1 March 2009 Time-gated flow cytometry: an ultra-high selectivity method to recover ultra-rare-event μ-targets in high-background biosamples
Dayong Jin, James A. Piper, Robert C. Leif, Sean Yang, Belinda C. Ferrari, Jingli Yuan, Guilan Wang, Lidia M. Vallarino, John W. Williams
Author Affiliations +
Abstract
A fundamental problem for rare-event cell analysis is auto-fluorescence from nontarget particles and cells. Time-gated flow cytometry is based on the temporal-domain discrimination of long-lifetime (<1 μs) luminescence-stained cells and can render invisible all nontarget cell and particles. We aim to further evaluate the technique, focusing on detection of ultra-rare-event 5-µm calibration beads in environmental water dirt samples. Europium-labeled 5-µm calibration beads with improved luminescence homogeneity and reduced aggregation were evaluated using the prototype UV LED excited time-gated luminescence (TGL) flow cytometer (FCM). A BD FACSAria flow cytometer was used to sort accurately a very low number of beads (<100 events), which were then spiked into concentrated samples of environmental water. The use of europium-labeled beads permitted the demonstration of specific detection rates of 100%±30% and 91%±3% with 10 and 100 target beads, respectively, that were mixed with over one million nontarget autofluorescent background particles. Under the same conditions, a conventional FCM was unable to recover rare-event fluorescein isothiocyanate (FITC) calibration beads. Preliminary results on Giardia detection are also reported. We have demonstrated the scientific value of lanthanide-complex biolabels in flow cytometry. This approach may augment the current method that uses multifluorescence-channel flow cytometry gating.

1.

Introduction

Specific detection of rare-event cells in a rapid and cost-effective fashion remains a challenging issue in biomedicine1, 2, 3, 4, 5 and environmental analysis.6, 7, 8 A rare-event is broadly defined as one that occurs at a frequency of less than 1 in 10,000. Many biological and clinical procedures require accurate detection of rare target cells at frequencies in the range 1 in 1,000,000 or below.3 For example, residual cancer cells in bone marrow or peripheral blood stem cells preparations used in bone marrow transplantations are a major concern,9 and it is vital to be able to remove such cells prior to the procedure.10 This requires a detection level of one residual cancer cell per 107 of bone marrow or peripheral blood stem cells. Fetal cells present in maternal blood during pregnancy are an ideal source of genetic material for noninvasive prenatal diagnosis; however, the target fetal nucleated red blood cells (NRBCs) need to be detected against the maternal cells at extremely low frequencies of 1 in 107 to 109 (Refs. 3, 11, 12). In water safety inspection, due to the very small number of organisms needed to cause infection, the methods of analysis must be sufficiently sensitive to detect a single microorganism (e.g., Cryptosporidium parvum and Giardia lamblia) in as many as 10 liters of water containing millions of nontarget microorganisms and particles.13

Flow cytometry is the only technique capable of rapidly identifying rare cells from a large sample. It relies on scattering or fluorescence measurements that are made while the cells or particles pass, usually in single file, through a capillary flow cell. Commercial conventional flow cytometry instruments that are equipped with multiple spectral channels are capable of real-time rapid analysis with rates reaching 100,000cellss (Ref. 6). However, detecting such rare events poses significant challenges in terms of accuracy and comes at an unacceptable cost.10, 14 The fundamental challenge is autofluorescence noise from nontarget particle/cells in the complex biological samples.13, 15, 16, 17

We have previously reported a practical flow cytometry method based on the time-gated detection of long-lifetime luminescence-labeled targets.18, 19 The use of time-gated detection of target events in temporal, spectral, and spatial 3-D domains excluded those background particles or cells that otherwise would have been included because their ranges (gates) of light scattering and autofluorescence matched that of the targets. This time-gated detection method enhanced the likelihood of detection of micron-size targets present at a low concentration. This technique showed great potential for the detection of trace amounts of micron-size targets by flow cytometry. However, these studies were limited by the lack of long-luminescent-lifetime calibration beads, which are required for a number of essential experimental activities, such as counting particles, obtaining a histogram of the distribution of particle luminescent intensities, and determining the frequency of rare-event recovery. In this work, a prototype lot of 5-μm europium-labeled beads with long-lived luminescence (340μs) was utilized20 to evaluate the performance of time-gated luminescence (TGL) flow cytometry, with specific focus on the capability of rare-event recovery in environmental samples.

2.

Materials and Methods

2.1.

Time-Gated Luminescence Flow Cytometer

The details of both the concepts and prototype operation of the TGL flow cytometer (FCM) were reported previously.18, 19 Briefly, Fig. 1 shows the basic geometric layout for the current development of the TGL FCM. The targets of interest are labeled with a long-lived luminescent probe, typically a lanthanide complex, and then are excited with a pulsed LED (pulse duration up to 100μs ) as they flow in single-cell profile through the hydrodynamically focused up-flowing laminar stream [Fig. 1a]. Once the excitation pulse has extinguished, the autofluorescence fades rapidly (within 0.1μs ). During this time, the luminescence from the targets loses little of its original intensity and continues as the targets progress downstream from the excitation spot. The implementation of periodically pulsed illumination and of time-delayed gated detection has been carefully designed to achieve 100% spatial detection of TGL particles [Fig. 1b, Table 1 ], taking into account of the flow speed (3.2ms) , illumination aperture (530μm) , detection aperture (8mm) , luminescence-label lifetime (340μs) , pulsed illumination (100μs) , and gated-detection-timing sequences.

Fig. 1

Schematic geometric layout of TGL FCM. (a) UV LED excitation at flow cell: In the sample flow stream, the solid round (red) particles represent target europium particles/cells, and the other larger round (pattern), oval, and triangle particles represent nontarget autofluorescent particles/cells. The smallest illumination spot size from focused UV LED light was 0.53mm along the flow stream, more than 20 times larger than in conventional flow cytometers that use laser illumination, thus covering many cells instead of a single cell per illumination region. (b). Instead of continuous illumination and detection as in conventional flow cytometers, pulsed excitation and time-gated detection sequence in antiphase are applied to monitor flow sections. This allows both the position and the size of the detection spot to be related to the flow rate and TGL repetition rate in order to discriminate long-lifetime luminescence-labeled cells. The image at the flow cell was magnified (15×) via the signal collection objective optics after projection onto the image plane, which is inverted by the objective. (c) UV exposure compensation in time (upper axis) and distance (lower axis): A cell entering into the excitation spot during the initial UV pulse (100μs) being less exposed can be compensated by the secondary UV pulse before exiting the UV illumination region. (Color online only.)

024023_1_044902jbo1.jpg

Table 1

Operational parameters of the TGL flow cytometer.

UV LED pulsedexcitation(NCCU033A)Output wavelength 365nm
Current level 1.0A
Peak intensity at flow cell 15.0mW
CPMT gateddetection(MH1372)Quantum yield at 610nm (europium)0.09
Gain 2×106
Preamplifier gain ( 400-kHz bandwidth) 105VA
OpticsExcitation filter (SCOTTS UG5 glass) <400-nm pass
Dichroic beamsplitter (Zeiss, FT395)Reflect UV; pass visible
Emission filter (New Focus, 9514B) 610-nm to 630-nm pass
Excitation aperture 9mm diameter
Condenser lens 40mm diameter; f=2mm
Emission aperture 9mm long; 1.5mm wide
Flow cell objective magnification ×15
Objective numerical aperture (NA)1.17 (oil-immersion)
Fluidics: vacuumsuctionFlow velocity 3.2ms1
Flow cell dimension 430μm×180μm
Timing sequenceTGL cycle rate 6.45kHz
Excitation pulse duration 100μs
Gate delay time 10μs
Gated detection duration 40μs
Geometricalparameters atflow cellExcitation spot (530μm)2
One excitation profile distance 850μm
Flow section distance 496μm
Detection region distance 528μm
Spatially delayed distance 350μm
Geometricalparameters atimage sideDetection window distance 8mm
Detection window delay distance 5mm
Image aperture width 1.5mm

According to the flow-section model of TGL flow cytometry,18 the continuous flow stream was conceived as divided into continuously adjacent sections, so that ideally each TGL cycle (consisting of pulsed excitation, time delay, and time-gated detection) can be applied to each section for examining target events within, in order to achieve 100% detection efficiency of sample targets. In this work, a cell takes 166μs to travel across a UV illumination region of 530μm (see Table 1); the long pulsed UV-excitation duration (100μs) resulted in the different position-dependent levels for UV exposure (cells receiving different percentage of pulsed UV exposure) along the flow stream.18 Fortunately, due to the rapid TGL repetition rate (6.45kHz) , the cells being less exposed from the initial pulse can be compensated by the secondary UV pulse before exiting the UV illumination region, so that the whole flow stream is covered by UV illumination without “dead time.” Figure 1c illustrates such excitation compensation for a cell entering into the excitation spot delayed by 50μs since switching-on of a UV pulse: for the first 50μs , the cell was outside the UV illumination spot, and then was illuminated only by the second-half UV pulse for 50μs after traveling into the excitation spot. During the UV-off period of 55μs [from 100μsto155μs , upper axis Fig. 1c], the cell traveled for 176μm [from 160μmto336μm , under axis Fig. 1c] within the excitation spot. Before exiting the UV excitation region, the second UV pulse will illuminate the cell for another 61μs [from 155μsto216μs , upper axis Fig. 1c]. In total, the cell will accumulate UV exposure for 111μs . At the TGL repetition rate of 6.45kHz and flow velocity of 3.2ms , all the cells/particles can be excited twice when traveling through the 530μm UV illumination region. Different position-dependent levels for UV exposure for analytes still exist because luminescence is emitted between the two pulses and cells/particles entering the illumination region at different positions (times in the cycle) receive different proportions of UV from the two pulses.

A new-generation UV LED (maximum CW output power of 250mW at 365nm , NCCU033A; Nichia Corp., Japan) generated 15-mW peak intensity at the flow cell, so the calculated UV excitation efficiency from the full 100-μs pulse was improved18 from 3.53% to the current 7.56%. At a TGL repetition rate of 6.45kHz , each flow section covers 496μm , so the detection region distance is set for 528μm by detection aperture of 8mm on the image plane [Fig. 1b]. Thus, each TGL section is monitored for 528-μm detection window, within which any time-gated signal pulses with duration of more than 10μs (filtered by software) can be counted as valid events. In principle, since the particles traveling at flow velocity of 3.2ms , the coincidence for the same particle being counted twice in any two adjacent TGL cycles (in the boundary of two adjacent flow sections) is extremely small. In practice, however, a filtering function was programmed in the software to count only events appearing > 300μs away from each other; hence, no long-lifetime luminescent target cells/particles are being counted twice.

Figure 2 shows the optical layout of the UV LED excited TGL FCM. The high-intensity UV LED is mounted on an aluminum plate, which provides improved heat dissipation. The light emitted at a large solid angle (> 45deg) is collected by a high numerical aperture (NA) condenser lens, with the image of the LED projected on the excitation aperture plane, which is positioned directly after a UV excitation filter. The approximately parallel beam is then reflected by the dichroic mirror and redirected into the flow cell objective. Table 1 lists the key parameters required for the current version of the TGL FCM to achieve optimum conditions. The TGL repetition rate was 6.45kHz with excitation pulses of 100μs , followed by a 10-μs time-resolving period and a 45-μs gated-detection period. The average flow velocity was calibrated as to 3.2ms (sample flow rate: 100μLmin ). The UV LED was focused to generate a (530-μm)2 illumination spot with 15mW peak power on the sample stream, and the channel photomultiplier tube (CPMT) detector (MH 1372; PerkinElmer Optoelectronics, Germany) operated at a photoelectric gain of 2×106 . The detected anode current from the CPMT was converted to voltage signal by a current-voltage preamplifier (model DHPCA-200, FEMTO, Germany; 400-khz bandwidth; 105VA ). This signal was subsequently digitized by a 1.25-MHz data acquisition card (DAQ; NI PCI-6251, National Instruments Australia Corp., New South Wales, Australia). The memory buffers provided by both DAQ card and personal PC are sufficient for capturing the signal sequence acquired every second, continuously allowing constant real-time data analyzing and presentation in the Labview program. The real-time TGL event counting was programmed with Labview 7.1 Express (National Instruments Australia Corp, New South Wales, Australia).

Fig. 2

Layout of the UV LED excitation system that employs epi-illumination optics to excite the TGL FCM. (Color online only.)

024023_1_044902jbo2.jpg

2.2.

Europium Calibration Beads

As described in the companion publication,20 the prototype Newport Instruments Fire Red™ beads were aqueous suspensions of europium-complex-loaded polymer microspheres (beads), which showed low aggregation [see Fig. 3a of Ref. 20]. The 5.0-μm europium microspheres had a coefficient of variation (CV) of 7.0% [see Fig. 3b of Ref. 20] when measured with a flow cytometer (see Fig. 1 of Ref. 20), which was conventionally gated by wide-angle light-scatter (see Fig. 2 of Ref. 20). The CV of 7.0% establishes that the beads were reasonably uniform in size. The beads contained Eu3+ coordination complexes that have an excitation maximum at approximately 370nm and a very narrow emission near 620nm , with a luminescence lifetime of 320±30μs for beads suspended in water (see Fig. 7 of Ref. 20). Before each flow cytometry measurement, the sample was ultrasonicated for 4min to remove particle aggregation.

Fig. 3

Real-time counting of europium calibration beads by the UV LED–excited TGL FCM. (a) A 4-ms section of the TGL FCM pulse train during a quiescent period when no particles were present. The minima and maxima of the peaks indicate that the CPMT was in the off and on modes, respectively. The TGL flow-counting background level was less than 0.3V when there was no TGL event present during the time-gated detection periods. (b) TGL event temporal location. Each cross (+) represents the signal pulse height (V) and arrival time (s) when a TGL event passed the interrogation region of the TGL flow cytometer. (c) Histogram of the signal pulses above 0.5V . The counts shown on the left below 3V result from particles that were in the illumination zone during part of the dark period between LED pulses and thus were exposed to UV pulses for less than 100μs .

024023_1_044902jbo3.jpg

2.3.

Flow Cytometric Cell Sorting to Produce Calibration Beads

Spiked samples of both europium-labeled and FITC-labeled beads, consisting of 10, 100, or 1000 beads, were prepared for recovery determinations. Briefly, calibration spikes were prepared using the BD FACSAria flow cytometer (BD Biosciences, Macquarie Research Park, Sydney, Australia) in single-cell mode. A gate (R1) was defined around the single bead populations within a bivariate dot plot of side scatter scatter (SSC) versus forward scatter (FSC) and events detected in R1 were sorted directly into microfuge or FACS tubes for analysis. Sheath fluid consisted of undiluted Osmosol (Lab Aids Pty. Ltd. Narrabeen, New South Wales, Australia). Each spike calibration set was produced in triplicate for each analysis. Since it is easy to lose subpopulations of cells or particles against the walls of tubes, particularly given the extremely small number (10 or 100) of beads used in this work, the calibration beads were sorted directly into the final test tubes (microfuge tubes were used to contain the europium beads for TGL flow cytometer; FACS tubes were used to contain the FITC-labeled beads for the conventional flow cytometer) or onto the microscopy slides for microscopy confirmation of beads (see the following subsection) without sample transfer steps. In particular, for detection beads from environmental water mud concentrates (Secs. 3.3, 3.4), the prepared water mud samples (Sec. 2.6) were added into the microfuge tubes or FACS tubes containing the sorted rare-event beads.

2.4.

Microscopy Confirmation of Bead Spikes

The numbers of beads present in the 100- and 10-bead calibration sets were determined by epifluorescence microscopy using an Axioskop 2 microscope (Carl Zeiss, Sydney, Australia) with appropriate filters for the examination of FITC and europium emissions [4'-6–diamidino-2–phenylindole (DAPI) excitation filter and dichroic mirror]. For the europium-bead preparations, the spike beads sets were 97±1 and 10.3±0.7 for the 100- and 10-bead sample sets, respectively. For the FITC bead preparations, the spike beads sets were 101±1 and 9±1 for the 100- and 10-bead sample sets, respectively.

2.5.

Giardia Lamblia Cyst Labelling

As previously reported,21 an indirect-labeling method was employed to maximize luminescence. Streptavidin (SA) was first conjugated to bovine serum albumin (BSA), and then the SA-BSA conjugate was labeled with a highly luminescent Eu3+ complex, 4,4 -bis ( 1,1,1,2,2,3,3 -heptafluoro- 4, 6 -hexanedion- 6 -yl)-chlorosulfo-o-terphenyl- Eu3+ (BHHCT-Eu3+) , to produce the europium complex–labeled SA, SA(BSA)0.9(BHHCT-Eu3+)46 . As a result, an average of 46 BHHCT-Eu3+ molecules were conjugated to each SA molecule.21 The original 7-μL solution containing 1000 Giardia lamblia cysts (Biotech Frontiers Pty. Ltd., Sydney, Australia) was incubated for 17h at room temperature (20°C) with 10μL of mouse monoclonal anti-Giardia IgG antibody G203 (cyst-wall specific, 40μgmL ; Biotech Frontiers Pty. Ltd., Sydney, Australia). The sample was then diluted with 300μL of monoclonal antibody (mAb) buffer, which contains 2mM tetrasodium pyrophosphate (TSPP); (VWR, http://www.vwrsp.com/); 2% w/v BSA (Sigma-Aldrich), and 0.05% v/v Tween 80 (Sigma-Aldrich); the pH of the buffer was adjusted to 8.0 . The sample was subsequently washed twice with the mAb buffer by centrifugation at 12,000RPM for 5min , and the supernatant solution was removed. The sample was then incubated at room temperature for 2h with 30μL of a 1:150 dilution of Goat Anti-Mouse Light Chain Specific Biotin Conjugated IgG antibody (Catalogue Number AP200B, Millipore; Chemicon International, Millipore Bioscience Division, Australia). The sample was then washed three times with the mAb buffer. Subsequently, 20μL of the BHHCT-Eu3+-SA conjugate solution (40μgml) was added and allowed to react for 1h . This was followed by three washes with 400μL of the mAb buffer, after which 500μL of mAb buffer was added to prepare the Giardia samples for detection with the TGL FCM. The aim of this work was to evaluate, by the use of a spike-and-recovery method, the current status of the prototype TGL FCM for the detection of microbial cells against an autofluorescent background. No attempt was made to study real-time flow cytometry labeling protocols that would minimize the effect of nonspecific binding to other particles.

2.6.

Environmental Water Sample Preparation

A large volume of backwash filtered water from a drinking water treatment plant was prepared for use as a quality control sample. Additionally, a 10-L water sample was processed with a calcium carbonate flocculation modified filter to produce a set of 100-mL samples.22 Following concentration, the water samples were centrifuged for 10min at 3,500RPM , and the supernatant was discarded. Prior to use, the concentrates were centrifuged again for 10min at 13,000RPM , and the supernatant was discarded. The samples were then resuspended to their initial volume (100mL) in the mAb buffer and filtered, using a syringe, through a 38-μm stainless steel mesh filter (Metal Mesh Pty., Ltd., Sydney) housed within in a Swinnex filter unit (Millipore, North Ryde, Australia). The samples were forced through the filter to remove large aggregates.

3.

Results

3.1.

Real-Time Counting Histogram

Figure 3a shows one typical 4-millisecond section (26 TGL cycles) of the TGL FCM pulse train of a blank sample. It was known from previous work19 that, under the same operational settings, the real-time observation of signal trains for either pure samples or samples with a large amount of autofluorescence backgrounds showed no difference in terms of background level during the time-gated detection phase. In the experiment illustrated in Figure 3, a sample consisting only of europium calibration beads was tested. Figure 3a shows that when no TGL event was present in the time-gated detection periods (corresponding to flow sections that did not contain target beads), the TGL flow counting background level was less than 0.3V . With the UV LED switched off, only the CPMT dark count (electronic noise) was occasionally observed during the time-gated detection periods (data not shown). When the UV LED injection current was increased, the background level also increased, until it become independent of the LED injection current in the range between 0.2A and 1A . As previously noted,23, 24 most of the background detected in the time-gated phase was the result of the UV LED native emissions, which were detected by the CPMT detector in spite of it being shielded by a 610-nmto640-nm bandpass filter. In order to generate reliable TGL detection data, a pulse detection threshold of 0.5V was set for all work reported in this paper.

A concentration of approximately 30,000 europium-loaded beads/mL was sufficient to generate statistically valid histogram data. The arrival rate of target beads was calculated from the sample flow rate (100μLmin) to be approximately 50beadss . According to the concept paper,18 at this concentration, the coincidence for two particles arriving in the same flow section is calculated as 0.003%. In practice, there were particles aggregating together (see the second histogram peak in Fig. 3b of the companion paper, Ref. 20). Figures 3b and 3c show the real-time counting results for a one-minute period, with 3,265 TGL events counted. Figure 3b shows the TGL event temporal locations; each cross (+) marker represents the signal pulse height ( y axis; volts) and arrival time ( x axis; μs ) when a TGL event passed the detection region of the TGL FCM. Figure 3c shows the histogram for the signal pulses above 0.5V . Thresholding resulted in the clear area at the bottom of Fig. 3b.

The use of the new europium calibration beads and of the new UV LED excitation ( 15-mW peak power at the flow cell, compared to 7mW in the previously reported work19) allowed most of the TGL pulses to be clearly resolved from the background level (0.3V) , and the highest TGL signal pulse was 8.2V , corresponding to a maximum 27 to 1 signal-to-background ratio. In contrast to our previously reported results,19 the use of new calibration beads provided minimum aggregation after ultrasonication. If the CV is measured on the high amplitude side, which consists of data obtained from particles with a complete exposure to the UV excitation light, this CV falls to approximately 8%. This is in good agreement with the 7% obtained with the light-scatter gated luminescence analysis described in the companion article.20 These improvements resulted in experimental data that fit more closely to the predicted theoretical concepts.18 For example, 81% of the TGL target beads registered into higher channels (> 2V) and the peak channel (mode) was located at approximately 6.3V , with 62% of the events being located in the region between 4.5 and 8.2V . The different periods of exposure of the beads to the exciting light and the heterogeneity of the beads are the two factors most likely to contribute to the shape of the profile of the TGL flow counting histogram. Concerning the exposure of the beads, the peak population at higher channels (> 4.5V) reflected the number of europium calibration beads that were exposed to maximum UV LED excitation, 100μs . In turn, the long “tail” toward lower channels was due to the partial excitation of the europium beads by UV LED pulses, which resulted either from some beads entering the excitation region shortly after the 100-μs UV pulse initiation or from some beads exiting the excitation region shortly before the 100-μs UV pulse completion.18 The heterogeneity of the emissions of the beads could also result from differences in the volume and/or loading of the calibration beads (This was discussed in detail and shown in Fig. 3b of the companion article, Ref. 20). Since the histogram figure reported in Fig. 3b of the companion article was generated by the scattering-triggered TGL flow cytometer, each particle was illuminated by the same level of UV excitation, thus the histogram figure does not contain the long “tail” toward lower channels, as shown in the histogram Fig. 3c generated by the flow-section TGL FCM. This comparison suggests that the different periods of exposure to the exciting light primarily contributed the long “tail” towards lower channels. The fraction of events that were undetected in this work is expected to be reduced by planned improvements in the instrumentation, as described in Sec. 4.

3.2.

Detection Rates for Rare-Event Counting of Beads

Three samples, each containing 100 sorted beads, were examined. Figure 4 presents one of the typical results (sample no. 1). Since some target beads could adhere to both the sample tube and to the injection needle of the flow cell, a number of dilution-washing-mixing steps was performed during the rare-event recovery experiments. The original samples of sorted beads were diluted initially into 800μL with the mAb buffer, which resulted in most of the beads being detected during the first 8min [Fig. 4a]. Without pausing and while continuing the counting by the TGL FCM, the sample tube was washed by the addition of 800μL of mAb buffer and vortex mixing. This was followed by a further wash with 400μL of water. Counting was terminated 30min (1800s) after initiation. Figure 4b is a histogram of the 87 events counted. Due to the low number of target beads, the histogram profile was not sufficiently clear to show consistency with the counting result reported in the preceding subsection. However, the peak channel (mode) at about 6.4V was still quite clearly populated. The visual count obtained by microscopy for the nominal 100 -bead samples was 97±1 . The number of beads detected by the TGL FCM in the three nominal 100 -bead samples was 87, 91, and 85 (87.7 average; 3.5% CV), which is 91% ±3% of the number counted by microscopy.

Fig. 4

Detection of the microscopically confirmed 97±1 beads (sample no. 1) with the TGL FCM, resulting in 87 beads counted. (a) Temporal location of TGL events. (b) TGL FCM counting histogram.

024023_1_044902jbo4.jpg

Six samples containing 10 sorted beads were examined with the TGL FCM. Figure 5 presents one of the typical results (sample no. 3). The original sorted bead samples (100μL) were tested directly without dilution and resulted in the majority of the beads (6 counts for sample nos. 3 and 9 counts in some other samples) being detected in the first 1min (data not shown). Without pausing the TGL FCM and while continuing the counting, the sample tube was washed by the addition of 600μL of water and vortex mixing; the counting was terminated 8min after initiation. Figure 5, is a histogram of the 11 events counted. The number of beads detected by the TGL FCM in the six nominal 10 -bead samples was 12, 10, 10, 11, 9, and 11 (10.5 average; 10.0% CV), which is 100% ±20% of the 10±1 counted by microscopy.

Fig. 5

Detection of the microscopically confirmed 10±1 beads (sample no. 3) with the TGL FCM, resulting in 10 beads counted.

024023_1_044902jbo5.jpg

3.3.

TGL Detection of Rare-Event Beads from Environmental Water Samples

Six 10-bead samples were spiked into 200μL of water-contaminate concentrates (mud) to check the rare-event counting capability offered by the TGL flow cytometry technique with high-background samples. Figure 6a shows that the water concentrates did not increase the background level during the time-gated detection phase; Fig. 6b presents the histogram of one of the typical results (sample no. 4). Samples that contained both beads and mud background particles (300μL) were tested directly without dilution. This resulted in 9 counts in the first 3min (event temporal location not shown). Without pausing the TGL FCM and while continuing the counting, the sample tube was then washed by the addition of 400μL of water and vortex mixing; counting was terminated 8min (480s) after initiation. Figure 6b is a histogram of the 11 events counted. The number of beads detected by the TGL FCM in the six nominal 10 -bead samples was 13, 11, 9, 11, 9, and 8 (10.2 average; 18.0% CV), which is 100%±30% of that measured by microscopy. There were no counts in the muddy water background samples.

Fig. 6

Detection of the microscopically confirmed 10±1 beads (sample no. 4) from muddy water with the TGL FCM, resulting in 11 beads counted. (a) The TGL flow-counting background level when no TGL event was present in the time-gated detection periods. (b) TGL FCM counting histogram.

024023_1_044902jbo6.jpg

3.4.

Conventional Flow Cytometric Detection of Rare-Event Beads from Environmental Water Samples

Figure 7 shows a comparison of the recovery data obtained with FITC beads from a pure sample versus a sample in muddy water. These measurements were obtained with a conventional BD FACSCalibur flow cytometer with two-channel regional gating (R1) of the side-scatter (SSC) and green fluorescence. The test procedure was the same as in the TGL FCM evaluations that followed the FACS sorting-microscopic confirmation-spike-flow cytometry evaluation steps. With samples of pure beads, the FACSCalibur detected an acceptable 82% (data not shown) and 100% [Fig. 7a] of the 101±1 and 9±1 FITC beads counted by microscopy. However, with environmental samples containing nontarget (background) particles, the two-channel-based conventional flow cytometry technology lost its discrimination power [Figs. 7b and 7c]. Over 106 nontarget autofluorescent particles were detected above the green fluorescence trigger threshold [already increased from 180 units to 450 units, in contrast to the pure bead counting shown in Fig. 7a]. The result was 507 and 314 FCM counts in the gated target region (R1) for samples that contained only 100 beads and 10 beads, respectively. We demonstrated that FACSCalibur failed to detect rare-event FITC beads in a highly autofluorescence sample, which is a typical example of the challenges encountered in the detection of rapid rare-event microbial cells in environmental samples that contain a large number of nonspecific, intrinsically autofluorescent background cells or particles.

Fig. 7

Detection of FITC beads from muddy water, using a conventional BD FACSCalibur flow cytometer with two-channel regional gating (R1) of the side-scattering (SSC) and green fluorescence (FL1). The counts within the R1 gate are shown in red and outlined in black. (a) Shows that in the case of pure-beads sample counting, very good detection rates (100%) could be achieved for 10 -bead samples. (b) and (c) Represent two typical results for test samples containing a large number of nontarget particles in muddy water from environmental sources. Region R1 is outlined in white. In these tests, the two-channel-based conventional flow cytometry method lost its discrimination power, as approximately 407 counts and 304 nonspecific counts in the targeted region masked the 100 and 10 target FITC beads, respectively, present in the samples. (Color online only.)

024023_1_044902jbo7.jpg

A number of issues need to be addressed here: The detection capability by conventional flow cytometry may be enhanced in the spectral domain by using dual-color coded beads or long Stokes-shift dyes such as PE-Cy5 or PE-Cy7 (or the Alexa analog) or Quantum Dots coded beads (in fact, the best choice would be to use green or red Qdots™ with UV or violet excitation) to avoid greenish autofluorescent backgrounds. Previous research25 demonstrated the presence of algal cells, which autofluoresce in the red. The presence of photosynthetic organisms is consisted with their findings that FITC excited at 488nm was the most suitable label for oocysts (detection) in untreated water from 10liter concentrate, and the fluorochromes CY3, phycoerythin (PE), and tetramethylrhodamine B thioisocyanate (TRITC) excited at 542nm were the most suitable labels for oocysts in drinking water, where presumably the algal material had been removed. The differences in lifetimes between lanthanide complexes (hundreds of μs ) and conventional fluorophores (20 or less nanoseconds) precluded any reasonable comparison of the two types of labels with the same instrument. At best, only approximately 3% of the europium emission would occur in the detection zone of a conventional flow cytometer. Excitation of the europium-labeled beads would require equipping the FACSCalibur with a 325-nm or 351-nm CW ion laser. Conversely, because of the short lifetime of a quantum dot or a conventional fluorophore, the emissions of these labels would disappear during the 10μs time-resolving period before the detector of the TGL FCM becomes fully activated.

3.5.

Giardia Lamblia Cyst Detection by TGL FCM

The original 100-μL samples containing approximately 200 europium-complex-labeled Giardia cysts in muddy water were tested with the current version of the TGL FCM, and 27 events were detected with an approximate detection rate of 13%. (Figure 8 is an example of the preliminary data on Giardia cyst detection.) The current prototype failed to detect the majority of the targeted Giardia cysts, and all detected events registered into lower channels (<2volts) . The main reason for this result is due to the low excitation efficiency (currently only 8% ) from UV LED; we expect strong excitation from coherent light sources or future development of high power UV LEDs. Second, according to our previous experience with TGL microscopy bio-imaging using europium complex as labels,24, 26 the luminescence intensity of europium complex is increased in aqueous solution by the use of a luminescence enhancement micellular solution.27, 28 Hence, at the next stage of this project, we are going to optimize the europium-labeling protocol including both the antibody-antigen biotin-SA labeling efficiency and the luminescence enhancement buffer for flow cytometry application. However, this result still shows the potential of the TGL flow cytometry technique for use in microbial cell detection, since major improvements in the instrumentation and chemistries can be achieved, as described in the next section.

Fig. 8

Preliminary results on the detection of Giardia cysts from muddy water using the TGL FCM.

024023_1_044902jbo8.jpg

4.

Discussion

Previous reports1, 6, 10, 16 have demonstrated the challenges encountered by the available flow cytometry technologies for the detection of ultra-rare-events ( μ -beads) when a large number of nontarget background particles/cells is also present. In contrast, the new technique for flow cytometry TGL detection reported in this paper has proved successful for the accurate detection of as few as 10 target μ -beads within a population of more than 1,000,000 background particles. In contrast to the conventional multispectrum flow cytometry, the TGL flow cytometry technique here presented a distinctive scenario of extracting temporal (target signal lifetime) fingerprints from target cells. Although in the family of temporal flow cytometry techniques, phase-sensitive flow cytometry29, 30, 31, 32, 33, 34 has been reported, discrimination of individual cells in flow cytometry using the lifetime difference in the nanosecond range could be decreased because of a number of factors that alter the lifetime values, e.g., staining concentration and analysis condition26 leading to a complex and expensive setup. Lifetime difference permits the detection of the relative amounts of fluorophores with overlapping spectra present. Careful studies on the excitation and emission spectra of a large number of mud samples would be required to determine whether a fluorophore exists that can be sufficiently separated from the autofluorescence of all of the muds to permit background-free detection. The magnitude of the cost of these studies probably is prohibitive. Since time-gated detection of the exceptional long lifetime and large Stokes shift of europium dye probes renders invisible all the background autofluorescence from the particle/cells, ultra-rare-event detection is possible. This has resulted in a practical simple mechanical-optical setup. In conventional flow cytometry, the frequency of target cells per total cells/particles is an essential parameter to calculate rare-cell sampling statistics27; while in the flow-section TGL flow cytometry, thanks to the independence of the number of nontarget particles/cells, the frequency of target cells per TGL cycles/sections is the equivalent parameter in order to calculate the rare-cell sampling statistics. For example, when detecting 100 target beads (Sec. 3.2) in 1800s (30min) with 11,610,000 decisions (TGL cycles/sections) made, the TGL decision rarity is calculated as one in 116,100, and a theoretical accuracy (assurance) limit can be calculated as 95% (as described in Ref. 34). Our experimental result of 91%±3% closely agrees with the theoretical limit. For the case of 10-bead recovery, the potential error in predicting the correct rare-event frequency gets increasingly larger as the total number of measured rare events becomes very small, as described by Rosenblatt,27 which also well explained our results of 100%±30% . In this work, the development and evaluation of an LED-excited TGL FCM for rare-event counting were based on the availability of low-cost UV LEDs and of new europium-labeled calibration beads, as well as on the sorting capacity of the BD FACSAria flow cytometer, which provided accurate and reliable counts of rare-events in test samples.

The ability of the LED TGL FCM to detect real biological agents can be further increased by improvements in both the instrument design and the detection chemistries. Continued improvements in excitation energy are expected because of the recent, rapid progress in the commercial development of UV LEDs (led by Nichia, Japan), which resulted in both shorter excitation wavelengths and higher peak powers. For example, LEDs with the following properties have been reported recently: 375-nm LED in 2000; 365-nm , 2-mW LED in 2002; 365-nm , 100-mW LED in 2005; and 365-nm , 250-mW in 2007. The cost of UV lasers, such as diode-pumped solid-state (DPSS) Q-switched or diode-based UV lasers, has also decreased sufficiently to make their use commercially feasible.

Improvement in the optics should increase the optical efficiency, which with the currently available UV LED TGL FCM was only 8% (Ref. 18). The precision of the measurements could be increased significantly by employing an emission detector with a faster turn-on and turn-off, which would permit an increase in the number of excitation-detection cycles. This would lessen the effect of partial illumination of the cells or particles; it would also allow a greater portion of the emitted light to be detected.

Improvements in the detection chemistries will include the use of the europium Quantum Dye, in association with the mononegative anion of thenoyltrifluoroacetone (TTFA) as enhancer.26, 28 This complex has a maximum excitation wavelength that coincides with the emission (365nm) of the 250-mW UV LED; whereas the optimum excitation wavelength for the europium chelates used in this study is 336nm , which does not match the LED emission maximum. The presence of an excess of the Gd(TTFA)3 complex in a micellar solution has been shown to further increase the luminescence of the europium Quantum Dye.35, 36, 37 Polymer carrier technology,38, 39 silica nano-encapsulation technology,40 and the smaller functionalized beads described in the companion publication20 could also be applied independently to increase the luminescence.

5.

Conclusions

Although lanthanide-complex biolabels have been very popular for spectroscopy, microscopy-type bioassays, and bioimaging, their utility for flow cytometry had been previously questioned by Leif 26, 41 It has now been proven that these long-lifetime biolabels can be very useful for rapid cell detection when used with TGL flow cytometry–type instrumentation.

The conventional approach for achieving a highly selective discrimination of target cells in flow cytometry has focused on the use of multicolor biolabels and of multifluorescence-channel gating methods. This work has now demonstrated that rare-event target cells can be discriminated effectively from background by a time-gated method based on long-lifetime luminescent biolabels.

The instrumentation for TGL flow cytometry has the advantage that it allows the use of inexpensive, incoherent solid-state diode sources, thus reducing both the cost and the size of flow cytometers. The success of the tests reported in this paper will lead to further development and applications of the TGL flow cytometer technique in other fields.

Acknowledgments

The authors acknowledge the Australian ARC/NHMRC Fluorescence Applications in Biotechnology and Life Sciences (FABLS) network for a seeding project fund to support emerging new technology, the International Society for Analytical Cytology (ISAC) scholar program, the National Natural Science Foundation of China (No. 20575069), Newport Instruments’ Internal Development Funds, and Lidia Vallarino’s Gift Fund. Robert C. Leif and Sean Yang are employees of Newport Instruments, which is the supplier of the Fire-Red beads.

References

1. 

A. D. Donnenberg and V. S. Donnenberg, “Rare-event analysis in flow cytometry,” Clin. Lab Med., 27 (3), 627 –652 (2007). https://doi.org/10.1016/j.cll.2007.05.013 0272-2712 Google Scholar

2. 

A. L. Allan, S. A. Vantyghem, A. B. Tuck, A. F. Chambers, I. H. Chin-Yee, and M. Keeney, “Detection and quantification of circulating tumor cells in mouse models of human breast cancer using immunomagnetic enrichment and multiparameter flow cytometry,” Cytometry, Part A, 65A (1), 4 –14 (2005). https://doi.org/10.1002/cyto.a.20132 1552-4922 Google Scholar

3. 

S. Bajaj, J. B. Welsh, R. C. Leif, and J. H. Price, “Ultra-rare-event detection performance of a custom scanning cytometer on a model preparation of fetal nRBCs,” Cytometry, 39 (4), 285 –294 (2000). https://doi.org/10.1002/(SICI)1097-0320(20000401)39:4<285::AID-CYTO6>3.0.CO;2-2 0196-4763 Google Scholar

4. 

E. A. Jones, A. English, S. E. Kinsey, L. Straszynski, P. Emery, F. Ponchel, and D. McGonagle, “Optimization of a flow cytometry-based protocol for detection and phenotypic characterization of multipotent mesenchymal stromal cells from human bone marrow,” Cytometry, Part B, 70B (6), 391 –399 (2006). https://doi.org/10.1002/cyto.b.20118 1552-4949 Google Scholar

5. 

K. Yamaguchi, K. Itoh, T. Masuda, A. Umemura, C. Baum, Y. Itoh, T. Okanoue, and J. Fujita, “In vivo selection of transduced hematopoietic stem cells and little evidence of their conversion into hepatocytes in vivo,” J. Hepatol, 45 (5), 681 –687 (2006). https://doi.org/10.1016/j.jhep.2006.04.012 0168-8278 Google Scholar

6. 

B. C. Ferrari and D. Veal, “Analysis-only detection of Giardia by combining immunomagnetic separation and two-color flow cytometry,” Cytometry, Part A, 51A (2), 79 –86 (2003). https://doi.org/10.1002/cyto.a.10009 1552-4922 Google Scholar

7. 

D. S. Francy, O. D. Simmons, M. W. Ware, E. J. Granger, M. D. Sobsey, and F. W. Schaefer, “Effects of seeding procedures and water quality on recovery of Cryptosporidium oocysts from stream water by using US environmental protection agency method 1623,” Appl. Environ. Microbiol., 70 (7), 4118 –4128 (2004). https://doi.org/10.1128/AEM.70.7.4118-4128.2004 0099-2240 Google Scholar

8. 

W.-T. Liu and C. Lay, “Lab-on-a-chip devices for microbial monitoring and detection in water,” Water Sci. Technol.: Water Supply, 7 (2), 165 –172 (2007). https://doi.org/10.2166/ws.2007.051 1606-9749 Google Scholar

9. 

A. B. Deisseroth, Z. F. Zu, D. Claxton, E. G. Hanania, S. Q. Fu, D. Ellerson, L. Goldberg, M. Thomas, K. Janicek, W. F. Anderson, J. Hester, M. Korbling, A. Durett, R. Moen, R. Berenson, S. Heimfeld, J. Hamer, L. Calvert, P. Tibbits, M. Talpaz, H. Kantarjian, R. Champlin, and C. Reading, “Genetic marking shows that Ph(+) cells present in autologous transplants of chronic myelogenous leukemia (Cml) contribute to relapse after autologous bone-marrow in cml,” Blood, 83 (10), 3068 –3076 (1994). 0006-4971 Google Scholar

10. 

H. J. Gross, B. Verwer, D. Houck, R. A. Hoffman, and D. Recktenwald, “Model study detecting breast-cancer cells in peripheral-blood mononuclear-cells at frequencies as low as 10(7),” Proc. Natl. Acad. Sci. U.S.A., 92 (2), 537 –541 (1995). https://doi.org/10.1073/pnas.92.2.537 0027-8424 Google Scholar

11. 

K. L. Johnson, H. Stroh, K. Khosrotehrani, and D. W. Bianchi, “Spot counting to locate fetal cells in maternal blood and tissue: a comparison of manual and automated microscopy,” Microsc. Res. Tech., 70 (7), 585 –588 (2007). https://doi.org/10.1002/jemt.20439 1059-910X Google Scholar

12. 

D. W. Bianchi, G. K. Zickwolf, G. J. Weil, S. Sylvester, and M. A. DeMaria, “Male fetal progenitor cells persist in maternal blood for as long as 27years postpartum,” Proc. Natl. Acad. Sci. U.S.A., 93 (2), 705 –708 (1996). https://doi.org/10.1073/pnas.93.2.705 0027-8424 Google Scholar

13. 

D. A. Veal, D. Deere, B. Ferrari, J. Piper, and P. V. Attfield, “Fluorescence staining and flow cytometry for monitoring microbial cells,” J. Immunol. Methods, 243 (1–2), 191 –210 (2000). https://doi.org/10.1016/S0022-1759(00)00234-9 0022-1759 Google Scholar

14. 

K. Lemarchand, N. Parthuisot, P. Catala, and P. Lebaron, “Comparative assessment of epifluorescence microscopy, flow cytometry, and solid-phase cytometry used in the enumeration of specific bacteria in water,” Aquat. Microb. Ecol., 25 (3), 301 –309 (2001). https://doi.org/10.3354/ame025301 0948-3055 Google Scholar

15. 

H. M. Shapiro, Practical Flow Cytometry, 4th ed.John Wiley & Sons, Inc., Hoboken, NJ (2002). Google Scholar

16. 

B. C. Ferrari, K. Stoner, and P. L. Bergquist, “Applying fluorescence-based technology to the recovery and isolation of Cryptosporidium and Giardia from industrial wastewater streams,” Water Res., 40 (3), 541 –548 (2006). https://doi.org/10.1016/j.watres.2005.11.037 0043-1354 Google Scholar

17. 

R. G. McClelland and A. C. Pinder, “Detection of low-levels of specific salmonella species by fluorescent-antibodies and flow-cytometry,” J. Appl. Bacteriol., 77 (4), 440 –447 (1994). 0021-8847 Google Scholar

18. 

D. Jin, R. Connally, and J. Piper, “Practical time-gated luminescence flow cytometry: I. Concepts,” Cytometry, Part A, 71A (10), 783 –796 (2007). https://doi.org/10.1002/cyto.a.20450 1552-4922 Google Scholar

19. 

D. Jin, R. Connally, and J. Piper, “Practical time-gated luminescence flow cytometry: II. Experimental evaluation using UV LED excitation,” Cytometry, Part A, 71A (10), 797 –808 (2007). https://doi.org/10.1002/cyto.a.20449 1552-4922 Google Scholar

20. 

R. C. Leif, D. Jin, J. Piper, L. M. Vallarino, J. W. Williams, S. Yang, and R. M. Zucker, “Calibration beads containing luminescent lanthanide ion complexes,” J. Biomed. Opt., 14 (2), 024022 (2009). https://doi.org/10.1117/1.3103646 1083-3668 Google Scholar

21. 

J. L. Yuan, G. L. Wang, H. Kimura, and K. Matsumoto, “Highly sensitive time-resolved fluoroimmunoassay of human immunoglobulin E by using a new europium fluorescent chelate as a label,” Anal. Biochem., 254 (2), 283 –287 (1997). https://doi.org/10.1006/abio.1997.2444 0003-2697 Google Scholar

22. 

G. Vesey, J. S. Slade, M. Byrne, K. Shepherd, and C. R. Fricker, “A new method for the concentration of cryptosporidium oocysts from water,” J. Appl. Bacteriol., 75 (1), 82 –86 (1993). 0021-8847 Google Scholar

23. 

D. Jin, R. Connally, and J. Piper, “Long-lived visible luminescence of UV LEDs and impact on LED excited time-resolved fluorescence applications,” J. Phys. D, 39 (3), 461 –465 (2006). https://doi.org/10.1088/0022-3727/39/3/006 0022-3727 Google Scholar

24. 

R. Connally, D. Y. Jin, and J. Piper, “High intensity solid-state UV source for time-gated luminescence microscopy,” Cytometry, Part A, 69A (9), 1020 –1027 (2006). https://doi.org/10.1002/cyto.a.20326 1552-4922 Google Scholar

25. 

G. Vesey, D. Deere, M. R. Gauci, K. R. Griffiths, K. L. Williams, and D. A. Veal, “Evaluation of fluorochromes and excitation sources for immunofluorescence in water samples,” Cytometry, 29 (2), 147 –154 (1997). https://doi.org/10.1002/(SICI)1097-0320(19971001)29:2<147::AID-CYTO7>3.0.CO;2-A 0196-4763 Google Scholar

26. 

R. C. Leif, L. M. Vallarino, M. C. Becker, and S. Yang, “Increasing the luminescence of lanthanide complexes,” Cytometry, Part A, 69A (8), 767 –778 (2006). https://doi.org/10.1002/cyto.a.20321 1552-4922 Google Scholar

27. 

J. I. Rosenblatt, J. A. Hokanson, S. R. McLaughlin, and J. F. Leary, “Theoretical basis for sampling statistics useful for detecting and isolating rare cells using flow cytometry and cell sorting,” Cytometry, 27 (3), 233 –238 (1997). https://doi.org/10.1002/(SICI)1097-0320(19970301)27:3<233::AID-CYTO4>3.0.CO;2-F 0196-4763 Google Scholar

28. 

R. C. Leif, L. M. Vallarino, M. C. Becker, and S. Yang, “Increasing lanthanide luminescence by use of the RETEL effect,” Cytometry, Part A, 69A (8), 940 –946 (2006). https://doi.org/10.1002/cyto.a.20318 1552-4922 Google Scholar

29. 

H. H. Cui, J. G. Valdez, J. A. Steinkamp, and H. A. Crissman, “Fluorescence lifetime-based discrimination and quantification of cellular DNA and RNA with phase-sensitive flow cytometry,” Cytometry, Part A, 52A (1), 46 –55 (2003). https://doi.org/10.1002/cyto.a.10022 1552-4922 Google Scholar

30. 

C. Deka and J. A. Steinkamp, “Time-resolved fluorescence-decay measurement and analysis on single cells by flow cytometry,” Appl. Opt., 35 (22), 4481 –4489 (1996). https://doi.org/10.1364/AO.35.004481 0003-6935 Google Scholar

31. 

B. L. Sailer, J. G. Valdez, J. A. Steinkamp, Z. Darzynkiewicz, and H. A. Crissman, “Monitoring uptake of ellipticine and its fluorescence lifetime in relation to the cell cycle phase by flow cytometry,” Exp. Cell Res., 236 (1), 259 –267 (1997). https://doi.org/10.1006/excr.1997.3717 0014-4827 Google Scholar

32. 

J. A. Steinkamp and H. A. Crissman, “Resolution of fluorescence signals from cells labeled with fluorochromes having different lifetimes by phase-sensitive flow-cytometry,” Cytometry, 14 (2), 210 –216 (1993). https://doi.org/10.1002/cyto.990140214 0196-4763 Google Scholar

33. 

J. A. Steinkamp, N. M. Lehnert, J. F. Keij, and B. E. Lehnert, “Enhanced immunofluorescence measurement resolution of surface antigens on highly autofluorescent, glutaraldehyde-fixed cells analyzed by phase-sensitive flow cytometry,” Cytometry, 37 (4), 275 –283 (1999). https://doi.org/10.1002/(SICI)1097-0320(19991201)37:4<275::AID-CYTO4>3.0.CO;2-U 0196-4763 Google Scholar

34. 

J. A. Steinkamp, B. E. Lehnert, and N. M. Lehnert, “Discrimination of damaged/dead cells by propidium iodide uptake in immunofluorescently labeled populations analyzed by phase-sensitive flow cytometry,” J. Immunol. Methods, 226 (1–2), 59 –70 (1999). https://doi.org/10.1016/S0022-1759(99)00053-8 0022-1759 Google Scholar

35. 

A. J. Bromm Jr., R. C. Leif, J. R. Quagliano, and L. M. Vallarino, “Addition of a second lanthanide ion to increase the luminescence of europium(III) macrocyclic complexes,” Proc. SPIE, 3604 263 –272 (1999). https://doi.org/10.1117/12.349209 0277-786X Google Scholar

36. 

R. C. Leif, M. C. Becker, A. J. Bromm Jr., L. M. Vallarino, J. W. Williams, S. A. Williams, and S. Yang, “Optimizing the luminescence of lanthanide(III) macrocyclic complexes for the detection of anti-5BrdU,” Proc. SPIE, 4622 250 –261 (2002). https://doi.org/10.1117/12.468350 0277-786X Google Scholar

37. 

R. C. Leif, M. C. Becker, A. J. Bromm Jr., L. M. Vallarino, and S. Yang, “Fluorescence resonance energy transfer enhanced luminescence (FRETEL) of Quantum Dyes,” Proc. SPIE, 6092 29 –41 (2006). 0277-786X Google Scholar

38. 

R. C. Leif and L. M. Vallarino, “Conjugated polymer tag complexes,” 1 –101 (2001). Google Scholar

39. 

R. C. Leif, M. C. Becker, A. J. Bromm Jr., L. M. Vallarino, S. A. Williams, and S. Yang, “Increasing the luminescence of lanthanide(III) macrocyclic complexes by the use of polymers and lanthanide enhanced luminescence,” Proc. SPIE, 4260 184 –197 (2001). https://doi.org/10.1117/12.426770 0277-786X Google Scholar

40. 

J. Wu, G. L. Wang, D. Y. Jin, J. L. Yuan, Y. F. Guan, and J. Piper, “Luminescent europium nanoparticles with a wide excitation range from UV to visible light for biolabeling and time-gated luminescence bioimaging,” Chem. Commun. (Cambridge), 3 365 –367 (2008). https://doi.org/10.1039/b715054g 1359-7345 Google Scholar

41. 

R. C. Leif, S. P. Clay, H. G. Gratzner, H. G. Haines, K. V. Rao, and L. M. Vallarino, “Markers for instrumental evaluation of cells of the female reproductive tract: Existing and new markers,” The Automation of Uterine Cancer Cytology, 313 –344 (1976) Google Scholar
©(2009) Society of Photo-Optical Instrumentation Engineers (SPIE)
Dayong Jin, James A. Piper, Robert C. Leif, Sean Yang, Belinda C. Ferrari, Jingli Yuan, Guilan Wang, Lidia M. Vallarino, and John W. Williams "Time-gated flow cytometry: an ultra-high selectivity method to recover ultra-rare-event μ-targets in high-background biosamples," Journal of Biomedical Optics 14(2), 024023 (1 March 2009). https://doi.org/10.1117/1.3103770
Published: 1 March 2009
Lens.org Logo
CITATIONS
Cited by 37 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Flow cytometry

Particles

Ultraviolet radiation

Ultraviolet light emitting diodes

Luminescence

Europium

Calibration

Back to Top