.160.04±0.160.59±0.162.19  deg±8.07  deg0.5
3.7 to 4.20.6±0.150.02±0.150.6±0.151.12  deg±7.0  deg0.44
4.2 to 8.00.08±0.190.12±0.19<0.3361.22  deg±37.93  deg0.56

Table 3

Polarization results from the W X-ray tube for ϵ1.

Energy (keV)Q/I (%)U/I (%)PD (%)PAMDP (%)
ϵ1
2.0 to 3.250.86±0.50.31±0.5<1.4280.01  deg±15.65  deg1.52
3.25 to 4.250.51±0.230.22±0.23<0.7878.17  deg±11.95  deg0.7
4.25 to 5.250.43±0.210.55±0.210.7±0.2163.9  deg±8.58  deg0.64
5.25 to 8.00.27±0.190.24±0.19<0.5620.89  deg±14.92  deg0.58

Table 4

Polarization results from the Rh X-ray tube for ϵ1 and ϵ2.

Energy (keV)Q/I (%)U/I (%)PD (%)PAMDP (%)
ϵ1
2.0 to 2.62.3±0.794.42±0.794.98±0.7958.73  deg±4.53  deg2.39
2.6 to 3.21.49±0.465.12±0.465.33±0.4653.13  deg±2.47  deg1.4
3.2 to 4.05.4±0.4214.13±0.4215.12±0.4255.45  deg±0.79  deg1.27
4.0 to 7.07.38±0.2920.93±0.2922.2±0.2954.71  deg±0.37  deg0.88
ϵ2
2.0 to 2.60.85±1.660.53±1.66<2.6673.99  deg±47.41  deg5.04
2.6 to 3.21.74±0.945.96±0.946.21±0.9436.86  deg±4.34  deg2.85
3.2 to 4.06.16±0.8613.63±0.8614.95±0.8632.83  deg±1.64  deg2.6
4.0 to 7.08.93±0.5920.61±0.5922.46±0.5933.29  deg±0.75  deg1.79

Table 5

Polarization results from the Fe X-ray tube for ϵ1 and ϵ2.

Energy (keV)Q/I (%)U/I (%)PD (%)PAMDP (%)
ϵ1
2.0 to 3.05.82±0.6613.28±0.6614.5±0.6656.82  deg±1.3  deg1.99
3.0 to 4.05.53±0.315.44±0.316.4±0.354.86  deg±0.52  deg0.9
4.0 to 5.08.59±0.3123.06±0.3124.61±0.3155.22  deg±0.36  deg0.93
5.0 to 8.011.93±0.4532.5±0.4534.62±0.4555.08  deg±0.38  deg1.38
ϵ2
2.0 to 3.05.93±1.3115.8±1.3116.87±1.3134.72  deg±2.23  deg3.98
3.0 to 4.05.09±0.614.29±0.615.17±0.635.2  deg±1.13  deg1.82
4.0 to 5.08.81±0.6224.09±0.6225.65±0.6234.96  deg±0.69  deg1.88
5.0 to 8.013.31±0.9232.83±0.9135.42±0.9133.97  deg±0.75  deg2.8

The PD was seen to increase with energy for the right angle X-ray tubes (Fe and Rh), whereas for head-on X-ray tubes (Ca and W), the X-rays are unpolarized, and the measured PA in both cases seems to be parallel to the direction of the incoming electrons, as what was expected in the energy range of the GPD. In the case of the Rh and Fe X-ray tube, the polarization increases from around 5% to 22% and around 15% to 35% across the IXPE energy band of 2 to 8 keV (see Tables 4 and 5). Even though the increasing trend of polarization is as expected, the absolute values of the measured polarization are lower than what is expected for an ideal case of single scattering of the electron in a crystal. However, in our case, the mean free path of the electron with an energy of a few keV in Rh and Fe (on the order of nm) would be much less than the thickness of the anode target (a few tens of microns), increasing the probability of multiple scattering of the single electron giving rise to multiple bremsstrahlung photons. This effect would, in turn, dilute and decrease the PD. It was previously seen that, even for very thin targets and at larger energies, the measured polarization was smaller than the theoretically expected value.23 In the case of Ca and W X-ray tubes, we obtained an upper limit closer to 1% or less (see Tables 2 and 3). At some energies, we have a measurement slightly above the MDP, but smaller than 1%, and it could be due to some unknown systematic such as the multiple scattering mentioned above. The shift in the PA for ϵ2 by 90 deg with respect to ϵ1 further indicates that the polarization is coming from the source rather than a detector systematic such as the spurious modulation (which indeed we subtract). If the polarization was a detector effect, the PA should not have rotated by 90 deg while rotating the illuminating source. It can also be seen that the highest energy in the photon spectrum matches the applied high voltage of the tube.

The polarization measured at ϵ1 is fitted with the theoretical function [Eqs. (2022)], allowing the applied voltage (V) and angle between the incoming electron and the outgoing photon (θ) left for the fit to be determined. The results of the fit are shown in Fig. 12. The solid red line is the fit to the data, and the green lines indicate the expected theoretical value based on V and θ. For the case of the perpendicular X-ray tubes, the opening angle of the anode crystal with respect to the aperture toward the detector is 11 deg, and hence to account for it within the theoretical estimate, we shade the space around the line assuming an error of 11 deg from the tube. However, the deviation is well beyond this limit. Table 6 outlines the fit results for all X-ray tubes. In the case of the parallel X-ray tubes, the fitted θ is closer to zero, and the fitted V is 3-4 times higher than the real applied voltage. This is mostly because of the non-zero values of polarization. In the case of the perpendicular X-ray tubes, both the θ and V are different from the real applied values. V is higher by a factor of 2, and θ is smaller by a factor of 3. This mismatch of values could be due to the multiple electron scattering effects.

Fig. 12

(a)–(d) The polarization fraction as a function of the photon energy for all X-ray tubes (at ϵ1). The black points show the measured polarization. The solid green line indicates the expected value from the theory, and the shaded green region in the case of Fe and Rh tubes shows the range due to the non-zero opening angle subtended by the crystal on the detector. The red line indicates the values of the best fit. The negative polarization fraction means that the PA is parallel to the electron direction, and the positive indicates the PA perpendicular to the incoming electron and outgoing photon.

JATIS_9_3_038002_f012.png

Table 6

Parameters of the fit (applied voltage V, angle between incoming electrons and outgoing photons θ) to the measured polarization as a function of photon energy. The real values are also shown to highlight the discrepancy.

X-ray tubeVreal (kV)θrealVfit (kV)θfit
Ca5.40 deg19.1±1.90.5  deg±6.5  deg
W8.00 deg23.8±0.20.05  deg±0.01  deg
Rh6.290 deg13.4±9.028.9  deg±23.6  deg
Fe6.190 deg12.7±10.835.3  deg±40.7  deg

5.

Conclusions

In this work, we calculated the polarization of the continuum bremsstrahlung emission from X-ray tubes used to calibrate the GPD onboard IXPE. The theoretical predictions of polarization from bremsstrahlung depend on the angle between the incoming electron and the emitted photon (θ0). When θ0=0  deg, the polarization is supposed to be constant at 0 deg, and when θ0=90  deg, the polarization is supposed to decrease until a certain energy and then increase. In the case of X-ray tubes, the θ0 depends on the geometrical configuration of the X-ray tube. We used two X-ray tubes with θ0=0  deg and two X-ray tubes with θ0=90  deg.

For perpendicular X-ray tubes, we find that the measured PDs are significantly lower than the ones predicted by theory, despite a good match between the observed PA and the overall trend of the PD with energy in the 2 to 8 keV range. This mismatch of the absolute value could be due to the multiple electron scattering in the anode crystal, which is relevant in 2 to 8 keV. This could not be tested as the manufacturer does not provide the dimensions of the anode crystal. Nonetheless, this scenario is plausible, considering that the mean free path of electrons with energies in the range of a few keVs in Ca, W, Rh, and Fe is on the order of nanometers, whereas the thickness of the anode could exceed micrometers. For parallel X-ray tubes, we obtained upper limits consistent with the predictions.

Our work experimentally shows the differences in the continuum polarization of two different configurations of X-ray tubes. The fact that the general trend of PD and PA is coherent with what is expected is positive. However, our work aimed to quantitatively fix the values of practical implementation to be used to calibrate detectors of IXPE or any other detector. These measurements can now form a base while using these X-ray tubes in the future to understand the response of any X-ray polarimeter to unpolarized X-rays.

Acknowledgments

The IXPE is a joint US and Italian mission. The US contribution is supported by the National Aeronautics and Space Administration (NASA) and led and managed by its Marshall Space Flight Center (MSFC), with industry partner Ball Aerospace (Grant No. NNM15AA18C). The Italian contribution is supported by the Italian Space Agency (Agenzia Spaziale Italiana, ASI) (Grant Nos. ASI-OHBI-2017-12-I.0, ASI-INAF-2017-12-H0 and ASI-INFN-2017.13-H0) and its Space Science Data Center (SSDC) (Grant Nos. ASI-INAF-2022-14-HH.0 and ASI-INFN 2021-43-HH.0), and by the Istituto Nazionale di Astrofisica (INAF) and the Istituto Nazionale di Fisica Nucleare (INFN) in Italy. This research used data products provided by the IXPE Team (MSFC, SSDC, INAF, and INFN) and distributed with additional software tools by the High-Energy Astrophysics Science Archive Research Center at NASA Goddard Space Flight Center.

References

1. 

P. Soffitta et al., “The instrument of the imaging x-ray polarimetry explorer,” Astron. J., 162 208 https://doi.org/10.3847/1538-3881/ac19b0 ANJOAA 0004-6256 (2021). Google Scholar

2. 

M. C. Weisskopf et al., “Imaging x-ray polarimetry explorer: prelaunch,” J. Astron. Telesc. Instrum. Syst., 8 (2), 026002 https://doi.org/10.1117/1.JATIS.8.2.026002 (2022). Google Scholar

3. 

F. Muleri et al., “The IXPE instrument calibration equipment,” Astropart. Phys., 136 102658 https://doi.org/10.1016/j.astropartphys.2021.102658 APHYEE 0927-6505 (2022). Google Scholar

4. 

R. Ferrazzoli et al., “In-flight calibration system of imaging x-ray polarimetry explorer,” J. Astron. Telesc. Instrum. Syst., 6 048002 https://doi.org/10.1117/1.JATIS.6.4.048002 (2020). Google Scholar

5. 

A. Di Marco et al., “Calibration of the IXPE focal plane x-ray polarimeters to polarized radiation,” Astron. J., 164 103 https://doi.org/10.3847/1538-3881/ac7719 ANJOAA 0004-6256 (2022). Google Scholar

6. 

J. Rankin et al., “An algorithm to calibrate and correct the response to unpolarized radiation of the x-ray polarimeter onboard IXPE,” Astron. J., 163 39 https://doi.org/10.3847/1538-3881/ac397f ANJOAA 0004-6256 (2022). Google Scholar

7. 

R. L. Gluckstern and M. H. Hull, “Polarization dependence of the integrated bremsstrahlung cross section,” Phys. Rev., 90 1030 –1035 https://doi.org/10.1103/PhysRev.90.1030 PHRVAO 0031-899X (1953). Google Scholar

8. 

S. Sembay et al., “Defining high-energy calibration standards: IACHEC (International Astronomical Consortium for High-Energy Calibration),” Am. Inst. Phys. Conf. Ser., 1248 593 –594 https://doi.org/10.1063/1.3475350 (2010). Google Scholar

9. 

K. K. Madsen et al., “IACHEC cross-calibration of Chandra, NuSTAR, Swift, Suzaku, XMM-Newton with 3C 273 and PKS 2155-304,” Astron. J., 153 2 https://doi.org/10.3847/1538-3881/153/1/2 ANJOAA 0004-6256 (2017). Google Scholar

10. 

M. C. Weisskopf, R. F. Elsner and S. L. O’Dell, “On understanding the figures of merit for detection and measurement of x-ray polarization,” Proc. SPIE, 7732 77320E https://doi.org/10.1117/12.857357 PSISDG 0277-786X (2010). Google Scholar

11. 

T. E. Strohmayer and T. R. Kallman, “On the statistical analysis of x-ray polarization measurements,” Astrophys. J., 773 103 https://doi.org/10.1088/0004-637X/773/2/103 ASJOAB 0004-637X (2013). Google Scholar

12. 

F. Kislat et al., “Analyzing the data from X-ray polarimeters with Stokes parameters,” Astropart. Phys., 68 45 –51 https://doi.org/10.1016/j.astropartphys.2015.02.007 APHYEE 0927-6505 (2015). Google Scholar

13. 

R. Bellazzini et al., “Micropattern gas detector for x-ray polarimetry,” Proc. SPIE, 4843 372 –382 https://doi.org/10.1117/12.459380 PSISDG 0277-786X (2003). Google Scholar

14. 

R. Bellazzini et al., “Novel gaseous x-ray polarimeter: data analysis and simulation,” Proc. SPIE, 4843 383 –393 https://doi.org/10.1117/12.459381 PSISDG 0277-786X (2003). Google Scholar

15. 

B. K. Agarwal, X-Ray Spectroscopy: An Introduction, 2nd ed.Springer Berlin Heidelberg( (2013). Google Scholar

16. 

H. A. Kramers, “XCIII. On the theory of x-ray absorption and of the continuous x-ray spectrum,” Lond. Edinburgh Dublin Philos. Mag. J. Sci., 46 (275), 836 –871 https://doi.org/10.1080/14786442308565244 (1923). Google Scholar

17. 

W. Duane and F. Hunt, “On x-ray wave-lengths,” Phys. Rev., 6 166 –172 https://doi.org/10.1103%2FPhysRev.6.166 PHRVAO 0031-899X (1915). Google Scholar

18. 

A. Sommerfeld, “Über die Beugung und Bremsung der Elektronen,” Annalen der Physik, 403 257 –330 https://doi.org/10.1002/andp.19314030302 ANPYA2 0003-3804 (1931). Google Scholar

19. 

W. Heitler and F. Sauter, “Stopping of fast particles with emission of radiation and the birth of positive electrons,” Nature, 132 892 https://doi.org/10.1038/132892a0 (1933). Google Scholar

20. 

P. Kirkpatrick and L. Wiedmann, “Theoretical continuous x-ray energy and polarization,” Phys. Rev., 67 321 –339 https://doi.org/10.1103/PhysRev.67.321 PHRVAO 0031-899X (1945). Google Scholar

21. 

R. L. Gluckstern, M. H. Hull and G. Breit, “Polarization of bremsstrahlung radiation,” Phys. Rev., 90 1026 –1029 https://doi.org/10.1103/PhysRev.90.1026 PHRVAO 0031-899X (1953). Google Scholar

22. 

L. Baldini et al., “Design, construction, and test of the gas pixel detectors for the IXPE mission,” Astropart. Phys., 133 102628 https://doi.org/10.1016/j.astropartphys.2021.102628 APHYEE 0927-6505 (2021). Google Scholar

23. 

R. W. Kuckuck and P. J. Ebert, “Linear polarization of low-energy-electron bremsstrahlung,” Phys. Rev. A, 7 456 –462 https://doi.org/10.1103/PhysRevA.7.456 (1973). Google Scholar

Biography

Ajay Ratheesh received his PhD in astronomy and astrophysics through the joint PhD program offered by Sapienza University of Rome, University of Rome “Tor Vergata,” and INAF, Italy, in 2021. Currently, he is a post-doctoral researcher affiliated with INAF-IAPS in Rome. He is a member of the IXPE collaboration.

Biographies of the other authors are not available.

CC BY: © The Authors. Published by SPIE under a Creative Commons Attribution 4.0 International License. Distribution or reproduction of this work in whole or in part requires full attribution of the original publication, including its DOI.
Ajay Ratheesh, John Rankin, Enrico Costa, Ettore Del Monte, Alessandro Di Marco, Sergio Fabiani, Fabio La Monaca, Fabio Muleri, Alda Rubini, Paolo Soffitta, Luca Baldini, Massimo Minuti, Michele Pinchera, and Carmelo Sgrò "Polarization properties of X-ray tubes used for Imaging X-ray Polarimetry Explorer calibration," Journal of Astronomical Telescopes, Instruments, and Systems 9(3), 038002 (27 July 2023). https://doi.org/10.1117/1.JATIS.9.3.038002
Received: 19 February 2023; Accepted: 28 June 2023; Published: 27 July 2023
Advertisement
Advertisement
KEYWORDS
X-rays

Polarization

Electrons

Photons

Modulation

Polarimetry

X-ray detectors

RELATED CONTENT

A possibility to extend the IXPE energy band