Open Access
14 February 2022 Modular head-mounted cortical imaging device for chronic monitoring of intrinsic signals in mice
Mark Christian Guinto, Makito Haruta, Yuki Kurauchi, Taisuke Saigo, Kazuki Kurasawa, Sumika Ryu, Yasumi Ohta, Mamiko Kawahara, Hironari Takehara, Hiroyuki Tashiro, Kiyotaka Sasagawa, Hiroshi Katsuki, Jun Ohta
Author Affiliations +
Abstract

Significance: Intrinsic optical signals (IOS) generated in the cortical tissue as a result of various interacting metabolic processes are used extensively to elucidate the underlying mechanisms that govern neurovascular coupling. However, current IOS measurements still often rely on bulky, tabletop imaging systems, and there remains a dearth of studies in freely moving subjects. Lightweight, miniature head-mounted imaging devices provide unique opportunities for investigating cortical dynamics in small animals under a variety of naturalistic behavioral settings.

Aim: The aim of this work was to monitor IOS in the somatosensory cortex of wild-type mice by developing a lightweight, biocompatible imaging device that readily lends itself to animal experiments in freely moving conditions.

Approach: Herein we describe a method for realizing long-term IOS imaging in mice using a 0.54-g, compact, CMOS-based, head-mounted imager. The two-part module, consisting of a tethered sensor plate and a base plate, allows facile assembly prior to imaging sessions and disassembly when the sensor is not in use. LEDs integrated into the device were chosen to illuminate the cortical mantle at two different wavelengths in the visible regime (λcenter: 535 and 625 nm) for monitoring volume- and oxygenation state-dependent changes in the IOS, respectively. To test whether the system can detect robust cortical responses, we recorded sensory-evoked IOS from mechanical stimulation of the hindlimbs (HL) of anesthetized mice in both acute and long-term implantation conditions.

Results: Cortical IOS recordings in the primary somatosensory cortex hindlimb receptive field (S1HL) of anesthetized mice under green and red LED illumination revealed robust, multiphasic profiles that were time-locked to the mechanical stimulation of the contralateral plantar hindpaw. Similar intrinsic signal profiles observed in S1HL at 40 days postimplantation demonstrated the viability of the approach for long-term imaging. Immunohistochemical analysis showed that the brain tissue did not exhibit appreciable immune response due to the device implantation and operation. A proof-of-principle imaging session in a freely behaving mouse showed minimal locomotor impediment for the animal and also enabled estimation of blood flow speed.

Conclusions: We demonstrate the utility of a miniature cortical imaging device for monitoring IOS and related hemodynamic processes in both anesthetized and freely moving mice, cueing potential for applications to some neuroscientific studies of sensation and naturalistic behavior.

CC BY: © The Authors. Published by SPIE under a Creative Commons Attribution 4.0 Unported License. Distribution or reproduction of this work in whole or in part requires full attribution of the original publication, including its DOI.
Mark Christian Guinto, Makito Haruta, Yuki Kurauchi, Taisuke Saigo, Kazuki Kurasawa, Sumika Ryu, Yasumi Ohta, Mamiko Kawahara, Hironari Takehara, Hiroyuki Tashiro, Kiyotaka Sasagawa, Hiroshi Katsuki, and Jun Ohta "Modular head-mounted cortical imaging device for chronic monitoring of intrinsic signals in mice," Journal of Biomedical Optics 27(2), 026501 (14 February 2022). https://doi.org/10.1117/1.JBO.27.2.026501
Received: 6 September 2021; Accepted: 24 January 2022; Published: 14 February 2022
Lens.org Logo
CITATIONS
Cited by 5 scholarly publications.
Advertisement
Advertisement
KEYWORDS
Information operations

Imaging devices

Sensors

Light emitting diodes

Brain

Head

Image sensors

RELATED CONTENT

Readout architectures for high speed CMOS image sensor
Proceedings of SPIE (August 21 2013)
A 33M pixel wide color gamut image capturing system using...
Proceedings of SPIE (February 19 2013)
Optical system with a curved detector for wide field high...
Proceedings of SPIE (September 12 2021)
A QVGA size CMOS time of flight range image sensor...
Proceedings of SPIE (January 26 2006)
Characterization and system modeling of a 5-Mpixel CMOS array
Proceedings of SPIE (February 21 2007)
In vitro and in vivo on chip biofluorescence imaging using...
Proceedings of SPIE (February 06 2006)

Back to Top