Total internal reflection fluorescence microscopy (TIR-FM) is an important tool in life science. Taking advantage of a very thin optical excitation depth formed by an evanescent wave above a glass substrate, TIR fluorescence detection achieves an exceptional axial resolution and allows for the study of important cellular processes, in particular at or near the cellular membrane.1 Prominent biological applications of TIR-FM include real-time in vivo observations of dynamics of molecular motors,2, 3 or of various membrane trafficking events, such as vesicle fusion or signal transduction after receptor binding.4, 5 To extend the sensitivity and observation of such cellular assays, it is important to combine TIR-FM with fluorescence measurement and single-molecule fluctuation analysis. Fluorescence correlation spectroscopy enables to expose biological reactivity such as binding events on the basis of temporal fluctuations in the fluorescence emission of single molecules.6, 7 Dual-color fluorescence cross-correlation spectroscopy (FCCS) represents a further development of fluctuation spectroscopy, which achieves a significantly improved selectivity and applicability.8 FCCS institutes two differently colored labels and realizes the distinction between coinciding and separated occurrence of these labels. It thus introduces a useful tool to monitor various biological assays including molecular binding between two differently labeled binding partners. Typically, FCCS is based on a confocal epi-illumination microscope with multicolor laser excitation and single-photon detection. Nonetheless, for several applications, TIR illumination can provide further benefits, for instance when ligands interacting with receptors immobilized onto a surface are to be studied.9, 10, 11, 12 First, in TIR-FCCS the molecules are only excited near the glass-sample interface, which simplifies the discrimination between freely diffusing and immobilized particles. Second, the most important prerequisite for efficient single-molecule detection is a high fluorescence count rate per molecule (CPM). Using TIR excitation and exploiting the anisotropic emission of the fluorescence generated at the glass-sample interface, much higher can be extracted from the investigated molecules.13 In this work, we present for the first time a dual-color excitation scheme for TIR single-molecule detection, which enables us to perform FCCS at glass-sample interfaces with high sensitivity. The improved performance of TIR-FCCS was verified on a synthetic binding assay, including dual-labeled double-stranded DNA molecules in aqueous solution, demonstrating its potential for biological and cellular applications. Figure 1 shows our dual-color TIR-FCCS setup. Two linear polarized laser beams were expanded to an diameter of and collinearly aligned to the microscope objective (BE & BS; beam expansion ). The laser powers were controlled by neutral density filters. An achromatic lens focused the beams into the back-focal plane (BFP) of the high numerical aperture (NA) oil immersion objective ( -Plan-Fluar with Immersoil™ 518F, Carl Zeiss Jena, Jena, Germany), which resulted in circular spots with diameters at the cover slide-sample interface of (blue) and (red), respectively. In the BFP, a beam foci offset of from the optical axis resulted in a super-critical angle illumination, i.e., a -confinement below resulting from the evanescent field excitation. Due to the epi-illumination, the fluorescent light was collected with the same high NA objective toward the pinholes and the single-photon detectors. The pinholes were realized by two multimode fibers with a core diameter of . Dichroic mirrors and bandpass filters blocked the back-reflected laser light by more than OD10 and provided an excellent filtering of the green and red fluorescence. The fluorescence photons were detected by fiber-coupled single photon counting modules (SPCM-AQR-14-FC, PerkinElmer Optoelectronics, Wiesbaden, Germany; ASY50/105 silica fibers, Thorlabs Inc., Grünberg, Germany), whose signals (number of photons over time) were recorded and correlated with a USB hardware correlator (Flex02-08D, Correlator.com, Bridgewater, New Jersey). A focusing lens, a plate, and a dichroic mirror were moved in one block by a linear translator for adjusting the position of the beam foci in the BFP of the objective. In this way, the excitation angle could be adjusted without altering the optical path length between the focusing lens and the objective. The configuration can easily be changed to a conventional confocal epi-illumination by removing the focusing lens and by centering the collimated beams in the BFP. As a synthetic binding assay, we used free rhodamine green (RhG) and Cy5 fluophores, and a 40 base pair double-stranded DNA labeled with Alexa488 and Cy5 (Zeiss cross-correlation standard). The laser powers were optimized with the dsDNA sample and the background was measured with a NaCl/EDTA/TRIS pH 8.0 buffer. The cross talk between the detection channels was calibrated with free RhG and Cy5, respectively. For cross-correlation measurements, different mixtures of the double-labeled dsDNA sample and free RhG/ Cy5 solutions were used and investigated with identical laser powers during a time interval of . Plasma cleaning of the cover slides improved the results due to a strong suppression of unspecific binding at the glass surface. For free diffusion the following model equation was used:13 Here, indices and represent the green and red detection channels; hence and are the autocorrelations of the signals in the green and red detection channels, respectively, and is the cross-correlation of both signals. The index represents the diffusing species: for RhG, for Cy5, and for the dsDNA. Also, is the measured background count rate, is the total count rate, is the CPM in channel of species , are the number of molecules in the effective sampling volume, and are the triplet amplitudes and the triplet correlation times, respectively, are the offsets at infinite lag time , and describes the diffusion and is given by13 where and are the axial and lateral diffusion times of species , respectively, and the scaled complementary error function is . A multidimensional least-squares Gaussian-Newtonion algorithm was used to fit the experimental data to these model equations.For these first experiments, only relative concentrations of the species were of interest. Therefore, a unique diffusion time was used, neglecting channel differences due to an imperfect overlap of the excitation and detection volumes. The triplet term accounts for the triplet contribution in the initial correlation amplitude . The measured background signals were and , respectively. Typically, autofluorescence from the immersion oil as well as from the glass slide were the main contributions to and , respectively. The crosstalk CPM of RhG in the red channel was found to be and assumed constant during the following evaluations. Cy5 did not show any crosstalk in the green channel, hence . Therefore, the CPM were replaced by the crosstalk factor for simplification. Figure 2 shows the correlation results for a mixture of the free fluorophores (negative control). The count rates of this control sample were and . The CPMs were and , respectively. Figure 3 shows the results for the Cy5 and Alexa488 labeled dsDNA correlation (positive control). We measured and , from which and were extracted. The increased count rate for bound Cy5 is attributed to a reduction of conformational changes leading to higher fluorescence emission.14, 15 For , the fit residuals were lower than . For smaller lag times, shot noise and afterpulsing reduced the signal-to-noise ratio. Figure 4 summarizes the TIR-FCCS measurements on a titration series of double-labeled dsDNA in a mixture of free RhG/ Cy5, such that the total fluorophore concentration was approximately constant. The measured fraction of dsDNA scaled linearly with the mixed fraction from (negative control) to (positive control). The measurements showed a relative scatter of . The measured fraction was at best one-third of the mixed fraction. This can be explained by an excess of molecules with a single green label, possibly further enhanced by photobleaching of the red label during the two-color excitation.15 The overlap of the sampling volumes was estimated to , which is close to the theoretical maximum. To reduce the influence of photobleaching, we used excitation intensities of . The measured CPM were about two times higher than with a confocal epi-illumination and a 1.20 NA water immersion objective at identical excitation intensities.13, 16 The diffusion times of the dsDNA were axially and laterally, respectively. Given a penetration depth of and a waist radius of , the diffusion constant was calculated to , which is about 22% of the estimation for a rodlike molecule with a diameter of and a length of . We attribute this to an increased hydrodynamic drag near the interface and weak unspecific binding, respectively. This study demonstrates for the first time single-molecule FCCS measurements based on an epi-illumination TIR concept. This TIR-FCCS concept offers distinct advantages to confocal FCCS for coincidence assays at solid/liquid surfaces, in particular by virtue of the much higher fluorescence collection efficiencies and the confinement of the excitation field to the surface of interest. AcknowledgmentsWe thank Aram Mooradian of Novalux for the Protera laser and his continous interest in our work, Klaus Weisshart of Carl Zeiss for contributing the cross-correlation sample, and Kai Hassler and Rudolf Rigler for many expert discussions. This work was supported by the Swiss National Science Foundation (SNSF) Grant No. 200021-103333. Jerker Widengren acknowledges support from the Swedish Strategic Research Foundation (SSF, BioX). ReferencesD. Axelrod,
“Total internal reflection fluorescence microscopy in cell biology,”
Traffic (Oxford, U. K.), 2 764
–774
(2001). 1398-9219 Google Scholar
Y. Sako and
T. Uyemura,
“Fluorescence correlation spectroscopy with single-molecule sensitivity on cell and model membranes,”
Cell Struct. Funct, 27 357
–365
(2002). 0386-7196 Google Scholar
Y. Sako and
T. Yanagida,
“Single-molecule visualization in cell biology,”
Nat. Rev. Mol. Cell Biol., 4 SS1
–SS5
(2003). 1471-0072 Google Scholar
D. Toomre and
D. J. Manstein,
“Lighting up the cell surface with evanescent wave microscopy,”
Trends Cell Biol., 11 298
–303
(2001). 0962-8924 Google Scholar
V. Beaumont,
“Visualizing membrane trafficking using total internal reflection fluorescence microscopy,”
Biochem. Soc. Trans., 31 819
–823
(2003). https://doi.org/10.1042/BST0310819 0300-5127 Google Scholar
R. Rigler,
Ü. Mets,
J. Widengren, and
P. Kask,
“Fluorescence correlation spectroscopy with high count rate and low background - Analysis of translational diffusion,”
Eur. Biophys. J., 22 169
–175
(1993). https://doi.org/10.1007/BF00185777 0175-7571 Google Scholar
S. Weiss,
“Fluorescence spectroscopy of single biomolecules,”
Science, 283 1676
–1683
(1999). https://doi.org/10.1126/science.283.5408.1676 0036-8075 Google Scholar
K. Bacia,
I. V. Majoul, and
P. Schwille,
“Probing the endocytic pathway in live cells using dual-color fluorescence cross-correlation analysis,”
Biophys. J., 83 1184
–1193
(2002). 0006-3495 Google Scholar
A. M. Lieto,
R. C. Cush, and
N. L. Thompson,
“Ligand-receptor kinetics measured by total internal reflection with fluorescence correlation spectroscopy,”
Biophys. J., 85 3294
–3302
(2003). 0006-3495 Google Scholar
P. Schwille,
J. Korlach, and
W. W. Webb,
“Fluorescence correlation spectroscopy with single-molecule sensitivity on cell and model membranes,”
Cytometry, 36 176
–182
(1999). https://doi.org/10.1002/(SICI)1097-0320(19990701)36:3<176::AID-CYTO5>3.0.CO;2-F 0196-4763 Google Scholar
R. L. Hansen and
J. M. Harris,
“Measuring reversible adsorption kinetics of small molecules at solid/liquid interfaces by total internal reflection fluorescence correlation spectroscopy,”
Anal. Chem., 70 4247
–4256
(1998). https://doi.org/10.1021/ac980925l 0003-2700 Google Scholar
N. L. Thompson,
T. P. Burghardt, and
D. Axelrod,
“Measuring surface dynamics of biomolecules by total internal-reflection fluorescence with photobleaching recovery or correlation spectroscopy,”
Biophys. J., 33 435
–454
(1981). 0006-3495 Google Scholar
K. Hassler,
M. Leutenegger,
P. Rigler,
R. Rao,
R. Rigler,
M. Gösch, and
T. Lasser,
“Total internal reflection fluorescence correlation spectroscopy (TIR-FCS) with low background and high count-rate per molecule,”
Opt. Express, 13 7415
–7423
(2005). https://doi.org/10.1364/OPEX.13.007415 1094-4087 Google Scholar
J. Widengren and
P. Schwille,
“Characterization of photoinduced isomerization and back-isomerization of the cyanine dye Cy5 by fluorescence correlation spectroscopy,”
J. Phys. Chem. A, 104 6416
–6428
(2000). https://doi.org/10.1021/jp000059s 1089-5639 Google Scholar
C. Eggeling,
J. Widengren,
L. Brand,
J. Schaffer,
S. Felekyan, and
C. A. M. Seidel,
“Analysis of photobleaching in single-molecule multicolor excitation and Förster resonance energy transfer measurements,”
J. Phys. Chem. A, 110 2979
–2995
(2006). 1089-5639 Google Scholar
J. Mertz,
“Radiative absorption, fluorescence, and scattering of a classical dipole near a lossless interface: a unified description,”
J. Opt. Soc. Am. B, 17 1906
–1913
(2000). 0740-3224 Google Scholar
|