1.IntroductionIntravoxel incoherent motion (IVIM) is a post-processing algorithm for diffusion magnetic resonance imaging (MRI) scans introduced by Le Bihan et al.;1 it uses multiple -values to simultaneously assess tissue water diffusion and tissue perfusion. Prolonged scan times, ease of use, and direct quantification of perfusion afforded by contrast-enhanced scans largely supplanted IVIM as a clinical option. However, because newer, faster scanner hardware and software have substantially reduced scan and image processing times, many groups2–6 are beginning to revisit IVIM perfusion-diffusion images as a tool for evaluating neurovascular disease. If IVIM can be established as a reliable, quantitative scan protocol that requires under 5 min of scan time and no contrast, quantitative perfusion can be added to any neurovascular scan protocol. This could provide cross-sectional and longitudinal tissue perfusion maps to allow physicians to ascertain changes in perfusion that are associated with chronic disease progression, such as intracranial atherosclerotic disease, carotid stenosis, and moyamoya disease, as well as treatment response.7–9 Furthermore, if cerebral blood flow can be quantified in ml/100 g/min (qCBF) without organ-dependent capillary segment length and dimension assumptions, IVIM has the potential to allow multi-site cross-sectional comparison in addition to within-patient longitudinal comparison in organs other than the brain.10–12 The IVIM sequence and post-processing presented here presents a clinically feasible scan time of 5-min to collect images for the calculation of parametric perfusion, diffusion, and water transport time (WTT) MR images. We hypothesize that IVIM perfusion values from multi--value diffusion with a standard IVIM bi-exponential can be quantified (ml/100 g/min) using WTT, agree with neutron capture microspheres in a setting of altered hemodynamics, and demonstrate similar sensitivity to physiologic changes. As the central volume principle of blood flow being equal to blood volume/transit time is predicated on the passage of bolus, by comparison of an IVIM model that assumes instantaneous bolus mimicking the central volume principle through isotropic Gaussian diffusion, we hypothesize that IVIM WTT should correlate with dynamic susceptibility contrast (DSC) mean transit time (MTT). Agreement would further answer questions of if observed “perfusion” changes are reflecting the flow of the perivascular cerebrospinal fluid (CSF) or the movement of blood in the capillary bed.13 Further, using the same IVIM acquisition, we hypothesize that IVIM diffusion positive infarct volumes will be accurate compared to MRI mean diffusivity (MD) infarct volumes as a reference standard. The purpose of this study is thus to assess the accuracy and sensitivity of a novel quantification of 5-min simultaneous IVIM perfusion and diffusion across three physiologic states. 2.Materials and Methods2.1.IVIM Quantification via Water Transport TimePerfusion and diffusion values are calculated from the multiple -value IVIM scan using software developed in-house with a two-step fitting algorithm built using previous literature.1,10,11 In our implementation, the diffusion regime () is fit to the mono-exponential , and the perfusion regime is fit to the remainder , resulting in the following standard bi-exponential1,10,11,14 on a voxel-by-voxel basis: where is the pseudo-diffusion coefficient, is the blood volume fraction (CBV), and is the diffusion coefficient.6With the bi-exponential fit returning an approximation of a pseudo-diffusion coefficient representing the fast component of water motion, the movement of perfusing water can be modeled as a one-dimensional Gaussian. WTT is then defined as the time for 50% of the original perfusing water to have left a voxel. With three directions taken, this can be written in spherical coordinates for three-dimensional (3D) pseudo-diffusion, set equal to 0.5 and solved for WTT as Quantitative CBF can be calculated by substituting this definition of WTT into the central volume principle, where WTT is solved from Eq. (2) as a function of , as shown in the Appendix. Fast-water fraction () per WTT () is proportional to the cerebral blood flow in ml/100 g/min with assumed values of and water content fraction , which is converted to ml/100 g/min as Using Eq. (3) on IVIM bi-exponential parameters from Eq. (1) returns IVIM maps of qCBF in ml/100 g/min for comparison to reference standard values. 2.2.Comparison with Capillary GeometryQuantitative perfusion values derived from Eq. (3) were compared directly to those derived based on capillary geometry.10 Quantification by capillary geometry requires an assumed water fraction of 0.65, density of , total capillary length of 2 mm, and average segment length = 0.11 mm. We performed a regression analysis to ascertain agreement with reference standard neutron capture microspheres in the same IVIM images 2.3.Pre-Clinical ModelAll experiments were conducted using a pre-clinical canine model of both controlled hypercapnia and MCAO.15,16 The 2-day experimental protocol was approved by the University of Chicago Institutional Animal Care and Use Committee and reported in compliance with ARRIVE guidelines. The University of Chicago is an AAALAC International accredited institution adhering to the following guidelines, regulations, and policies: (a) Guide for the Care and Use of Laboratory Animals (National Research Council), (b) USDA Animal Welfare Act and Animal Welfare Regulations, and (c) Public Health Service Policy on Humane Care and Use of Laboratory Animals. The canine model has several advantages compared to rodents as a model for assessment of infarct progression using perfusion and diffusion imaging studies. Canines have a gyrencephalic neocortex with similar ratio of white to gray matter17–19 as well as comparable pial arteriolar network organization,20 critical in modeling collateral arterial blood supply and predicting infarct evolution,21 to humans. In addition, the canine model provides tissue volumes necessary to evaluate both core and penumbra during ischemia using microsphere deposition. The neurovascular structure of canines also allows a range of endovascular devices and interventional radiology techniques allowing minimal invasion with real-time visualization of occlusion, preventing imaging artifacts and traumatic cerebrovascular reaction from open surgical occlusion. Finally, the middle cerebral artery occlusion proposed is a relatively inexpensive alternative to nonhuman primate models of acute ischemia. 2.4.Experimental ProtocolIn each animal, two experiments were performed on consecutive days to validate our algorithm in a setting of cerebrovascular response to hypercapnia and MCAO and the effect of flow augmentation on collateral arterial networks after MCAO.15,21 We present an analysis of these experiments in which neutron capture microspheres and IVIM were acquired during quiescent physiologic conditions. On the first day, canines were imaged at baseline (i.e., normocapnia) (target to 35 mmHg) and hypercapnia (5% to 7% carbogen respiration, target 60 to 70 mmHg) induced by carbogen gas inhalation (5% to 7% 95% ). End tidal was monitored to assess minute to minute variations in exhaled levels, and true arterial was measured through direct arterial blood sampled from the descending aorta at the time of perfusion measurements. On the second day, canines underwent permanent endovascular middle cerebral artery occlusion under fluoroscopic guidance. Occlusion was then verified via selective ipsilateral and contralateral internal carotid and vertebral arteriography.15 Subjects were transported to the MRI suite for imaging studies within 60 min of occlusion. We evaluated experimental flow augmentation during MCAO via simultaneous administration of a vasopressor and vasodilator intended to augment CBF;22 therefore, highly dynamic and abnormally high perfusion throughout the brain post-occlusion is expected by experimental design. Throughout the experiments, physiologic parameters were maintained within normal range (excluding for experiments involving intentional hypercapnia and blood pressure for experiments involving flow augmentation). IVIM and microspheres included in the study were taken apart on day two post-occlusion to mitigate perfusion fluctuation. Furthermore, physiologic parameter variation (heart rate, blood pressure, ) between microsphere and IVIM imaging was monitored. Large variance in physiologic parameters within these timing windows is noted, and subjects were removed from analysis [Fig. 3(b)] due to the confounding effect. 2.5.Microsphere AcquisitionStable-isotope microsphere CBF in ml/100 g/min was acquired using neutron activation.15,23–25 Each injection consisted of 4 ml of stable isotope-labeled microspheres (STERIspheres, BioPal Inc, Medford, Massachusetts) into the left ventricle via a 5 Fr pigtail catheter over 10 s and reference blood (20 ml) was withdrawn at for analysis (World Precision Instruments, Sarasota, Florida). Brains were excised and sectioned into regions of interest (ROIs), Fig. 1, and analyzed through neutron activation at an independent laboratory (BioPal Inc. Medford, Massachusetts). 2.6.MRI AcquisitionAll MRI scans were performed on a 3T MRI scanner (Ingenia, Philips) with canines in a headfirst, prone position using a 15 channel receive-only coil. Diffusion weighted images (DWI) for IVIM were collected with 10 b-values from 0 to (0, 111, 222, 333, 444, 556, 667, 778, 889, 1000) and three orthogonal directions to ensure a clinically feasible scan time (5:38 min). IVIM scans were prescribed to cover the entire head (two-dimensional single shot EPI, TR/TE = 3056/91 ms, number of slices = 50, slice thickness = 2 mm, FOV = 160 mm/matrix = 128, total scan time = 332 s, SENSE factor = 2, 3 directions). DSC images were acquired using a T1-Bookend method15 within 30 min of IVIM (FOV/matrix = 220 mm/224, voxel size = 220/224, single shot, EPI, fat saturated, number of slices = 5, slice thickness = 6 mm, TR/TE = 315/40, flip angle = 75 deg, 200 time points) and delay and dispersion corrected26 to return MTT. Diffusion tensor imaging (DTI) for infarct volume progression was prescribed to cover the entire head (TR/TE = 2993/83 ms, FA = 90 deg, BW = 1790 Hz/pixel, no. of slices = 50, slice thickness = 2 mm, FOV = 160 mm/matrix = 128, -values = 0, , 32 directions). 2.7.Perfusion ComparisonHemispheric ROIs, shown in Fig. 1, were manually drawn on three central slices on the diffusion image by a trained operator who was blinded to all perfusion values for comparison to microsphere perfusion. Three IVIM slices were averaged to match microsphere thickness (6 mm). To remove high values from misfitting as demonstrated in previous studies,10 were excluded from analysis, and the sum squared residuals were analyzed. Manual avoidance of subarachnoid space and ventricles, along with leave-one-out cross validation T2 map thresholds, was used to remove CSF-dominated voxels from analysis. Kwong et al.13 suggested using an inversion recovery saturation pulse to remove proposed artifactual high perfusion values in the brain from the glymphatic system. In this study, a simulation was written to explore the effects of this saturation pulse on the blood signal. 2.8.Transit Time ComparisonAnatomic middle cortical ROIs post-MCAO were drawn by an operator blinded to physiologic status on pre-processed images of IVIM and DSC ipsilateral and contralateral to the occlusion. 2.9.Infarct Volume ComparisonMD analogous to a more widely used apparent diffusion coefficient (ADC) was used to determine the true infarct volume with the automatic threshold of . A previously determined automatic IVIM infarct threshold of from leave one out cross validation was applied to calculate IVIM infarct volume.27 2.10.Statistical AnalysisLinear regression between IVIM and microsphere perfusion was analyzed and pooled across all physiologic states, and the coefficient of determination (), -values (), slopes (), offsets (), and corresponding Bland–Altman mean differences and 95% CIs are reported. A mixed effects model was applied with modality as the fixed effect and subject nested as the random effect. Tukey’s boxplots per physiologic state with mean, first and third quartiles, and whiskers were produced with hemispheric paired comparison. Statistical agreement within physiologic states was studied with mean paired difference and Wilcoxon signed rank, and sum squared residuals of the IVIM bi-exponential fits to Eq. (1) are reported. Lin’s concordance correlation coefficient (CCC) was used to determine agreement between multiple quantitative measures incorporating both accuracy and precision between two readings along the 45 deg line through the origin.28 In addition, CCCs variability across subjects was shown with the reported subject-wise CCC. All statistical significance was determined at the 5% () level for hypothesis testing. 3.ResultsA total of (mean age = 0.72y, mean weight = 25.1 kg, 5 female, 1 male) experiments were analyzed with further detail in Table 1. Representative images are shown in Fig. 2(a) for quantitative perfusion images during normocapnia, hypercapnia, and post-MCAO, in Fig. 2(b) for transit time post-MCAO, and in Fig. 2(c) for diffusion-positive maps post-MCAO. Average sum squared residuals of the IVIM signal fits (Table 1) show higher average error at hypercapnia. Table 1Of the six cases studied, the number of cases available for each physiologic state is shown in the top row, with all measurements within 30 min of each other. Two cases post-MCAO had slices removed from analysis due to zeros being returned by the microsphere vendors. Statistical analysis per physiologic state is shown in the middle row: mean paired difference and corresponding signed rank within each physiologic state, as well as the mean and standard deviation of the sum squared residuals for the voxels included in the analysis (D*>0.10) removed following historical threshold. The mixed effects model holding subjects as random variables is shown in the bottom row.
Figures 3(a) and 3(b) show linear regression, Lin’s CCC, and corresponding Bland–Altman of IVIM and microsphere perfusion. Tukey’s box plots in Fig. 3(c) demonstrate average and range of perfusion at normocapnia, hypercapnia, and post-MCAO split into ipsilateral and contralateral hemispheres. There were no significant differences in hypercapnia or post-occlusion hemispheres, but a statistically significant difference was seen at normocapnia (Table 1). The mixed effects model showed correlation existed holding subjects as random effects as well, supporting IVIM sensitivity to changing physiology (Table 1). Linear regression of middle cortical asymmetry of IVIM WTT post-occlusion returned correlation to DSC MTT [Fig. 4(a)] and Tukey’s Box plots show an expected increase in transit time on the ipsilateral hemispheres, though with a different mean and range of values [Fig. 4(b)]. Automatic infarct volume from IVIM returned strong linear correlation to MD (, , , ). Subject-wise Lin’s CCC for cases with all slices analyzed ranged from 0.26 to 0.86, with a mean of 0.56. The simulation of inversion recovery spin echo (TR = 5000 ms, TE = 37 ms) for CSF suppression of the remaining transverse signal from CSF and blood returned inversion time of 1850 ms to be the best time for CSF suppression. However, this optimal TI for CSF suppression also suppressed the blood signal to 18% of the original when both T1 and T2 are included. In the capillary geometry based analysis, we found linear regression (slope = 0.60, intercept = 57.2, , Lin’s CCC = 0.68 [0.59, 0.75]) with a Wilcoxon signed-rank statistic across all states equal to 308 and . BA analysis found a mean difference and 95% CI of compared to the water transit time model, which found mean difference and 95% CI of −11[−78, 54]. The WTT model demonstrates comparable sensitivity and an improved mean difference without the assumptions of capillary dimension in vessel-dependent models. 4.DiscussionThis study found that a 5-min IVIM 10 -value sequence can produce quantitative cerebral perfusion values via WTT in a clinically acceptable scan time across a range of relevant physiologic conditions. The sensitivity of IVIM perfusion metrics to physiologic changes in tissue suggest that IVIM images can have impact in the study and progression tracking of cerebrovascular disease.9 Furthermore, IVIM images are sensitive to changes in qCBF in response to inhalation (cerebrovascular reactivity), without repeated administration of the contrast agent, an advantage in the evaluation of chronic disease.10,14 Further, the study suggests a possible quantification factor that relies solely on assumed randomly oriented isotropic diffusing water transport and not on organ-dependent capillary segment length parameters. All comparisons between IVIM and microspheres across three physiologic states returned strong linear regression correlation and were further supported by the agreement shown with Lin’s CCC analysis. Prior studies compare IVIM cerebral perfusion to MR CBF measurements with varying levels of success.10,29,30 Several studies compare DSC CBV and IVIM ,10,29 and one study () compares relative IVIM CBF and relative DSC CBF.30 In comparison, this study found a strong correlation () of quantitative IVIM CBF in ml/100 g/min to neutron capture microsphere qCBF in ml/100 g/min. Mixed effects modeling showed that this sensitivity exists holding subjects as random effects as well (Table 1). This linear regression suggests sensitivity across physiologic states, and paired agreement within physiologic states supports the quantification of hemispheric perfusion without capillary parameter assumptions in a multi-day controlled pre-clinical model, which to the best of our knowledge has not been published previously. By showing correlation to an absolute quantitative method, the work supports a method that allows absolute comparison across multiple subjects, days, and experimental groups. In comparison, standard relative perfusion requires comparison to an internal reference that is confounded if the internal reference varies.22 However, the correlation also highlights an example of the remaining difficulty of CSF removal: partial volume contamination of CSF may falsely add to the perfusion signal at normocapnia, shown in Fig. 3(c), and lead to higher average perfusion with at normocapnia, though this study does not explore this. Although the standard IVIM model is bi-exponential, assuming -values from 222 to 1000 is mono-exponential, and -values are mono-exponential, the examination of tri-exponential behavior with more than just two fluid compartments is needed. Notably, although not explored in this paper, voxels with partial volume contamination such as those along the periphery of the cortex and close to the boundaries of the ventricles showed tri-exponential behavior, potentially from a three-compartment model. Further, the higher residuals at hypercapnia show increased perfusion causing a poorer bi-exponential fit, a potential factor in the lower sensitivity of IVIM to increased perfusion. WTT used to quantify perfusion showed linear correlation with delay and dispersion corrected DSC MTT asymmetry in the affected middle cortical territory post-MCA occlusion. In the interpretation of the DSC or CTP perfusion values clinically, time-based metrics used in bolus track perfusion calculation are often used due to the high sensitivity that these measures have to pathology. As IVIM and bolus-track DSC are very different approaches, the “water transit time” is derived to be an analog of MTT to study if IVIM based on fast water diffusion relates to MTT as determined by blood flow through a vascular bed as determined by DSC. The correlation of WTT and MTT provides another quantifiable metric to link the two methods for quantifying perfusion. Ii is important to note that comparison of IVIM WTT to DSC requires that DSC undergoes delay and dispersion correction, i.e., a local AIF, to account for IVIM AIF-independence. Without this correction of DSC, the correlation of qCBF and transit time would be weaker. The difference in transit time range observed is notable [Fig. 3(b)], likely due to the difference in properties being measured and endogenous versus exogenous measurement. Further, zeros are noted where is small as its inverse (WTT) is inversely proportionally high. To prevent influencing the results, these voxels are assigned 0 in Fig. 2(b), rather than being assigned an estimated WTT, and are not included in the analysis. The agreement of transit time supports the IVIM model with the assumption of instantaneous bolus mimicking the central volume principle through isotropic Gaussian diffusion if the DSC perfusion uses a local AIF. Previous studies have also reported positive correlation between ADC and IVIM in ischemic and normal regions (, )31 and (, ).32 Although this study’s agreement was lower than prior studies (slope = 1.01, ), it should be noted that the IVIM and DTI scans that were compared were taken from separate MR scans during a period of dynamic infarct expansion, whereas the previous studies used the same images to get ADC and IVIM . The use of two separate scans and sequences for IVIM and MD infarct volume was chosen to best compare IVIM and a separate standard clinical infarct volume. This combined with the IVIM qCBF from the same IVIM scan, not included in previous studies, demonstrates that the simultaneous perfusion and diffusion components from a single IVIM bi-exponential fit agreed with corresponding perfusion and diffusion reference standard values. IVIM imaging exploits the fact that diffusion-weighting is sensitive to water motion and consequently is able to identify where in the brain water is sequestered inside cells by cytotoxic edema.33 At very low -values (), diffusion-weighting is predominantly sensitive to faster water motion, such as capillary level blood flow.11,14,34–36 Questions remain regarding the origin of fast water motion signal, with Kwong et al.13 showing that the use of an inversion recovery pulse significantly reduced the cortical gray matter pseudo-diffusion fraction, which suggests that much of the bi-exponential behavior resulted from CSF contamination rather than perfusion.13 The simulation of inversion recovery spin echo suppression shows that it will suppress both CSF and blood from both T1 and T2 effects. Further, following the Monro–Kellie hypothesis, at hypercapnia, the dilation of cerebral capillaries prompts an increase in cerebral blood volume and CBF and a decrease in CSF to maintain intracranial pressure.37 Therefore, if signal in the fast-water compartment of the bi-exponential IVIM signal were due solely to CSF contamination, hypercapnia from inhalation should return a decrease in the fast-water compartment, which was not observed in this study [Fig. 3(c)]. Avoidance of voxels in the ventricles and subarachnoid space from diffusion maps, along with leave-one-out cross validation T2 map thresholds, reduced the effect of human bias in ROI selection. This study provides evidence that observed IVIM signal changes reflect cytotoxic edema and tissue perfusion. However, endogenous contrast prevents proof of signal origin, and partial volume effects require further study especially in disease states. Arterial spin labeling (ASL) is a well-established, widely available, and fully quantitative means of imaging cerebral perfusion noninvasively.38 However, with ASL, delayed arterial arrival times due to proximal vessel occlusion in stroke studies remains a challenge.39,40 As such, IVIM being independent of arterial input functions or bolus kinetics is a singificant benefit of the sequence in comparison with both ASL and DSC-MRI when measuring neurovascular disease. Futher, IVIM is able to provide simultaneous identification of cytotoxic edema, unlike ASL or DSC-MRI. This study is not without limitations. The highly dynamic physiology created challenges for the acquisition of normocapnia, hypercapnia, and post-MCAO with MRI and invasive neutron capture microspheres within tight timing windows. Two cases showing large disagreement between IVIM and microsphere perfusion also display differences in BP and between the two measurements that agree with the discrepancy in perfusion agreement that is shown [Fig. 3(b)], though the causality cannot be proven. This supports the physiologic change between MRI and microspheres that contribute to scatter in the linear regression. Difficulty in perfect alignment of physical brain cutting segmentation and discrete MR imaging also contributes to added noise in the linear regression. Although sensitivity to physiologic change is observed, IVIM sensitivity appears weaker than the perfusion effects observed with microspheres. Notably, normocapnia returns higher IVIM perfusion compared with microspheres, possibly due to partial volume contamination from CSF. The exponential fitting error, especially regarding pseudo-diffusion coefficient, is a continuing problem. Although limits were placed to maintain pseudo-diffusion coefficients in the expected range following the historical literature,30 pseudo-diffusion error will have large effects due to the inverse relation to WTT, and the error is observed to increase at hypercapnia. Further, some data were lost beyond our control due to vendor error and dynamic physiology, highlighting the complexity of microsphere perfusion. Finally, translatability of this perfusion to humans is not known. 5.ConclusionsDiffusion positive volumes and quantitative cerebral perfusion from IVIM agreed with reference standard values over a range of physiologic conditions. The sensitivity of IVIM provides evidence that observed signal changes reflect cytotoxic edema and tissue perfusion. IVIM images were acquired from 5 min of non-contrast DWI and quantified with WTT independent of capillary geometric assumptions and returned agreement to neutron capture microspheres across a wide range of physiologic states, such as normocapnia, hypercapnia, and infarction. This supports the further development and refinement of IVIM for measuring noninvasive quantitative perfusion and diffusion. 6.AppendixStep-by-step derivation of WTT, the time at which 50% of the original molecules have diffused with diffusion-coefficient beyond a unit boundary, is given. In other words, this is an equation for the concentration of a species to calculate the time taken for the initial concentration to be reduced by half. If particles begin as a delta function in the center of a voxel, the way tracer kinetics is modeled as an instantaneous bolus in DSC, following the law of diffusion the distribution of particles will change over time according to a normal distribution as a function of , where is the diffusion coefficient and is the time it has been diffusing. WTT can be solved by integrating over one unit of a three-dimensional Gaussian centered at zero and setting it equal to 0.50 as follows: The concentration over time will decrease as water diffuses out of the sphere with diffusion coefficient . Solving for WTT is shown below. For simplicity, is substituted for , and the integral is solved by integration by parts. The limits of integration are set from 0 to 0.5 mm because the instantaneous impulse function is set at the center of a unit volume sphere and returns the following: Plugging this into Eq. (4) returns: Due to the nature of , this is not analytically solvable. Therefore, the use of a numerical approximation of is required. The approximation of from equation (7.1.27) of Ref. 41 returns an error and is given as follows (Fig. 5): Plugging this in returns the following equation, which can be solved for : This can be checked by plugging into Eq. (4) as follows: Returning to Gaussian diffusion parameters, if we substitute , we can solve for and get . Solving this for WTT in terms of returns This is the same as if we solved for WTT in terms of straight from the approximation of , but in terms of of the the 3D Gaussian. With this WTT derived and explained intuitively, we can plug this into the central volume principle (blood flow = blood volume/transit time) to return the calibration coefficient for the fast-moving component of IVIM volume and pseudo-diffusion coefficient : It should be emphasized that this is an approximation that is valid when capillary perfusion direction is randomly oriented and isotropic and dominates the anisotropic signal. The water content fraction of a voxel , the density , the numerical estimation used for solving the , and the assumption of it being a sphere all contribute to the quantification factor above. This is the trade-off in assumptions compared with capillary bed network assumptions, such as capillary segment length, diameter, and path length, that are organ-dependent and vary under physiologic stress. Using the WTT based perfusion model to quantify IVIM perfusion returns the following: Code and Data AvailabilityThe post processing code was written in MATLAB and is stored in Github and readily available. All images are stored local in easily read DICOM format. Physiologic monitoring and drug administration are recorded in Microsoft Excel worksheets. These records are time-stamped to allow for cross referencing with image acquisition times (provided in the DICOM headers). The archive of all data are available by request from the corresponding author. AcknowledgmentsThis study was funded by the National Institute of Neurological Disorders and Stroke (R01-NS093908), National Science Foundation (DGE-1746045), and American Heart Association (GRNT-20380798). ReferencesD. Le Bihan et al.,
“MR imaging of intravoxel incoherent/t motions: application to diffusion and perfusion in neurologic disorders,”
Radiology, 161
(2), 401
–407 https://doi.org/10.1148/radiology.161.2.3763909 RADLAX 0033-8419
(1986).
Google Scholar
G. Zhu et al.,
“Comparison of MRI IVIM and MR perfusion imaging in acute ischemic stroke due to large vessel occlusion,”
Int. J. Stroke, 15
(3), 332
–342 https://doi.org/10.1177/1747493019873515
(2020).
Google Scholar
C. Federau et al.,
“IVIM perfusion fraction is prognostic for survival in brain glioma,”
Clin. Neuroradiol., 27
(4), 485
–492 https://doi.org/10.1007/s00062-016-0510-7
(2017).
Google Scholar
S. Fujiwara et al.,
“Feasibility of IVIM parameters from diffusion-weighted imaging at 11.7T MRI for detecting ischemic changes in common carotid artery occlusion rats,”
Sci. Rep., 10
(1), 8404 https://doi.org/10.1038/s41598-020-65310-8 SRCEC3 2045-2322
(2020).
Google Scholar
M. Bergamino et al.,
“Preliminary assessment of intravoxel incoherent motion diffusion-weighted MRI (IVIM-DWI) metrics in Alzheimer’s disease,”
J. Magn. Reson. Imaging, 52
(6), 1811
–1826 https://doi.org/10.1002/jmri.27272
(2020).
Google Scholar
A. M. Paschoal et al.,
“Intravoxel incoherent motion MRI in neurological and cerebrovascular diseases,”
Neuroimage Clin., 20 705
–714 https://doi.org/10.1016/j.nicl.2018.08.030
(2018).
Google Scholar
C. P. Derdeyn et al.,
“Variability of cerebral blood volume and oxygen extraction: stages of cerebral haemodynamic impairment revisited,”
Brain, 125
(3), 595
–607 https://doi.org/10.1093/brain/awf047 BRAIAK 0006-8950
(2002).
Google Scholar
C. P. Derdeyn, Jr. R. L. Grubb and W. J. Powers,
“Re: stages and thresholds of hemodynamic failure,”
Stroke, 34
(3), 589 https://doi.org/10.1161/01.STR.0000058161.89420.F4 SJCCA7 0039-2499
(2003).
Google Scholar
C. P. Derdeyn, Jr. R. L. Grubb and W. J. Powers,
“Indications for cerebral revascularization for patients with atherosclerotic carotid occlusion,”
Skull Base, 15
(1), 7
–14 https://doi.org/10.1055/s-2005-868159 SBSUEL 1052-1453
(2005).
Google Scholar
R. Wirestam et al.,
“Perfusion-related parameters in intravoxel incoherent motion MR imaging compared with CBV and CBF measured by dynamic susceptibility-contrast MR technique,”
Acta Radiol., 42
(2), 123
–128 https://doi.org/10.1080/028418501127346459
(2001).
Google Scholar
D. Le Bihan and R. Turner,
“The capillary network: a link between IVIM and classical perfusion,”
Magn. Reson. Med., 27
(1), 171
–178 https://doi.org/10.1002/mrm.1910270116 MRMEEN 0740-3194
(1992).
Google Scholar
E. K. Englund et al.,
“Intravoxel incoherent motion magnetic resonance imaging in skeletal muscle: review and future directions,”
J. Magn. Reson. Imaging, 55
(4), 988
–1012 https://doi.org/10.1002/jmri.27875
(2022).
Google Scholar
K. K. Kwong et al.,
“CSF-suppressed quantitative single-shot diffusion imaging,”
Magn. Reson. Med., 21
(1), 157
–163 https://doi.org/10.1002/mrm.1910210120 MRMEEN 0740-3194
(1991).
Google Scholar
C. Federau et al.,
“Quantitative measurement of brain perfusion with intravoxel incoherent motion MR imaging,”
Radiology, 265
(3), 874
–881 https://doi.org/10.1148/radiol.12120584 RADLAX 0033-8419
(2012).
Google Scholar
Y. I. Jeong et al.,
“Absolute quantitative MR perfusion and comparison against stable-isotope microspheres,”
Magn. Reson. Med., 81 3567
–3577 https://doi.org/10.1002/mrm.27669 MRMEEN 0740-3194
(2019).
Google Scholar
G. A. Christoforidis et al.,
“An endovascular canine middle cerebral artery occlusion model for the study of leptomeningeal collateral recruitment,”
Investig. Radiol., 46
(1), 34
–40 https://doi.org/10.1097/RLI.0b013e3181f0cbc7 INVRAV 0020-9996
(2011).
Google Scholar
B. B. Thomsen et al.,
“Spontaneous ischaemic stroke lesions in a dog brain: neuropathologicharacterizationion and comparison to human ischaemic stroke,”
Acta Vet. Scand., 59
(1), 7 https://doi.org/10.1186/s13028-016-0275-7
(2017).
Google Scholar
K. Zhang and T. J. Sejnowski,
“A universal scaling law between gray matter and white matter of cerebral cortex,”
Proc. Natl. Acad. Sci. U. S. A., 97
(10), 5621
–5626 https://doi.org/10.1073/pnas.090504197
(2000).
Google Scholar
R. J. Traystman,
“Animal models of focal and global cerebral ischemia,”
ILAR J., 44
(2), 85
–95 https://doi.org/10.1093/ilar.44.2.85
(2003).
Google Scholar
G. McHedlishvili and N. Kuridze,
“The modular organization of the pial arterial system in phylogeny,”
J. Cereb. Blood Flow Metab., 4
(3), 391
–396 https://doi.org/10.1038/jcbfm.1984.57
(1984).
Google Scholar
N. Saadat et al.,
“Influence of simultaneous pressor and vasodilatory agents on the evolution of infarct growth in experimental acute middle cerebral artery occlusion,”
J. Neurointerv. Surg., 13 741
–745 https://doi.org/10.1136/neurintsurg-2020-016539
(2020).
Google Scholar
M. Liu et al.,
“Augmentation of perfusion with simultaneous vasodilator and inotropic agents in experimental acute middle cerebral artery occlusion: a pilot study,”
J. Neurointerv. Surg., 15 e69
–e75 https://doi.org/10.1136/jnis-2022-018990
(2022).
Google Scholar
C. P. Reinhardt et al.,
“Stable labeled microspheres to measure perfusion: validation of a neutron activation assay technique,”
Am. J. Physiol. Heart Circ. Physiol., 280
(1), H108
–H116 https://doi.org/10.1152/ajpheart.2001.280.1.H108
(2001).
Google Scholar
Jr. J. G. Kingma, D. Simard and J. R. Rouleau,
“Comparison of neutron activated and radiolabeled microsphere methods for measurement of transmural myocardial blood flow in dogs,”
J. Thromb. Thrombol., 19
(3), 201
–208 https://doi.org/10.1007/s11239-005-1201-4
(2005).
Google Scholar
J. Ahmed, M. K. Pulfer and R. A. Linsenmeier,
“Measurement of blood flow through the retinal circulation of the cat during normoxia and hypoxemia using fluorescent microspheres,”
Microvasc. Res., 62
(2), 143
–153 https://doi.org/10.1006/mvre.2001.2321 MIVRA6 0026-2862
(2001).
Google Scholar
J. J. Mouannes-Srour et al.,
“Correction for arterial-tissue delay and dispersion in absolute quantitative cerebral perfusion DSC MR imaging,”
Magn. Reson. Med., 68
(2), 495
–506 https://doi.org/10.1002/mrm.23257 MRMEEN 0740-3194
(2012).
Google Scholar
M. Liu,
“Quantitative IVIM perfusion at normocapnia, hypercapnia, acute stroke and stroke treatment predicts neutron capture microspheres,”
in Proc. ISMRM Workshop Perfusion MRI;,
(2022). Google Scholar
L. I. Lin,
“A concordance correlation coefficient to evaluate reproducibility,”
Biometrics, 45
(1), 255
–268 https://doi.org/10.2307/2532051 BIOMB6 0006-341X
(1989).
Google Scholar
W. C. Wu et al.,
“Caveat of measuring perfusion indexes using intravoxel incoherent motion magnetic resonance imaging in the human brain,”
Eur. Radiol., 25
(8), 2485
–2492 https://doi.org/10.1007/s00330-015-3655-x
(2015).
Google Scholar
C. Federau et al.,
“Measuring brain perfusion with intravoxel incoherent motion (IVIM): initial clinical experience,”
J. Magn. Reson. Imaging, 39
(3), 624
–632 https://doi.org/10.1002/jmri.24195
(2014).
Google Scholar
S. Suo et al.,
“Stroke assessment with intravoxel incoherent motion diffusion-weighted MRI,”
NMR Biomed., 29
(3), 320
–328 https://doi.org/10.1002/nbm.3467
(2016).
Google Scholar
Q. Q. Gao et al.,
“Quantitative assessment of hyperacute cerebral infarction with intravoxel incoherent motion MR imaging: initial experience in a canine stroke model,”
J. Magn. Reson. Imaging, 46
(2), 550
–556 https://doi.org/10.1002/jmri.25556
(2017).
Google Scholar
M. Iima and D. Le Bihan,
“Clinical intravoxel incoherent motion and diffusion MR imaging: past, present, and future,”
Radiology, 278
(1), 13
–32 https://doi.org/10.1148/radiol.2015150244 RADLAX 0033-8419
(2016).
Google Scholar
D. Le Bihan,
“What can we see with IVIM MRI?,”
Neuroimage, 187 56
–67 https://doi.org/10.1016/j.neuroimage.2017.12.062 NEIMEF 1053-8119
(2019).
Google Scholar
O. Jalnefjord et al.,
“Comparison of methods for estimation of the intravoxel incoherent motion (IVIM) diffusion coefficient (D) and perfusion fraction (f),”
MAGMA, 31
(6), 715
–723 https://doi.org/10.1007/s10334-018-0697-5
(2018).
Google Scholar
D. Le Bihan et al.,
“Separation of diffusion and perfusion in intravoxel incoherent motion MR imaging,”
Radiology, 168
(2), 497
–505 https://doi.org/10.1148/radiology.168.2.3393671 RADLAX 0033-8419
(1988).
Google Scholar
L. A. van der Kleij et al.,
“Arterial pressure changes during hypercapnia are associated with changes in brain parenchymal volume,”
Eur. Radiol. Exp., 4
(1), 17 https://doi.org/10.1186/s41747-020-0144-z
(2020).
Google Scholar
J. A. Detre and D. C. Alsop,
“Perfusion magnetic resonance imaging with continuous arterial spin labeling: methods and clinical applications in the central nervous system,”
Eur. J. Radiol., 30
(2), 115
–124 https://doi.org/10.1016/S0720-048X(99)00050-9 EJRADR 0720-048X
(1999).
Google Scholar
S. Haller et al.,
“Arterial spin labeling perfusion of the brain: emerging clinical applications,”
Radiology, 281
(2), 337
–356 https://doi.org/10.1148/radiol.2016150789 RADLAX 0033-8419
(2016).
Google Scholar
Y. Yao et al.,
“Intravoxel incoherent motion diffusion-weighted imaging in stroke patients: initial clinical experience,”
Clin. Radiol., 71
(9), 938.e11
–938.e16 https://doi.org/10.1016/j.crad.2016.04.019
(2016).
Google Scholar
M. Abramowitz and I. A. Stegun, Handbook of Mathematical Functions with Formulas, Graphs, and Mathematical Tables, U.S. Government Publishing Office, Washington, DC
(1964). Google Scholar
BiographyMira Liu, PhD, was a National Science Foundation graduate research fellow in the graduate program in medical physics at the University of Chicago. She is now continuing work on advanced perfusion and diffusion MRI in clinical translational abdominal radiology as a postdoctoral fellow at the Icahn School of Medicine at Mount Sinai. Niloufar Saadat, MD, was a postdoctoral research scientist at the University of Chicago studying flow augmentation stroke therapeutics. She is now a clinical radiologist at the University of Texas Southwestern Medical Center. Yong Jeong, PhD, was a postdoctoral research fellow at the University of Chicago with a PhD in biomedical engineering from Northwestern. He worked on quantitative dynamic susceptibility contrast and is currently working in private industry. Steven Roth, MD, FARVO, is a Michael Reese endowed professor of anesthesiology and associate head for research at the University of Illinois. Marek Niekrasz, DVM ACLAM Diplomate, was a professor of surgery, professor of veterinary medicine, director of surgery and large animal studies, and director of the Institutional Animal Care and Use Committee at the University of Chicago. He has since retired. Mihai Giurcanu, PhD, is a research associate professor in the Department of Statistics Biostatistics Laboratory where he works in collaboration with scientists and radiologists in the Department of Radiology. |
Microspheres
Diffusion
Capillaries
Magnetic resonance imaging
Blood
Linear regression
Statistical analysis