1.IntroductionAlthough conceived over 20 yr1 2 ago and developed in its modern form 10 yr ago,3 two-photon excitation (TPE) fluorescence microscopy can be considered a comparatively young technique in far-field fluorescence optical microscopy, taking advantages over both wide field and confocal laser scanning microscopy (CLSM),4 5 6 7 for the study of the three-dimensional (3D) and dynamic properties of biological systems.8 9 10 11 12 13 Moreover, a comparatively new form of spectroscopy, related to the use of tiny excitation volumes, has grown together with the development of confocal and two-photon excitation microscopy. In fact, the advent of two-photon excitation microscopes has fostered the application of techniques such as fluorescence correlation spectroscopy (FCS)14 15 16 17 or fluorescence lifetime to single biological molecules and to the characterization of tissues.18 19 20 21 22 23 It is worth noting that the development of mode-locked lasers, providing moderate average power at high repetition rates and ultrafast pulses,24 25 26 matched with the increased dissemination of confocal laser scanning microscopes, favored an incredible increase of demands for developing TPE architectures. It is our opinion, that the simplest and fastest method for implementing two- or multiphoton excitation microscope is to modify a commercial CLSM. Less inviting alternatives are to modify a common optical microscope or to buy a commercial TPE microscope.27 28 There are two popular approaches to realize TPE imaging architectures based on a CLSM, namely: descanned and nondescanned mode.29 20 30 31 32 33 34 35 28 The former uses the very same optical pathway and mechanism employed in CLSM. Pinholes are removed or set to their maximum aperture and the emission signal is captured using the galvanometric scanning mirrors. In order to obtain a better spatial resolution it is also possible to keep confocal pinholes, also considering a right balance between spatial resolution and detection sensitivity.36 37 When pinhole insertion is possible, the major advantage is attained in terms of axial resolution that can be ameliorated approximately of the 40.38 The latter approach allows a significant signal-to-noise ratio increase. The signal is collected using dichroic mirrors on the emission path or external detectors without passing through the galvanometric scanning mirrors. Unfortunately, this approach introduces some problems related to external detectors positioning and environmental light noise. This means that it is not immediate as the descanned solution. Beside the imaging capabilities of a two-photon setup that exploits directly the fluorescence output, we also consider images of the samples obtained through mapping of other spectroscopic properties, such as the decay time of the excited states or the relaxation rate of the fluctuation of fluorescence. These properties can be studied in a point-by-point way on a solid sample (for example a tissue) thereby reconstructing an image,20 23 or in solution with a laser beam kept at a fixed position and waiting for the molecules to pass through the excitation volume.14 17 We will refer in the following to this second type of spectroscopic application. Fluorescence correlation spectroscopy is a very sensitive tool to investigate a variety of properties of biomolecules in solutions, from translational and rotational diffusion to chemical reactions.15 39 40 Confocal detection with low power continuous wave (cw) laser sources is usually adopted to achieve the small excitation volumes needed for the study of the fluorescence fluctuations.41 42 43 More recently two-photon (and multiphoton) excitation with pulsed infrared (IR) lasers has been exploited to enhance the background rejection and reduce the excitation volumes with respect to the more conventional single-photon confocal techniques.3 20 In both cases the time dependence of the fluorescence has been widely characterized by means of correlation methods. The analysis of the autocorrelation function of the fluorescence signal as detected by a photomultiplier tube (PMT) or an avalanche photodiode (APD), gives two main parameters, the average number of molecules in the excitation volume N¯ and the molecular translational diffusion coefficient D.44 The probability distribution of the photon counting can be used, beside correlation functions, for the description of random processes45 but it has generally received less attention than the correlation methods. Theoretical and experimental studies44 have recently shown how to extract detailed information on N¯ and the molecular brightness ε from the photon counting histogram (PCH). This theoretical treatment implements the previous ones46 by taking explicitly into account the shape of the point spread function (PSF) and the motion of the tracer molecule. The experimental test has showed that this approach can also be applied to a mixture of species characterized by different quantum yields.47 2.TPE Architecture: General Considerations2.1.MicroscopyThe main components for TPE microscopy are almost identical to those of CLSM but some sagacity has to be used.28 27 48 Currently, the mode-locked Ti: sapphire laser25 is the preferred source, also for the small amplitude fluctuations produced that could become critical in a TPE application where the fluorescence intensity depends on the square of the excitation intensity. Moreover, the trend of the market is to produce compact and easy to use mode-locked lasers and this fact not only reduces the cost and problems related to maintenance but also the complexity of the operations related to laser functioning and the room needed. The features of infrared pulsed laser are high average power, 80 MHz high repetition rate, and 100 fs short pulse width. The use of short pulses and small duty cycles is mandatory to allow image acquisition in a reasonable time while using power levels that are biologically tolerable.49 50 51 A suitable laser source, to take advantage of any available fluorescent molecule, should also be widely tunable. The average power outputs are commonly in excess (200 mW–1 W) and allow therefore to serve different instruments at the very same time. The laser beam can be directly coupled to the scanning head, enclosures, and boxes should be used to allow safe operations. Some research groups are working on optical fiber coupling and different commercial systems are available for femtosecond and picosecond operating laser sources. However, some problems and drawbacks still remain.52 In principle, any confocal scanning unit can be converted to a TPE microscope.8 7 In practice, one needs the simplest confocal unit, optically speaking. The best solution is a confocal scanning head with a very reduced optical path and requiring easy manual operations for controlling dichroic positions, pinhole aperture, and excitation power. The different excitation wavelengths require replacement of a dichroic beam splitter that usually is inserted in any confocal unit and that separates fluorescence from excitation. In terms of filters, changes have also to be done in the output module where fluorescence is collected. In fact, the excitation of light needs to be suppressed by 6–10 orders of magnitude.53 The test can be performed by recording images from large fluorescent spheres or films of fluorescent molecules at different power levels to check for quadratic power behavior. If it is less than two some reflections are affecting the acquisition. Scanning mirrors should be metal coated (silver) and should have a good thermal resistance.27 2.2.SpectroscopySpectroscopy applications can be performed either on a tiny volume fixed on the sample plane or by performing the spectroscopic measurement while scanning the sample. The former is preferred for studies of the heterogeneity of molecular solutions47 or of the binding of different fluorescent species, while the latter allows obtaining lifetime imaging of the tissues (FLIM). In any case one of the main problems is again the rejection of the stray scattering light since it lowers the fluorescence fluctuation and the efficient detection of the low fluorescence levels involved in these studies. A part from the use of high rejection filters for the IR light, the use of high quantum yield detectors with low noise equivalent power (NEP) is the crucial issue for this point. The best detectors at the moment are avalanche photodiodes that have NEP<10−15 W/Hz 0.5 for a 0.5 mm diode area and at least 70 quantum yield at 630 nm.54 A second relevant issue in the design and the use of a two-photon excitation or confocal setup for spectroscopy is the full characterization of the excitation volume that greatly determines the relaxation time of the fluorescence fluctuations. The fluorescence signal detected through an optical setup can be expressed as the product of the fluorophore concentration in the object plane times the detection efficiency function42 that can be taken as an effective laser profile. As a matter of fact, in order to quantitatively apply fluorescence fluctuation spectroscopy, one must have a detailed knowledge of the shape of this effective excitation laser profile. In an ideal confocal setup, obtained for an infinitesimal size of the detection pinhole, this function is given by the square of the PSF of the objective,55 and for a finite size of the pinhole, this limiting PSF must be convolved to the transmission function of the pinhole.4 56 The effective excitation laser profile in a confocal setup has been the object of several theoretical and experimental works and it is assumed to have a Gaussian shape both in the object plane and in the optical path direction. So far, in a TPE architecture the excitation of the fluorescent molecules occurs because of the nonlinear response of the molecules to the electric field. It is common to assume that the effective excitation profile is simply given by the square of the (Gaussian–Lorentzian) laser beam focused by the objective. The estimate of the laser beam waist is usually obtained by measuring the fluorescence correlation function of a known dye. Both the decay time of this function and the signal-to-noise ratio are related to the excitation volume. However the estimate of the signal-to-noise ratio from the fluorescence correlation function is a delicate matter due to the contribution of shot noise, of retarded fluorescence, and of uncorrelated background to the short lag time part of the correlation decay. In Sec. 3 we give examples of the characterization of the excitation volume and a comparison to the PSF determined on the imaging path of the microscope. This is made by the analysis of the fluorescence correlation and of the photon counting histogram (PCH) method and by studying the dependence of the measured excitation volume on the laser beam diameter. The PCH method allows measuring the excitation volume without resorting to the a priori knowledge of the diffusion coefficient of the observed tracer molecule. 3.An Example of Realization of Multipurpose TPE ArchitectureThis section is related to the realization of a TPE microscope for imaging and spectroscopy applications achieved through minor modifications of a commercial CLSM, in which the ability to operate as a standard CLSM has been preserved.10 57 A panoramic view of the experimental setup is given in Figure 1(a). The core of the architecture is a mode-locked Ti:sapphire infrared pulsed laser (Tsunami 3960Spectra Physics Inc., Mountain View, CA, USA), pumped by a high-power (5 W @ 532 nm) solid state laser (Millennia V, Spectra Physics Inc., Mountain View, CA, USA). The Ti:sapphire laser output can be tuned across a range from 680 to 830 nm, with a bandwidth from 4 to 12 nm, allowing excitation of a variety of fluorescent molecules normally excited by visible and ultraviolet radiation.58 59 The restriction of the tunable range is given by the set of mirrors installed for our purposes. A dichroic mirror, optimized for high power ultrashort infrared pulses (CVI, USA), is used to bring in air the Tsunami beam to the scanning head. Before entering into the scanning head, beam average power is brought to values ranging from 2 to 50 mW using a neutral density rotating wheel (Melles Griot, USA). For an average power of 20 mW at the entrance of the scanning head, the average power before the microscope objective is about 9–13 mW and at the sample is estimated between 3 and 5 mW. At the focal volume a 1.5–2. times broadening occurs using a high numerical aperture objective and a reduced amount of optics within the optical path.32 60 61 The scanning and acquisition system for microscopic imaging is based on a commercial single-pinhole scanning head Nikon PCM2000 (Nikon Instr., Florence, Italy) mounted on the lateral port of a common inverted microscope, Nikon Eclipse TE300. Ths scanning head has a simple and compact light path that makes it very appropriate for conversion to a two-photon scope.62 28 Figure 1(b) outlines in a somplified scheme its optical design. The optical resolution performances of this microscope when operating in conventional confocal mode, and using a 100×/1.3 NA oil immersion objective, have been reported in detail elsewhere and are 178±21 nm laterally and 509±49 nm axially.63 The scanning head operates in the “open pinhole” condition. A series of emission custom made filters, that block infrared radiation (>650 nm) to an optical density of 6–7 within 50 mW of beam power incident on the filters themselves, are employed.53 In particular, the E650P filter (Chroma Inc., Brattleboro, VT, USA) constitutes the base for the other emission filters. One-photon and two-photon modes can be simply accomplished by switching from the single-mode optical fiber (one photon), coupled to a module containing conventional laser sources (Ar-ion, He–Ne green), to the optical path in air, delivering the Tsunami laser beam (two photon), as shown in Figure 2. A high throughput optical fiber delivers the emitted fluorescence from the scanning head to the PCM2000 control unit where PMTs (R928, Hamamatsu, Japan) are physically plugged. Axial scanning for confocal and TPE three-dimensional imaging is actuated by means of two different positioning devices, namely: a belt driven system using a dc motor (RFZ-A, Nikon, Japan) and a single objective piezoelectric nano-positioner (PIFOC P-721-17, Physik Instrumente, Germany). Acquisition and visualization operations are completely computer controlled by a dedicated software, EZ2000 (Coord, The Netherlands; http://www.coord.nl). The main controls are related to PMTs voltage, pixel dwell time, frame dimensions ( 1024×1024 maximum), field of scan (from 1 to 140 μm using a 100× objective). For the spectroscopy applications the laser beam (30 of the output power) is deviated by a beamsplitter just in front of the laser output and is sent to the rear port of the microscope by a beam steering, see Figure 1(a).57 The laser power at the object plane can be set by means of neutral filters. In order to achieve the smallest volumes we must expand the laser beam in order to fill almost completely the entrance pupil of the objective. We have found it is almost equivalent to expand the laser beam two or three times right at the entrance of the microscope with a beam expander, or to steer the beam along a longer path on the optical bench with a couple of fast mirrors and let the beam expand because of its intrinsic divergence. In both cases the beam reaches a diameter of approximately 3 mm which almost fills the entrance pupils of the employed objectives. The use of the beam expander should deteriorate the pulse width of the laser with respect to the use of fast mirrors, however we have not detected an appreciable difference in the fluorescence output with the two methods. The laser light is then reflected by a dichroic mirror (650 DCSPRX C72-38, Nikon, Japan) mounted on the dichroic holder sleeve of the TE300 microscope, and focused on the sample with a 100× oil objective ( NA=1.4, Plan Apochromat DICH 100× oil, working distance 0.19 mm, focal length 2 mm, Nikon, Japan). The fluorescence emission is collected by the same objective, passes through the dichroic mirror and through a band pass filter (depending on the fluorophore emission; for fluoresceine we use a HQ535/50, Chroma, VT, U.S.A.) to diminish the red and infrared light reflected toward the detector that is mounted on the bottom or the front ports. On the bottom port we detect it in single photon counting mode for the FCS applications, while on the front port we use an analog detector for the measurements of lifetimes and fluorescence polarization anisotropy decays. For single photon counting detection we have used a R928 PMT (Hamamatsu, Japan) and an avalanche photodiode (APD) (SPCM-AQ-151, EG&G, Wellesley, MA) finding that the performance of the APD is by far the best for the FCS application on highly dilute samples at low excitation power. This is probably due to the difference in the NEP of the two detectors and in the higher quantum yield of the APD compared to the photomultiplier (≅10). The measurements shown in the following have been collected with the APD. For the FCS measurements, the APD output is digitized by a photon counting module (ALBA, ISS, Urbana-Champaign, IL) and the TTL pulses are acquired by a 40 MHz acquisition board (ISS, Urbana-Champaign, IL). The correlation functions of the fluorescence fluctuations are computed online by the acquisition software and the raw data, i.e., photon counts per sampling time versus acquisition time, are stored and analyzed offline to obtain the PCHs. For the time resolved fluorescence measurements (lifetime or polarization anisotropy) the fluorescence light is detected through the front port of the microscope (180° with respect to the excitation beam direction) by a photomultiplier tube (R928, Hamamatsu, Bridgewater, NY), whose gain was modulated by biasing the second dynode stage at a radio frequency νn+360 Hz, with νn=n×80 MHz. The PMT signal is fed to an ISS lock-in amplifier board (ISS, Champaign, IL) for the measurement of the polarized modulation ratios and the phase differences of the fluorescence light with respect to the excitation laser beam. The synchronization is performed by a Lock-to-Clock system (model 3930, Spectra Physics, CA) which takes as a reference the signal of a fast diode detecting a small portion of the laser light. The 10 MHz output of the Lock-to-Clock is used as a frequency reference by the ISS board and for the synthesizer (Marconi Instruments, model 2023A, UK) that provides the rf bias for the modulation of the PMT gain. The polarization of the fluorescence is selected by means of a Glan–Thompson polarizer (extinction ratio <10−6, Bernardt-Halle, Berlin), and the fluorescence measurements are taken with the polarizer at the magic angle18 of 54.7°. The reference sample for the lifetime measurements is a basic solution (pH≅8) of Fluoresceine (Fluka Chemika, Cat. No. 46955), whose lifetime64 is ≅4.05 ns. The fluorescence anisotropy measurements are taken by rotating the polarizer between the directions parallel and perpendicular to that of the excitation light; the differential phase shifts and the polarized modulation ratios are now the ones of the fluorescence detected with parallel polarization taken with respect to the perpendicular component. For each harmonic of the laser repetition frequency a minimum of one and a maximum of five differential measurements are performed in order to obtain a differential phase shift error <0.2 and a polarized modulation ratio error <0.004. For each measurement a minimum of 20 to a maximum of 120 acquisitions are taken each corresponding to a single wave form (period≅2.78 ms); the integration time per measurement is therefore ≅55–330 ms. 3.1.TPE Microscope Behavior3.1.1.ImagingPoint spread function (PSF) measurements are referred to a planachromatic Nikon 100×, 1.4 NA immersion oil objective with enhanced transmission in the infrared region. Blue fluorescent carboxylate modified microspheres 0.1 μm diameter (F-8797, Molecular Probes, OR) were used. A drop of dilute samples of bead suspensions was spread between two coverslips of nominal thickness 0.17 mm. These microspheres constitute a very good compromise toward the utilization of subresolution point scatterers and acceptable fluorescence emission. An object plane field of 18 μm×18 μm was imaged in a 512×512 frame, at a pixel dwell time of 17 μs. Axial scanning was performed and 21 optical consecutive and parallel slices have been collected at steps of 100 nm. The x–y scan step was 35 nm. The scanning head pinhole was set to open position. The 3D data sets of several specimens were analyzed. The measured full width at half maximum (FWHM) lateral and axial resolutions were 210±40 and 700±50 nm, respectively.10 In order to be sure of operating in the TPE regime the quadratic behavior of the fluorescence intensity versus excitation power was also demonstrated. Another interesting issue is given by the demonstration of the localized photobleaching as compared with the single-photon confocal case. Large fluorescent microspheres (i.e., 22 μm diameter) can be used in a three-dimensional acquisition session. Figure 3(a) shows a three-dimensional selective photobleaching in the case of one-photon confocal (left side) and two-photon (right side) imaging modes. Photobleaching has been produced performing a zoom on a 1.96 μm2 squared area appreciatively in the center plane of the microsphere. Then three-dimensional imaging has been performed using confocal and TPE mode. The resulting images are intended only for qualitative comparison and the selective bleaching can be useful to have a better feeling with respect to the power commonly used. From Figure 3(a), in the case of confocal imaging it is possible to see a double-cone-like image according to imaging conditions occurring in confocal microscopy4 3 65 5 produced within the bead while in the TPE case the dark volume is extremely localized,29 33 66 67 demonstrating the main difference between these two three-dimensional imaging methods: it must be remembered that in confocal imaging, even though out of focus fluorescence is not detected, it is generated. Figure 3(b) shows a writing application of the TPE microscope. Figure 4 shows a TPE image of Saccharomices cerevisiae cells marked with DAPI. This image conjugates the autofluorescence of the sample with the DAPI–DNA fluorescence coming from the nucleus. In this figure imaging is coupled to active actions like the removal of cell layers increasing the power of the beam and residence time. 3.1.2.SpectroscopyHere, the excitation volume is measured from the diffusion of yellow–green fluorescent latex microspheres of diameter 64±9 nm (Polysciences, Fluoresbrite YG, Cat. No. 17149) suspended in de-ionized and filtered (0.22 μm, Millipore filter) water. It is essential that the concentration obtained by dilution from the mother solution (corresponding to ≅400 nM), is determined by light absorption after filtering with a 0.22 μm Millipore filter. The absorbance has been measured by getting rid of the scattering contribution. All the measurements reported hereafter have been performed at room temperature. The fluorescence correlation function gF(t)=〈δF(t)δF(0)〉/〈F(0)〉2 for two photon excitation involves the numerical computation of a one-dimension integral,68 however it can be accurately described by a simpler form of the type as it is shown in Figure 5 for the case λ exc =770 nm and w0=0.5 μm. The extrapolation at t=0 is gF(0)=κ/N EXC , where the contrast factor is κ=0.076, and the number of molecules is related to the number concentration 〈C〉, and the volume of the excitation profile V EXC =πw0 4/λ by N EXC =V EXC 〈C〉. The relaxation time obtained from the fit depends on the laser beam waist and the dye diffusion coefficient as τ=αw0 2/(8D), where α=1.15±0.02. 57As an example of the analysis of the FCS measurements, we show results taken at various the excitation powers in the range 1.3<P<7 mW (at the entrance of the objective). The best fit parameters, gF(0) and τ are reported in the upper panel of Figure 6. The behavior of gF(0) in Figure 6 is due to the increasing effect of the uncorrelated background bg with decreasing powers and can be accounted for by the relation16 gF(0)=(κ/V EXC 〈C〉)(〈F〉−bg)2/〈F〉2, where the dependence on the power P is in the total fluorescence rate 〈F〉 which scales as P2, and in the background bg, which scales as P. The average fluorescence rate 〈F〉 is a measured quantity and the background level is fit to a trend of the type, bg=bg0+βP. The best-fit curve is shown in the upper panel of Figure 6 and corresponds to g0=0.29±0.008, bg0=135±10, and β≅0.1 Hz/mW. The measured relaxation time does not vary appreciably and is constant τ=3.1±0.3 ms. From this value we can give a first estimate of the laser beam waist w0 by exploiting the relation τ=αw0 2/(8D) with α≅1.15. The diffusion coefficient of the microspheres D=7.45 μm 2/s is computed according to their nominal radius R=32 nm, at T=25 °C and by employing this value we find V EXC =πw0 4/λ=0.11±0.04 fL. On the other hand, from the g0=0.29±0.008 value and 〈C〉≅4 nM we can compute V EXC =κ/(g0〈C〉)=0.11±0.005 fL, in very good agreement with the estimate obtained from the relaxation time. A further important test of the two-photon setup for spectroscopy applications is the study of the dependence of the excitation volume on the beam diameter at the entrance pupil of the objective. We give an example of such an analysis in Figure 7, where a power law fit to the data is reported as a solid line on the data of the excitation volumes obtained with the analysis reported in the previous section at various values of the beam expansion. We can write57 the relation between V EXC and the beam expansion N as V EXC =πλ3(0.61/nN)4((f/Ф0)2+N2)2. If the ratio f/Ф0≫N max =3, the previous equation becomes The power law best fit of the data in Figure 7, V EXC =(4.8±0.9)N−(3.9±0.3) fL, gives an exponent very close to the expected fourth power. We can also estimate the laser beam radius at the entrance of the beam expander, Ф0=0.63±0.07 mm, by substituting the focal length of the objective, f=2 mm, and the oil refraction index, n=1.517 in the relation (πλ3)(0.61/n)4(f/Ф0)4=4.8±0.9 fL. This value of Ф0 is in good agreement with the Ti: sapphire laser specifications (beam radius at 1/e2 is <1 mm).How robust are these methods for the determination of the excitation volume? Undesired aggregation can affect both methods since we are basically counting the number of independent fluorescence entities. Even after extensive filtering of the samples one finds a small fraction of aggregates, which however could be easily picked out of the averages. This is shown for example in Figure 8, where we plot the PCH computed on a single 1.6 s trace in which one observes a huge fluorescence peak (see inset) and from which a double peaked histogram is computed (see filled squares in Figure 8). In the same plot, for comparison, we report the PCH computed on a second 1.6 s trace where no huge peaks are detected and the histograms are of the same type as those analyzed in the previous paragraphs. The anomalous behavior shown in Figure 8 is very likely due to the passage of aggregates through the excitation volume. Such events, however, are very rare indicating that the number of aggregates is rather low and the measurements affected by aggregates can be singled out by looking at the average rate of each measurement. A second and more critical experimental issue that can be attacked by looking at the fluorescence fluctuations, is whether the laser beam at the object plane is well shaped, i.e., it follows a Gaussian–Lorentzian shape. We have to this purpose performed a series of measurements versus the sample concentration in a case when approximately half of the laser beam cross section (which was expanded three times) was blocked at the entrance of the microscope by a misplaced diaphragm. The analysis of the auto correlation functions (ACFs) gives reasonable results although the residuals always indicate the presence of a second fast component that was more evident at higher sample concentrations (data not shown). If we force fit the ACFs to Eq. (1), the dependence of the gF(0) and the relaxation time τ upon the concentration is in reasonable agreement with the expected laws, i.e., gF(0) decreases with the concentration and τ is approximately constant (data not shown). However the estimate of the excitation volume from these two data sets disagree significantly, since V EXC =0.5±0.05 fL, from the gF(0) analysis, and V EXC =0.26±0.05 fL from the relaxation time. These values are sensibly larger than the result found when no perturbation of the laser cross section was present, V EXC ≅0.1 fL. It must be noted that beside the larger value of the excitation volume, the expectations of the theoretical trends of the best fit parameters are not fulfilled when the beam cross section is perturbed. 3.1.3.Two-Photon Lifetime DecaysA single component model can satisfactorily analyze the lifetime decays for the Rhodamine 6G (Fluka Chemika, Cat. No. 83698) solutions to give a lifetime τF=3.95±0.2 ns. The two quantities measured are the phase shifts and the demodulation ratios at the harmonics of the laser repetition frequency ≅80 MHz. These data were fit versus the frequency by a two-exponential decay scheme using the GLOBALS UNLIMITED software,69 in order to obtain the fluorescence lifetimes of the fluorophore. We can also show as an example of protein studies, the fluorescence decay of the Alexa dye (succymidil-esther derivative by Molecular Probes) bound to the Beta-lactoglobulin B (Sigma Chemical Co., Cat. no. L-8005, Lot. No. 13H7150) where a double exponential decay is found. The main decay component corresponds to a lifetime τF=3.4±0.4 ns, and there is a tiny amount (<1) of a faster decay (≅0.4 ns). 3.1.4.Microscope Polarization Response and Fluorescence Polarization AnisotropyTwo anisotropy components are often needed to satisfactorily interpret the differential phase shifts and polarized demodulation ratios. We ascribe these decays to the same fluorophore according to the time domain form: where τW is the libration time of the fluorophore in the binding site, τ ROT is the rotational time, and rW and r ROT are the limiting anisotropies of the two components. From τ ROT we obtain the rotational diffusion coefficient70 D ROT =(6τ ROT )−1. The anisotropy shape of the molecules must be taken into account for a correct interpretation of the rotational times. However, an apparent radius can be assigned by inverting the Stokes–Einstein equation where kB is the Boltzmann constant, T is the absolute temperature, and η is the solvent viscosity. All measurements were made at T=22.5 °C.In order to measure the fluorescence polarization anisotropy (FPA) it is essential to correctly measure the components of the fluorescence intensity with polarization directions parallel I∥ and perpendicular I⊥ to the direction of the polarization of the excitation light. Therefore the microscope transmission function for the light with different polarizations must be measured. In the FPA experiments the exciting light coming from the laser is polarized in the plane determined by the direction of the light entering the microscope from the epifluorescence port and the vertical direction. The reflections on the microscope optical components (dichroic beam splitters, mirrors) and the polarization of the excitation light are always in this plane. The fluorescence light from the sample, instead, is partially depolarized and the polarization direction of the fluorescence collected by the photomutiplier can be selected by a Glan–Thompson polarizer. We indicate by two subscripts the polarization of the excited light and the position of the emission polarizer, respectively, measured by the detector. For example, I VH corresponds to vertical polarized excitation light and horizontal position of the emission polarizer. We disregard for the moment the finite size of the excitation volume in which we assume that the polarization of the light is homogenous. For vertically polarized excitation light the observed fluorescence intensities are where SV and SH are the sensitivities of the microscope front port for the vertically and horizontally polarized components, respectively, and kV and kH are proportional to the excitation intensity with parallel and perpendicular polarization that is actually coming onto the sample. The ratio I VV /I VH gives To calculate the intensity ratio I∥/I⊥ we need to determine the so-called G factor, G=SV/SH. Differently from the usual configuration,71 we cannot use horizontal polarized excitation light in order to get perfectly unpolarized fluorescence, i.e., I∥=I⊥, in Eq. (4). In the present backscattering configuration we exploit the case of horizontally polarized excitation that is now described by the following relations: In order to measure the G factor we collect the fluorescence emission for all the four cases described by Eqs. (5), (7), and by combining these equations we obtain The rotation of the vertical polarization of the excitation light to the horizontal one is obtained by means of a beam steering and checked by exploiting the reflection of the light on a glass slide at the Brewster angle. The uncertainty on this measurement is ≅0.4°.As an example on our setup, the measurements on Rhodamine 6G solutions in 60 Glycerol/water give in the range 80–800 MHz a nearly constant value, G=0.864±0.003 and this has been used to correct all the anisotropy measurements reported here. We have tested the characterization of the response of the microscope to the light polarization by measuring the FPA decay on protein (Beta-lactoglobulin B) solutions at high dilution corresponding to the limit of ≅1 molecule per excitation volume or less. We have performed measurements of the fluctuation of the fluorescence, i.e., FCS, and of FPA. The analysis of the fluorescence ACFs provide us directly with the average number of molecules per excitation volume ≅κ/gF(0) that is in the range 0.4–17 (see inset of Figure 9). The average relaxation time τ=210±20 μs corresponds to a diffusion coefficient DT≅1.15w0 2/8τ=100±10 μm 2/s. The FPA measurements on the same solutions are shown in Figure 10 together with the best-fit functions according to the two-component model described by Eq. (3). The rotational time, τ ROT =9±1.5 ns, corresponds to a rotational diffusion coefficient D ROT =18±3 MHz. The implicit assumption that we have done on the origin of fast relaxation time τW is supported by its value (averaged for C>100 nM ) τW=400±40 ps, which could not correspond to the rotational relaxation time of the free dye. In fact, this should be ≅200 ps in water, according to its molecular weight. The check of the alignment and the polarization response characterization of the setup that we have described above is fulfilled by noticing that the value of the translational diffusion coefficient that is measured by FCS and FPA is in agreement with the size and shape of Beta lactoglubulin (BLG). The computations made by Garcia de la Torre et al.;72 indicate for the dimer form of BLG at T=20 °C a value DT≅78.5 μm 2/s, that corresponds to ≅84 μm2/s at T≅22.5 °C. Our experimental estimate DT=100±10 μm 2/s is ≅20 larger than the theoretical value and corresponds approximately to the value expected for the monomer of BLG. This discrepancy is due to the fact that most of the BLG dimers have dissociated into monomers at concentrations C<280 nM that corresponds to a mass concentration ≅0.01 g/L. In fact we expect ≅90 of BLG monomers in the solution at C=280 nM and at ionic strength =50 mM. Also our estimate of the rotational diffusion coefficient of the Beta-lactoglobulin B by FPA measurements corresponds closely to the prediction that can be done for the monomer of this protein. The theoretical predictions made by Garcia de la Torre et al.;72 for the average rotational diffusion coefficient of the BLG dimer reported at T≅22.5 °C gives D ROT ≅8 MHz. We expect the rotational diffusion of the BLG monomer to be roughly double this figure, since the rotational diffusion scales approximately with the inverse of the molecular volume. The theoretical prediction of ≅16 MHz for the BLG monomer is therefore in good agreement with our experimental estimate D ROT =18±3 MHz. We can now compare the estimate of the excitation volumes obtained on the ports for imaging and spectroscopy of the microscope. The PSF of the imaging port corresponds to a three-dimensional Gaussian and from the values of the FWHM lateral and axial resolutions ( wr0=210±40 nm and z0=700±50 nm ), we can estimate the excitation volume on the imaging port as V EXC ≅(π/4 ln(2))1.5 wr0 2z0≅0.03±0.013 fL. For both the imaging and the spectroscopy ports the values of the excitation volumes are close to the minimum that can be reached by optical setup. The ≅3 times smaller excitation volume observed for the imaging port is therefore to be ascribed to the different optical paths. For the spectroscopy port a collimated beam fitting the entrance aperture of the objective is used, while in the imaging path a point source is conjugated on the sample plane. We suggest that the parallel use of the two paths for microspectroscopy must be made with caution since the volumes from which the fluorescence is sampled are different in the two ports. 4.ConclusionTPE microscopy broke into the arena bringing its intrinsic three-dimensional resolution, the absence of background fluorescence, and the charming possibility of exciting ultraviolet excitable fluorescent molecules at infrared wavelengths increasing sample penetration. Moreover, TPE microscopy can be considered an essential imaging system for thick sample and live cell imaging and a powerful setup for the detection of rarely occurring molecular species and their mutual interaction based on the detection of the fluorescence fluctuations or the lifetime decay time. At present, the main disadvantages of TPE techniques come from the cost of femtosecond laser sources and the difficulty in predicting or measuring two-photon absorption spectra of the fluorescent molecules potentially utilizable. Because the idea of a TPE microscope is surprisingly straightforward to implement in this paper it has been discussed the conversion of a CLSM to a two-photon multipurpose architecture.73 74 75 76 77 78 79 80 81 82 83 84 AcknowledgementsThe authors would like to acknoledege Enrico Gratton. TPE architecture has been realized and it is mantained by INFM grants. REFERENCES
J. N. Gannaway
and
C. J. R. Sheppard
,
“Second harmonic imaging in the scanning optical microscope,”
Opt. Quantum Electron. , 10 435
–439
(1978). Google Scholar
C. J. R. Sheppard
and
R. Kompfner
,
“Resonant scanning optical microscope,”
Appl. Opt. , 17 2879
–2885
(1978). Google Scholar
W. Denk
,
J. H. Strickler
, and
W. W. Webb
,
“Two-photon laser scanning fluorescence microscopy,”
Science , 248 73
–76
(1990). Google Scholar
M. Gu
and
C. J. R. Sheppard
,
“Comparison of three-dimensional imaging properties between two-photon and single-photon fluorescence microscopy,”
J. Microsc. , 177 128
–137
(1995). Google Scholar
R. H. Webb
,
“Confocal optical microscopy,”
Rep. Prog. Phys. , 59 427
–471
(1996). Google Scholar
S. W. Hell
,
“Non linear optical microscopy,”
Bioimaging , 4
(3), 121
–224
(1996). Google Scholar
A. Diaspro
,
“Two-photon excitation microscopy,”
IEEE Eng. Med. Biol. Mag. , 18
(5), 16
–99
(1999). Google Scholar
A. Diaspro
,
“Two-photon microscopy,”
Microsc. Res. Tech. , 47 163
–212
(1999). Google Scholar
D. Magde
and
E. L. Elson
,
“Fluorescence correlation spectroscopy, II. An experimental realization,”
Biopolymers , 13 29
–61
(1974). Google Scholar
E. L. Elson
and
D. Magde
,
“Fluorescence correlation spectroscopy, I. Conceptual basis and theory,”
Biopolymers , 13 1
–27
(1974). Google Scholar
P. Schwille
,
J. Bieschke
, and
F. Oehlenschlager
,
“Kinetic investigation by fluorescence correlation spectroscopy—the analytical and diagnostic potential of diffusion studies,”
Biophys. Chem. , 66 211
–228
(1997). Google Scholar
T. W. J. Gadella
,
T. M. Jovin
, and
R. M. Clegg
,
“Fluorescence lifetime imaging microscopy (FLIM)-spatial resolution of microstructures on the nanosecond time-scale,”
Biophys. Chem. , 48 221
–239
(1993). Google Scholar
P. T. C. So, K. M. Berland, T. French, C. Y. Dong, and E. Gratton, “Two photon fluorescence microscopy: Time resolved and intensity imaging,” in Fluorescence Imaging Spectroscopy and Microscopy, X. F. Wang and B. Herman, Eds., Chemical Analysis Series, Vol. 137, pp. 351–373, Wiley, New York (1996).
J. Sytsma
,
J. Vroom
,
H. C. Gerritsen
, and
Y. K. Levine
,
“Development of a confocal laser scanning fluorescence microscope using two-photon excitation in combination with time gated detection,”
Proc. SPIE , 2412 110
–114
(1995). Google Scholar
K. M. Berland
,
P. T. So
,
Y. Chen
,
W. W. Mantulin
, and
E. Gratton
,
“Scanning two-photon fluctuation correlation spectroscopy: particle counting measurements for detection of molecular aggregation,”
Biophys. J. , 71 410
–420
(1996). Google Scholar
A. Squire
,
P. J. Verveer
, and
P. I. H. Bastiaens
,
“Multiple frequency fluorescence lifetime imaging microscopy,”
J. Microsc. , 197 136
–149
(2000). Google Scholar
D. E. Spence
,
P. N. Kean
, and
W. Sibbett
,
“60-fsec pulse generation from a self-mode-locked Ti:sapphire laser,”
Opt. Lett. , 16 42
–45
(1991). Google Scholar
W. G. Fisher
,
E. A. Watcher
,
M. Armas
, and
C. Seaton
,
“Titanium: sapphire laser as an excitation source in two-photon spectroscopy,”
Appl. Spectrosc. , 51
(2), 218
–226
(1997). Google Scholar
A. Majewska
,
G. Yiu
, and
R. Yuste
,
“A custom-made two-photon microscope and deconvolution system,”
Pflugers Archiv-Eur. J. Physiol. , 441
(2/3), 398
–408
(2000). Google Scholar
S. M. Potter
,
C. M. Wang
,
P. A. Garrity
, and
S. E. Fraser
,
“Intravital imaging of green fluorescent protein using 2-photon laser-scanning microscopy,”
Gene , 173 25
–31
(1996). Google Scholar
K. Konig
,
H. Liang
,
M. W. Berns
, and
B. J. Tromberg
,
“Cell damage in near infrared multimode optical traps as a result of multiphoton absorption,”
Opt. Lett. , 21 1090
–1092
(1996). Google Scholar
C. Soeller
and
M. B. Cannell
,
“Construction of a two-photon microscope and optimisation of illumination pulse duration,”
Pflugers Arch. , 432 555
–561
(1996). Google Scholar
A. Periasamy
,
P. Skoglund
,
C. Noakes
, and
R. Keller
,
“An evaluation of two-photon excitation versus confocal and digital deconvolution fluorescence microscopy imaging in Xenopus morphogenesis,”
Microsc. Res. Tech. , 47 172
–181
(1999). Google Scholar
C. Soeller
and
M. B. Cannell
,
“Two-photon microscopy: Imaging in scattering samples and three dimensionally resolved flash photolysis,”
Microsc. Res. Tech. , 47 182
–195
(1999). Google Scholar
A. Diaspro
,
M. Corosu
,
P. Ramoino
, and
M. Robello
,
“Adapting a compact confocal microscope system to a two-photon excitation fluorescence imaging architecture,”
Microsc. Res. Tech. , 47 196
–205
(1999). Google Scholar
R. Gauderon
,
P. B. Lukins
, and
C. J. R. Sheppard
,
“Effect of a confocal pinhole in two-photon microscopy,”
Microsc. Res. Tech. , 47 210
–214
(1999). Google Scholar
M. Gu
and
C. J. R. Sheppard
,
“Effects of a finite sized pinhole on 3D image formation in confocal two-photon fluorescence microscopy,”
J. Mod. Opt. , 40 2009
–2024
(1993). Google Scholar
M. Eherneberg
and
R. Rigler
,
“Rotational Brownian motion and fluorescence intensity fluctuations,”
Chem. Phys. , 4 390
–401
(1974). Google Scholar
S. R. Aragon
and
R. Pecora
,
“Theory of dynamic light scattering from large anisotropic particles,”
J. Chem. Phys. , 66 2506
–2516
(1977). Google Scholar
H. Quian
and
E. L. Elson
,
“Analysis of confocal laser-microscope optics for 3-D fluorescence correlation spectroscopy,”
Appl. Opt. , 30 1185
(1991). Google Scholar
R. Rigler
,
U. Mets
,
J. Windengren
, and
P. Kask
,
“Fluorescence correlation spectroscopy with high count rate and low background: analysis of translational diffusion,”
Eur. Biophys. J. , 22 169
–175
(1993). Google Scholar
D. E. Koppel
,
F. Morgan
,
A. E. Cowan
, and
J. H. Carson
,
“Scanning concentration correlation spectroscopy using the confocal laser microscope,”
Biophys. J. , 66 502
–507
(1994). Google Scholar
Y. Chen
,
J. D. Muller
,
P. T. C. So
, and
E. Gratton
,
“The photon counting histogram in fluorescence fluctuation spectroscopy,”
Biophys. J. , 77
(1), 553
–567
(1999). Google Scholar
H. Quian
and
E. L. Elson
,
“Distribution of molecular aggregation by analysis of fluctuation moments,”
Proc. Natl. Acad. Sci. U.S.A. , 85 5479
–5483
(1990). Google Scholar
J. D. Mu¨ller
,
Y. Chen
, and
E. Gratton
,
“Resolving heterogeneity on the single molecular level with the photon-counting histogram,”
Biochem. J. , 78 474
–486
(2000). Google Scholar
A. Scho¨nle
and
S. W. Hell
,
“Heating by absorption in the focus of an objective lens,”
Opt. Lett. , 23 325
(1998). Google Scholar
K. Konig
,
“Multiphoton microscopy in life sciences,”
J. Microsc. , 200 83
–104
(2000). Google Scholar
M. Ghioni
,
S. Cova
,
F. Zappa
, and
C. Samori
,
“Compact active quenching circuit for fast photon counting with avalanche photodiodes,”
Rev. Sci. Instrum. , 67
(10), 3440
–3448
(1996). Google Scholar
G. Chirico
,
F. Olivini
, and
S. Beretta
,
“Fluorescence excitation volume in two-photon microscopy by autocorrelation spectroscopy and photon counting histogram,”
Appl. Spectrosc. , 54
(7), 1084
–1090
(2000). Google Scholar
C. Xu
,
W. Zipfel
,
J. B. Shear
,
R. M. Williams
, and
W. W. Webb
,
“Multi-photon fluorescence excitation: new spectral windows for biological nonlinear microscopy,”
Proc. Natl. Acad. Sci. (U.S.A.) , 93 10763
–10768
(1996). Google Scholar
C. Xu
and
W. W. Webb
,
“Measurement of two-photon excitation cross sections of molecular fluorophores with data from 690 to 1050 nm,”
J. Opt. Soc. Am. , 13 481
–491
(1996). Google Scholar
P. E. Hanninen
and
S. W. Hell
,
“Femtosecond pulse broadening in the focal region of a two-photon fluorescence microscope,”
Bioimaging , 2 117
–121
(1994). Google Scholar
M. Muller
,
J. Squier
, and
G. J. Brakenhoff
,
“Measurements of femtosecond pulses in the focal point of a high numerical aperture lens bt two-photon absorption,”
Opt. Lett. , 20 1038
–1040
(1995). Google Scholar
A. Diaspro
,
S. Annunziata
,
M. Raimondo
,
P. Ramoino
, and
M. Robello
,
“A single-pinhole CLSM for 3-D imaging of biostructures,”
IEEE Eng. Med. Biol. Mag. , 18
(4), 106
–110
(1999). Google Scholar
A. Diaspro
,
S. Annunziata
,
M. Raimondo
, and
M. Robello
,
“Three-dimensional optical behavior of a confocal microscope with single illumination and detection pinhole through imaging of subresolution beads,”
Microsc. Res. Tech. , 45
(2), 130
–131
(1999). Google Scholar
B. Bianco
and
A. Diaspro
,
“Analysis of the three dimensional cell imaging obtained with optical microscopy techniques based on defocusing,”
Cell Biophys. , 15 189
–200
(1989). Google Scholar
O. Nakamura
,
“Three-dimensional imaging characteristics of laser scan fluorescence microscopy: Two-photon excitation vs. single-photon excitation,”
Optik (Stuttgart) , 93 39
–42
(1993). Google Scholar
G. H. Patterson
and
D. W. Piston
,
“Photobeleaching in two-photon excitation microscopy,”
Biophys. J. , 78 2159
–2162
(2000). Google Scholar
K. M. Berland
,
P. T. C. So
, and
E. Gratton
,
“Two-photon fluorescence correlation spectroscopy: method and application to the intracellular environment,”
Biophys. J. , 68 694
(1995). Google Scholar
J. M. Beechem
and
E. Gratton
,
“Fluorescence spectroscopy data analysis environment: a second generation global analysis program,”
Proc. SPIE , 909 70
(1988). Google Scholar
J. Garcia de la Torre
,
M. L. Huertas
, and
B. Carrasco
,
“Calculation of hydrodynamic properties of globular proteins from their atomic-level structure,”
Biophys. J. , 78 719
(2000). Google Scholar
M. Albota
et al.;,
“Design of organic molecules with large two-photon absorption cross sections,”
Science , 281 1653
–1656
(1998). Google Scholar
G. Baldini
,
S. Beretta
,
G. Chirico
,
H. Franz
,
E. Maccioni
,
P. Mariani
, and
F. Spinozzi
,
“Salt induced association of Beta-lactoglobulin studied by salt light and X-ray scattering,”
Macromolecules , 32 6128
(1999). Google Scholar
V. E. Centonze
and
J. G. White
,
“Multiphoton excitation provides optical sections from deeper within scattering specimens than confocal imaging,”
Biophys. J. , 75 2015
–2024
(1998). Google Scholar
W. Denk
,
“Two-photon excitation in functional biological imaging,”
J. Biomed. Opt. , 1 296
–304
(1996). Google Scholar
A. Diaspro
and
M. Robello
,
“Two-photon excitation of fluorescence in three-dimensional microscopy,”
Eur. J. Histochem. , 43
(3), 70
–79
(1999). Google Scholar
T. Furuta
,
S. S.-H. Wang
,
J. L. Dantzker
,
T. M. Dore
,
W. J. Bybee
,
E. M. Callaway
,
W. Denk
, and
R. Y. Tsien
,
“Brominated 7-hydroxycoumarin-4-ylmethyls: Photolabile protecting groups with biologically useful cross-sections for two photon photolysis,”
Proc. Natl. Acad. Sci. U.S.A. , 96 1193
–1200
(1999). Google Scholar
S. W. Hell
,
K. Bahlmann
,
M. Schrader
,
A. Soini
,
H. Malak
,
I. Gryczynski
, and
J. R. Lakowicz
,
“Three-photon excitation in fluorescence microscopy,”
J. Biomed. Opt. , 1 71
–74
(1996). Google Scholar
H. J. Koester
,
D. Baur
,
R. Uhl
, and
S. W. Hell
,
“Ca2+ fluorescence imaging with pico- and femtosecond two-photon excitation: Signal and photodamage,”
Biophys. J. , 77 2226
–2236
(1999). Google Scholar
K. Ko¨nig
,
P. T. C. So
,
W. W. Mantulin
, and
E. Gratton
,
“Cellular response to near-infrared femtosecond laser pulses in two-photon microscopes,”
Opt. Lett. , 22 135
–136
(1997). Google Scholar
K. Ko¨nig
,
T. W. Becker
,
P. Fischer
,
I. Riemann
, and
K. J. Halbhuber
,
“Pulse-length dependence of cellular response to intense near-infrared laser pulses in multiphoton microscopes,”
Opt. Lett. , 24 113
–115
(1999). Google Scholar
E. H. K. Stelzer
,
S. Hell
, and
S. Lindek
,
“Nonlinear absorption extends confocal fluorescence microscopy into the ultra-violet regime and confines the illumination volume,”
Opt. Commun. , 104 223
–228
(1994). Google Scholar
D. W. Wokosin
,
V. E. Centonze
,
S. Crittenden
, and
J. G. White
,
“Three-photon excitation fluorescence imaging of biological specimens using an all-solid-state laser,”
Bioimaging , 4 208
–214
(1996). Google Scholar
|
CITATIONS
Cited by 32 scholarly publications and 3 patents.
Luminescence
Confocal microscopy
Microscopes
Spectroscopy
Polarization
Head
Molecules