Significance: Glioblastoma (GBM) is the most common and aggressive malignant brain tumor in adults. With a worldwide incidence rate of 2 to 3 per 100,000 people, it accounts for more than 60% of all brain cancers; currently, its 5-year survival rate is <5 % . GBM treatment relies mainly on surgical resection. In this framework, multimodal optical spectroscopy could provide a fast and label-free tool for improving tumor detection and guiding the removal of diseased tissues. Aim: Discriminating healthy brain from GBM tissues in an animal model through the combination of Raman and reflectance spectroscopies. Approach: EGFP-GL261 cells were injected into the brains of eight laboratory mice for inducing murine GBM in these animals. A multimodal optical fiber probe combining fluorescence, Raman, and reflectance spectroscopy was used to localize in vivo healthy and tumor brain areas and to collect their spectral information. Results: Tumor areas were localized through the detection of EGFP fluorescence emission. Then, Raman and reflectance spectra were collected from healthy and tumor tissues, and later analyzed through principal component analysis and linear discriminant analysis in order to develop a classification algorithm. Raman and reflectance spectra resulted in 92% and 93% classification accuracy, respectively. Combining together these techniques allowed improving the discrimination between healthy and tumor tissues up to 97%. Conclusions: These preliminary results demonstrate the potential of multimodal fiber-probe spectroscopy for in vivo label-free detection and delineation of brain tumors, and thus represent an additional, encouraging step toward clinical translation and deployment of fiber-probe spectroscopy. |
1.Introduction1.1.GlioblastomaGlioblastoma (GBM) is the most common and aggressive malignant brain tumor in adults, accounting for more than 60% of all cases.1 Current treatment options at diagnosis include surgical resection, radiation, and chemotherapy2 but—despite advances in this field—it remains largely incurable; its 5-year survival rate3 is . Extensive and complete surgical resection of GBM is difficult due to the high invasiveness of these tumors. However, multiple studies4,5 have demonstrated the importance of aggressive surgical resection—when possible—with trends toward better outcomes in those patients with a greater extent of resection. 1.2.Multimodal Optical SpectroscopyIn this framework, multimodal optical spectroscopy could provide a fast and label-free tool for improving tumor detection and guiding the removal of diseased tissues.6–8 In fact, laser-induced fluorescence9 spectroscopy of tissue native autofluorescent molecules has been widely used to discriminate normal and diseased tissues. This method has the ability to provide quantitative and real-time biochemical information related to tissue pathology and has the potential for minimally invasive diagnosis, considering the recent advancements in fiber-optic probes. Moreover, diffuse reflectance and Raman spectroscopy could provide additional information for improving the diagnosis of tumors. Diffuse reflectance10 employs a broadband light source and it is based on the ratiometric measurement of the spectrum of light back-scattered by the tissue, as compared with the illumination spectrum. The measured spectrum allows characterizing tissue chromophores and scatterers,11 which in turn are related to the physiologic conditions of the tissue under investigation. On the other hand, Raman spectroscopy12 is based on inelastic scattering processes in which non-resonant photons contribute to the excitation of vibrational molecular modes. The measurement of the energy lost in the process allows measuring a vibrational spectrum, typical of the specific molecule involved in the process. Hence, the multimodal combination of different spectroscopic techniques13,14 allows investigating tissue composition, either at structural, molecular, and metabolic level, without recurring to labels and in a relatively small amount of time (, or even less). Moreover, machine learning15,16 methods can be used for providing unbiased, operator-independent, and automated clinical/biomedical classification based on the spectral data collected with such approach. 1.3.In Vivo Detection of GBMIn this proof-of-concept study, we used multimodal fiber-probe spectroscopy for in vivo examination of mouse brain tumors induced by murine GBM cell line. In particular, we used label-guided fluorescence for detecting tumor areas and, then, Raman and reflectance spectroscopies for examining and classifying both healthy and diseased tissues. Finally, we developed a scoring algorithm based on principal component analysis (PCA) and linear discriminant analysis (LDA) for discriminating the two tissue types based on their spectral profiles. 2.Materials and MethodsThe main goal of this study was to explore the use of Raman and reflectance fiber-probe spectroscopies for in vivo detection of GBM in an animal model (i.e., laboratory mice). The following sections will describe animal preparation, our experimental setups, data collection, and analysis. 2.1.Animal ModelWe induced the development of a glioma tumor in laboratory mice in order to compare the spectra of healthy and diseased brain tissues, as described in the following paragraphs. All animal care and experimental procedures were performed in strict accordance with the recommendations of the Italian Ministry of Health (Dlg. 26/14) and according to protocol 143-2017 PR approved by the Ministry of Health on February 13, 2017. 2.1.1.Cell culture preparationThe experiment was conducted using EGFP-GL261 tumor cells, a murine glioma cell line with characteristics similar to human GBM and with stable expression of the green fluorescent protein (EGFP). GL261 cells were transfected with EGFP using the electroporation technique. The cells were detached using 0.05% trypsin-EDTA (Gibco), washed with 5 ml of PBS and centrifuged at 1200 rpm for 5 min; then, they were re-suspended in of resuspension buffer (Invitrogen) and of DNA were added. Afterward, the cells were electroporated with two 20-ms pulses at 1200 V using Neon Transfection System pipette tips (ThermoFisher Scientific). The amount of DNA used is around 10 to for each electroporation. The transfected cells were plated and maintained in Dulbecco’s modified Eagle’s medium (DMEM, Gibco) containing 10% fetal bovine serum, 1% penicillin–streptomycin ( of penicillin and of streptomycin, Gibco), 1 mM sodium pyruvate (Gibco) and 10 mM HEPES (Gibco) in a humidified atmosphere of 5% . The culture medium was replaced with fresh medium 2 to 3 times per week. Once the confluence reaches to about 90%, the cells were harvested using 0.05% trypsin-EDTA. The cells were pelleted by centrifugation at 1200 rpm for 5 min (Sorvall T3 Centrifuge, ThermoScientfic). The pellet was re-suspended with PBS and the resulting solution was injected into the mouse brain using a glass capillary. 2.1.2.Mice preparationAdult C57BL/6 mice were maintained at a mice facility with controlled temperature and light, where they received food and water ad libitum. The animals were anesthetized with Avertin ( body weight) by intraperitoneal injection. Hair at the top of the head was removed using a hair depilatory cream and fixed on the stereotaxic stage. A single skin cut of to 1 cm was made using a pair of scissors and the exposed sub-cutaneous tissue was gently removed. A hand drill was used to thin the skull at the injection site, which is the visual cortex. Two microliters of EGFP-GL261 cell suspension containing in total were aspired into the glass capillary and slowly injected into the mouse cortex. The glass capillary was inserted at at first and then retracted before injection of cell suspension. A total of 5 min was spent for the injection and additional 2 min before removing the glass capillary. The surface of the injection site was then washed with PBS and skin was closed with simple stitches. Iodine antiseptic cream was applied on the stitches to prevent infection. To relieve pain, 5 drops of paracetamol () per 100 ml of drinking water was provided for the animals after surgery. All the instruments used were sterilized. Another surgery was carried out under anesthesia induced by Avertin ( body weight) for placing a cranial window. The mice were placed on a heating pad and fixed on the stereotaxic stage. A longitudinal incision of the skin was performed between the occiput and the forehead, followed by skin and sub-cutaneous tissue removal around the injected site. Using a hand drill, the skull was thinned to open a 5-mm-diameter window. Forceps were used to break and remove the skull gently. The dura mater was continuously rinsed with artificial cerebro-spinal fluid: 125 mM NaCl, 25 mM , 1 mM , 2.5 mM KCl, 1.25 , 2 mM , and 10 mM glucose. A coverslip (5-mm diameter and 0.17-mm thickness) was used to cover the window and fixed using strong adhesive glue and self-curing dental cement carefully. We chose a quartz coverslip because, when excited at 785 nm, standard glass emits fluorescence at wavelengths corresponding to the “fingerprint region” of Raman spectroscopy. Again, all the instruments used were sterilized. Finally, another cranial window was placed at a certain distance from the injection site, in order to have an optical access to healthy brain tissue. 2.1.3.Optical measurementsEight animals were included in this study. After two weeks from the injection of EGFP-GL261 cells, each mouse was anesthetized and placed under the stereotaxic device (Fig. 1) for examination through our optical fiber probe (Sec. 2.3.1), whose distal end was placed perpendicularly in contact with each of the two quartz windows. Few animals were inspected also in the following days, in order to record tumor progression; in total, we performed 13 experiments. First, fluorescence spectroscopy was used for real-time () detection of tumor areas by taking advantage of the characteristic EGFP signal. The probe was placed in the point of maximum EGFP intensity; then, fluorescence, Raman, and reflectance spectra were acquired from the same region. Spectral measurements were performed also on the second optical window (i.e., on control brain tissue). Total integration times were selected in order to exploit the dynamic range of the CCD and to reduce noise. For fluorescence acquisitions, they ranged between 0.5 and 5 s—depending on the presence of EGFP and the amount of autofluorescence—obtained by averaging 10 recordings and using 5- and 3-mW excitation power at 378 and 445 nm, respectively. Each Raman measurement required averaging 10 recordings, resulting in 25- to 50-s integration times using 75-mW power. Reflectance spectra were acquired by averaging 100 recordings in 0.1 to 5 s, using typical output power (from a nominal bulb power of ). The inspection typically required a total of less than 4 min for each animal, resulting in 1 spectrum for each spectroscopic technique and for each region. Tumor areas were later examined through a multiphoton microscope (Sec. 2.3.2) in order to observe EGFP-labelled cells and to estimate tumor size. 2.2.Pellet Preparation and MeasurementsWe investigated the presence of EGFP spectral contributions in Raman and reflectance measurements by examining two pellets with our fiber probe: one made of wild-type (WT) GL261 cells and the other one of EGFP-GL261 cells. Each pellet was produced starting from two 100-mm tissue culture dishes containing GL261 cells that reached 100% confluence. The medium was discarded from the dishes and then they were first washed with Dulbecco phosphate-buffered saline (DPBS) without calcium and magnesium (Gibco 14190136). Trypsin-EDTA 0.05% (Gibco 25300054) was added to the dishes, and then cells were incubated for 2 to 3 min at 37 °C. After that, trypsin was blocked by adding cell culture medium and cells were resuspended to a single-cell suspension by pipetting with a 1000-ml pipette. Then the cells were transferred to a 15-ml conical tube and centrifuged at . Then the pellet was measured in a 4% paraformaldehyde solution. 2.3.Experimental Setups2.3.1.Fiber-probe systemSpectral measurements were performed through the experimental setup described in Ref. 6. As shown in Fig. 1, this system consists of four excitation sources, a quadri-furcated fiber probe (EMVision LCC, Loxahatchee, Florida), and a spectrograph (MicroHR, HORIBA Scientific, Edison, New Jersey) equipped with a 600-lines/mm grating and with an open-electrode CCD camera (Syncerity, HORIBA Scientific, Edison, New Jersey) cooled at . In particular, we used two laser diodes (TEC 042, Sacher Lasertechnik, Maburg, Germany) for exciting fluorescence at 378 and 445 nm, respectively; a narrow-band laser diode at 785 nm (FC-785-350-MM2-PC-1-0-RM, RGBLase, Fremont, California) for near-infrared (NIR) Raman spectroscopy; and a halogen lamp (HL-2000-LL, Ocean Optics, Dunedin, Florida) for reflectance spectroscopy. The probe, which is connected to all light sources, consists of a bundle of 11 optical fibers: a central fiber ( core diameter) for delivering 785 nm; 7 surrounding fibers ( each) for Raman collection; 2 fibers ( each) for delivering 378/445 nm and white-light from the lamp, respectively, and 1 fiber () for collecting both fluorescence/reflectance signals, all of them placed near to the central fiber. A narrow bandpass filter centered at 785 nm was placed in front of the Raman delivering fiber in order to clean the laser line, blocking unwanted fluorescence and Raman signals generated inside the optical fiber itself. A ring-shaped long-pass optical filter was placed at the distal end of the probe, preventing 785-nm laser light from entering the seven collection fibers. A motorized filter wheel allows changing emission filter according to the spectral technique used: a long pass filter at 400 nm (Longpass 400 nm, Melles Griot, Albuquerque, New Mexico) for fluorescence spectroscopy excited at 378 nm; a long pass at 458 nm (LP02-458RS-25, Semrock, Rochester, New York) for fluorescence spectroscopy excited at 445 nm; a long pass at 785 nm (LP02-785RE-25, Semrock) for Raman spectroscopy; and a neutral density filter with for diffuse reflectance. These filters prevent laser light from entering the monochromator and lamp light from saturating the detector. The device was designed to be easily handled (as a pen) for fast examination of suspicious tissue areas. 2.3.2.Two-photon excited fluorescence microscopeHigh-resolution two-photon excited fluorescence (TPEF) in vivo imaging was performed on a Prairie Ultima Multiphoton microscope equipped with a mode-locked Ti:sapphire laser (Chameleon Ultra II, Coherent, Santa Clara, California). Laser excitation was set at 910 nm for EGFP detection and a maximum power of was used for data acquisition. Images were acquired with a water immersion lens (XLUMPLFLN20XW, Olympus, Tokyo, Japan; magnification, ) at resolution and zoom = 1, leading to field of view with linear resolution of . 2.4.Data Analysis2.4.1.Pre-processing of spectraAbout 26 spectra—13 from healthy mice brains and 13 from GBM tissues—were recorded for each optical technique (fluorescence excited at 378 and 445 nm, Raman, reflectance). The transmission spectrum of the fiber-probe setup was used for correcting all the recorded spectra in order to remove artifacts. Also, further processing was performed on Raman and reflectance datasets. In particular, we used a sixth-order polynomial fitting through an automated routine (Vancouver Raman algorithm17) for removing NIR fluorescence and background signals from the Raman spectra, which were later normalized to the Raman band (corresponding to and deformations18) intensity. Reflectance spectra were corrected for the lamp emission spectrum and normalized to the intensity recorded at 570 nm, which is one of the oxy/deoxy-hemoglobin isosbestic points. 2.4.2.Spectral classification analysisThis study aims to discriminate healthy brain and glioma tissues in a label-free modality, e.g., based on the information recorded through Raman and reflectance spectroscopies. In this framework, PCA and LDA are well-known methodologies19–21 used for the classification of spectral data;6,22,23 we implemented each method in a separate custom-made routine on LabVIEW (National Instruments, Austin, Texas). PCA was used for reducing noise and redundant information by selecting four principal components (PCs) of each dataset and discarding all the others. Also, a decreased dimensionality reduces the risk of overfitting the recorded data and, thus, overestimating classification accuracy. LDA, then, was employed to build a linear combination of the selected PCs in order to maximize discrimination between different tissue classes. We specifically designed an LDA routine for generating a classifier and maximizing the area under its receiver operating characteristic curve (auROCc), a parameter useful for estimating the classification accuracy.24,25 For each spectrum, the algorithm creates a linear combination by multiplying each selected PC loading by a random number between 0 and 100, then summing all the four products into a score. The ROC curve of normal and tumor scores and the corresponding auROCc values are then calculated and recorded. The algorithm iterates such routine, each time comparing the new auROCc value with the maximum value among those ones already calculated, and stops after consecutive iterations (e.g., ) where no further maximum is found. Finally, the linear combination providing the best accuracy is saved and can be used to classify a group of unknown spectra. A cross-validation function can be selected for dividing the dataset in ()-folds, using ()-folds for training the classification algorithm and 1 fold for testing it, and repeating the process ()-times by rotating the folds in the two groups. Two additional analytical methods were applied to the recorded spectra: ratiometric scoring (i.e., calculating the ratio between photon counts recorded at two different wavelengths/wavenumbers) and slope analysis (i.e., applying a linear fit to a region of the spectrum in order to derive its slope). Both analyses were implemented in LabVIEW. 2.4.3.Image analysisTPEF images were reconstructed and analyzed through ImageJ (National Institutes of Health, Bethesda, Maryland). The radial (i.e., parallel to brain surface) extension of each tumor was estimated by measuring the average distance between opposite margins. Instead, tumor axial (i.e., perpendicular to brain surface) extension could not be estimated due to the limits in NIR light penetration (and visible fluorescence back-propagation) in highly scattering media such as brain tissue; roughly, the microscope was not able to image deeper than . 3.Results3.1.EGFP Fluorescence, Raman, and Reflectance SpectraWe used the fiber-probe system for spectroscopic examination of both WT- and EGFP-GL261 pellets. Figure 2 shows our results. As expected, UV- and blue-excited fluorescence spectra of the EGFP pellet [Figs. 2(a) and 2(b), respectively] are dominated by the label emission peak at 510 nm; instead, autofluorescence signal is relatively negligible. Moreover, we found stronger EGFP signal in blue-excited spectra, as their excitation wavelength (445 nm) is closer to the maximum absorption peak (490 nm) of the protein. Raman spectra [Fig. 2(c)] are mainly characterized by the phenylalanine bands26,27 at 1000, 1030, and , showing no significant differences between WT and EGFP cell pellets. Instead, reflectance spectra have different trends in the 450 to 500 nm and 500 to 540 nm ranges—corresponding, respectively, to EGFP absorption and emission—while the two signals show good correspondence at longer wavelengths and mostly overlap. Hence, we found that EGFP presence does not affect Raman measurements through our setup, but it may do so in reflectance recordings. Although the effects of the fluorescent label are not clearly visible in tissue reflectance acquisitions [Fig. 3(g)], we decided to restrict the analysis of those spectra (Sec. 3.2) in the 540 to 733 nm range. 3.2.Brain Tumor Detection in Mice3.2.1.Tumor localizationTumor areas in mice brain were detected through their EGFP emission peak excited at 378 and 445 nm [Figs. 3(a) and 3(b), respectively]. In fact, after placing each animal on the stereotaxic stage, we continuously acquired fluorescence spectra while moving the fiber probe along the tumor window; then, we positioned the probe where the fluorescent label signal was maximized. We found that EGFP increased day by day, as tumor size increased over time; also, we recorded the autofluorescence spectrum of healthy brain tissues from the center of the other quartz window for a qualitative comparison against spectra of EGFP-labelled tumor. However, only a small number (maximum 3) of time points was recorded for each animal, due to the limited time window—few days—available for measurements before its death; moreover, in order to reduce the stress for the animal, a 1-day pause was interposed between every two sessions. Tumor radial extension was correlated with fluorescence intensity excited at 445 nm and emitted at 510 nm, corresponding to the EGFP emission peak [Fig. 4(c)]. These variables seem to be related (, Pearson’s for a linear fit): normal dots are clustered in one small area, while tumor dots follow an approximately linear trend. However, there are few considerations to be made. In the first place, peak intensity is calculated relatively to tissue autofluorescence emitted at 660 nm; thus, biological variation may introduce a certain degree of “noise” in the reported values. Second, the intensity of EGFP emission depends on the spatial convolution between tumor volume (i.e., its radial and axial extensions) and the excitation volume of the probe. Since the radial laser spot is , fluorescence intensity for tumors larger than that size depends only on their axial extension; it can be expected that, in such cases, EGFP intensity increases with tumor depth. Then, if radial and axial extensions grow proportionally together, this is a possible explanation for seeing a linear trend in Fig. 4(c). 3.2.2.Raman resultsWe recorded two Raman spectra from each animal inspection, i.e., from both healthy and tumor brain regions. Figure 3(d) shows the average spectra of these tissue types, which are characterized by the typical Raman peaks of amino acids (such as proline and valine bands18 at ), nucleic acids28 (1060 and ), lipids (), amide III and tryptophan29 (), and amide I27 (). By normalizing spectral intensity to the band, the major differences seem to reflect higher amino acids, tryptophan, and amide III relative (i.e., with respect to the lipid content) concentrations in healthy brain, and higher amide I relative concentration in tumor tissues. These findings could be partially explained with the altered cell metabolism in gliomas, which causes an increased rate of lipid synthesis during their growth30 and lower levels of amino acids.31 A classifier based on ratiometric scoring resulted in 77% and 79% auROCc (data not shown) by calculating the ratio between Raman photon counts registered at the and, respectively, 985- and bands. The first 15 PCs of the Raman dataset were calculated through standard PCA. We selected four of them, based both on the discrimination capability of each one (evaluated through its auROCc) and the improvement provided to the PCA-LDA analysis (again, measured as increased/decreased auROCc value). A classification model was generated from the linear combination of PC3, PC7, PC10, and PC13; the resulting classifier was denoised and reported in Fig. 3(d) in order to show the correlation between its major peaks and the observable differences between normal and tumor spectra. Finally, each Raman spectrum was assigned a “score” [Fig. 3(e)] based on the classifier, whose auROCc was found to be 92% [Fig. 3(f)] when training the algorithm on all dataset spectra. Then, in order to validate this result, we performed an automated 4-fold cross-validation (4f CV) analysis—10 normal and 10 tumor samples for training, 3 normal and 3 tumor samples for testing—resulting in 86% average auROCc (see Table 1). Table 1Specificity/sensitivity and auROCc values obtained using the classification models based on PCA-LDA. First and second rows show the results from the separate classification of Raman and reflectance datasets, respectively, while the third row reports their combined analysis. The last column on the right reports auROCc values obtained through 4f CV.
3.2.3.Reflectance resultsFigure 3(g) shows the average reflectance spectra recorded from normal murine brain and GL261 tumor areas. Each spectrum was normalized to an isosbestic point of oxy- and deoxy-hemoglobin (570 nm) in order to take advantage of their different molar extinction coefficients, to qualitatively compare their relative abundance in the two examined tissue types and, hence, to derive information on the corresponding oxygen levels. In fact, one of the most characteristic features of these spectra regards oxy-hemoglobin absorption peaks32 (540 and 575 nm), indicating that such substance is one of the major absorbers. Normal and tumor spectra intersect each other also in the others’ isosbestic points33 (452, 529, 545, and 584 nm), with the only exception of 500 nm. It is unclear whether the presence of EGFP may be partially responsible for missing the latter intersection, because a similar behavior can be observed in Ref. 14 between label-free normal and cancer brain tissue spectra. The major differences in Fig. 3(g) take place between 600 and 730 nm: murine gliomas show higher relative absorption in this region, suggesting a higher concentration of deoxy-hemoglobin with respect to oxy-hemoglobin, thus lower oxygenation with respect to normal brain. Indeed, this is consistent with the typical hypoxic environment34 of tumors. These patterns are so prominent that a classifier based on ratiometric scoring resulted in 82% auROCc (data not shown) by calculating the intensity ratio between 570 and 690 nm. Another feature can be observed at longer wavelengths (), where healthy spectra have slightly higher slope than tumors. It is known35 that cell nuclei dimension affects reflectance spectra in this region, specifically with respect to the reduced scattering coefficient (which is inversely proportional36 to the square root of scatterers diameter). Thus, as reflectance intensity is an approximately linear37 function of such coefficient, the ratio between normal and tumor spectral slopes could reflect their different nuclear size. We implemented a linear fitting algorithm in LabVIEW for deriving normal and tumor slopes between 625 and 730 nm [Fig. 4(f)], finding a normal-to-tumor slope ratio of . Formally, this value is consistent with typical tumor-to-normal root square nuclear size ratio (for example, we calculated a ratio from data reported in Ref. 38), although the reported standard error is relatively high. A classifier based on slope values resulted in 73% auROCc. Finally, the same PCA-LDA methodology used for the analysis Raman spectra was applied to the reflectance dataset. The first 15 PCs were calculated through standard PCA, then 4 of them (PC1 to PC3 and PC5) were used to generate a classification model. The PCA-LDA classifier [reported, after denoising, in Fig. 3(g)] reached 93% auROCc [Table 1, Figs. 3(h) and 3(i)] when training the algorithm on all dataset spectra; a 4f CV resulted in almost the same accuracy (92%). Moreover, we evaluated the correlation between our model and tumor diameter [Fig. 4(d)] or EGFP intensity [Fig. 4(e)]. Using a linear fit, a positive correlation was observed in both cases (Pearson’s and 0.552, respectively); however, the regression line did not result in good fitting of the data ( and 0.305). Thus, the classifier based on reflectance spectroscopy seems capable of detecting the tumor and potentially sensitive to its diameter, although more experiments and training of the model would be required to quantitatively evaluate its size. 3.2.4.Multimodal resultsWe reached high sensitivity (92%) in classifying Raman spectra and high specificity (100%) for reflectance ones. Since Raman and reflectance spectroscopies provide complementary information, combining their results could improve tissue characterization. Thus, the scores resulting from their classification models were linearly combined through the same LDA routine in order to maximize discrimination between normal and tumor areas. We found a sensible increase in auROCc value (97% from the whole dataset and 95% from the cross-validation), while the suggested threshold corresponded to 92% sensitivity and specificity. Tumor multimodal scores were also correlated to GBM size and EGFP signal, but the results from both linear fits were worse than the corresponding analyses on reflectance spectra. Such outcome reflected the fact that Raman scores were totally uncorrelated to tumor dimensions, hence their contribution to multimodal scoring reduced the correlation observed from reflectance data. 4.DiscussionSurgical resection plays a major role in current GBM treatment, hence the necessity of improving the detection of brain tumor areas in a fast and reliable way is key to a positive prognosis. Multimodal optical spectroscopy could be a practical tool for such purpose, as demonstrated in this preliminary study, where we employed in vivo Raman and reflectance spectroscopies in order to discriminate healthy brain from tumor tissue in a mouse GBM model. This approach led to the classification of healthy and tumor tissues in a label-free modality, based only on their intrinsic molecular composition (Raman) and oxygenation (reflectance). Our findings suggest that these spectroscopic techniques are able to recognize the altered metabolism of mouse glioma and its hypoxic environment. In particular, Raman spectroscopy could be used to automatically evaluate lipids and amino acids concentrations from the analysis of the fingerprint region, as they seem to be important biomarkers for brain tumor recognition. It should be noted that these results are consistent with 10 years of successful Raman applications7,39–42 (both ex vivo43 and in vivo14) to the detection of brain tumors, of whom many studies involving handheld fiber probes such as ours and involving similar numbers of patients/subjects. For example, Jermyn et al.39 and Jermyn et al.14 used fiber-probe Raman spectroscopy for in vivo detection of human brain tumors from 17 and 15 patients, respectively, obtaining auROCc. More generally, Zhang et al.42 summarized the results from 6 studies—mostly ex vivo—performed between 2005 and 2015, concluding that Raman spectroscopy provided 98% auROCc for glioma discrimination. While there are similarities between the presented study and previous works, there are also notable differences regarding number of acquisitions per subject (e.g., 10 or more spectra, instead of a single one) and detection instrumentation (i.e., CCDs with higher sensitivity and cooled down to lower temperatures), which may explain the comparatively lower signal-to-noise ratio of our spectra. Another major difference regards the analytical methods adopted, which can strongly affect classification accuracies; for instance, we found the analysis of Raman spectra to be sensibly dependent on the criteria for PCs selection. Finally, the presented experiment was designed differently from the known literature, as EGFP-labelling was used in order to precisely localize murine GBM areas and to record their Raman signatures. In our view, this approach could be followed in other oncologic studies based on animal models. Reflectance spectroscopy has been used for glioma detection, too, either in single modality44,45 or multimodal combination11,13,14 with fluorescence and/or Raman. Some studies, for example, used optical parameters derived from reflectance spectra (e.g., attenuation, absorption, and scattering coefficients) in order to correct the corresponding normal and tumor fluorescence spectra. Jermyn et al.14 improved in vivo brain cancer detection up to 98% auROCc by analyzing both Raman and attenuation-corrected autofluorescence spectra. Similarly, Valdes et al.11 applied absorption/scattering corrections to fluorescence spectra collected in vivo from 10 patients; notably, a fluorescence biomarker (PpIX) was used in order to provide contrast between control and tumor areas, resulting in 94% auROCc from the analysis of corrected fluorescence recordings. Other works, instead, discriminated brain tumors by directly analyzing reflectance data. Lin et al.45 analyzed reflectance spectra collected from 12 patients, obtaining 77% auROCc in control/tumor discrimination by looking at the normalized intensity recorded at 820 nm. Du Le et al.13 examined ex vivo samples excised from 7 patients, obtaining 100% sensitivity and 80% specificity in discriminating GBM from low-grade gliomas by looking at the intensity recorded at 650 nm. Therefore, the results presented in this study are consistent with the known literature. In fact, we successfully applied reflectance spectroscopy for probing tissue oxygenation and, also, deriving information on scatterers’ size. Moreover, the stark and label-free contrast between normal and diseased reflectance spectra allowed developing a classification algorithm with high accuracy. Further investigation of brain tissues could involve measuring absolute absorption, since tumors typically stimulate the formation of new blood vessels (angiogenesis), but the methodology applied in this work was not enough standardized to perform such a task. The use of a halogen lamp as excitation source represents a simpler and cheaper solution with respect to laser-based techniques; moreover, both detection system and data analysis could be eventually simplified into a narrow-band ratiometric approach. Considering the high speed of these spectral recordings and its relatively low implementation cost, such technique offers an enormous translational potential as it fits with all the requirements for a clinical deployment. However, the known literature and this study suggest that a multimodal approach—e.g., combining reflectance with laser spectroscopy—is crucial for improving sensitivity and specificity in tumor detection beyond the capability of each single technique. Further, multimodal spectroscopy could be exploited for scanning large brain areas upon implementation within real-time fiber probes46 or imaging systems.47 In conclusion, the in vivo results presented here represents an additional, although preliminary, encouraging step toward clinical translation and deployment of Raman and reflectance spectroscopies for a fast, label-free detection, and delineation of human brain tumors. DisclosuresThe authors declare that there are no conflicts of interest regarding the publication of this article. They also acknowledge that the presented results were partially disclosed in prior proceedings papers.48,49 AcknowledgmentsThe research leading to these results has received funding from Tuscany Region in the framework of the project GLIOMICS (research program PAR FAS 2007-2013, Bando Salute 2014), from EU Horizon 2020 research and innovation program (Grant H2020-ICT-2016-1 732111 PICCOLO and Grant 654148 Laserlab-Europe), and from Ente Cassa di Risparmio di Firenze (private foundation). ReferencesF. Hanif et al.,
“Glioblastoma multiforme: a review of its epidemiology and pathogenesis through clinical presentation and treatment,”
Asian Pac. J. Cancer Prev., 18
(1), 3
–9
(2017). https://doi.org/10.22034/APJCP.2017.18.1.3 Google Scholar
M. E. Davis,
“Glioblastoma: overview of disease and treatment,”
Clin. J. Oncol. Nurs., 20
(5), S2
–S8
(2016). https://doi.org/10.1188/16.CJON.S1.2-8 Google Scholar
H. Xu et al.,
“Geographic variations in the incidence of glioblastoma and prognostic factors predictive of overall survival in US adults from 2004–2013,”
Front. Aging Neurosci., 9 352
(2017). https://doi.org/10.3389/fnagi.2017.00352 Google Scholar
D. Kuhnt et al.,
“Correlation of the extent of tumor volume resection and patient survival in surgery of glioblastoma multiforme with high-field intraoperative MRI guidance,”
Neuro-Oncology, 13 1339
–1348
(2011). https://doi.org/10.1093/neuonc/nor133 Google Scholar
C. Roder et al.,
“Maximizing the extent of resection and survival benefit of patients in glioblastoma surgery: high-field iMRI versus conventional and 5-ALA-assisted surgery,”
Eur. J. Surg. Oncol., 40 297
–304
(2014). https://doi.org/10.1016/j.ejso.2013.11.022 Google Scholar
E. Baria et al.,
“Label‐free grading and staging of urothelial carcinoma through multimodal fibre‐probe spectroscopy,”
J. Biophotonics, 12
(11), e201900087
(2019). https://doi.org/10.1002/jbio.201900087 Google Scholar
J. Desroches et al.,
“Development and first in‐human use of a Raman spectroscopy guidance system integrated with a brain biopsy needle,”
J. Biophotonics, 12
(3), e201800396
(2019). https://doi.org/10.1002/jbio.201800396 Google Scholar
S. Anand et al.,
“Multimodal fiber-probe spectroscopy allows detecting epileptogenic focal cortical dysplasia in children,”
J. Biophotonics, 10
(6–7), 896
–904
(2017). https://doi.org/10.1002/jbio.201600136 Google Scholar
C. Croce and G. Bottiroli,
“Autofluorescence spectroscopy and imaging: a tool for biomedical research and diagnosis,”
Eur. J. Histochem., 58
(4), 2461
(2014). https://doi.org/10.4081/ejh.2014.2461 Google Scholar
S. Kawauchi et al.,
“Diffuse light reflectance signals as potential indicators of loss of viability in brain tissue due to hypoxia: charge-coupled-device-based imaging and fiber-based measurement,”
J. Biomed. Opt., 18
(1), 015003
(2013). https://doi.org/10.1117/1.JBO.18.1.015003 JBOPFO 1083-3668 Google Scholar
P. A. Valdes et al.,
“Combined fluorescence and reflectance spectroscopy for in vivo quantification of cancer biomarkers in low and high-grade glioma surgery,”
J. Biomed. Opt., 16
(11), 116007
(2011). https://doi.org/10.1117/1.3646916 JBOPFO 1083-3668 Google Scholar
M. Schmitt and J. Popp,
“Raman spectroscopy at the beginning of the twenty-first century,”
J. Raman Spectrosc., 37 20
–28
(2006). https://doi.org/10.1002/jrs.1486 JRSPAF 0377-0486 Google Scholar
V. N. Du Le et al.,
“Dual-modality optical biopsy of glioblastomas multiforme with diffuse reflectance and fluorescence: ex vivo retrieval of optical properties,”
J. Biomed. Opt., 22
(2), 027002
(2017). https://doi.org/10.1117/1.JBO.22.2.027002 JBOPFO 1083-3668 Google Scholar
M. Jermyn et al.,
“Highly accurate detection of cancer in situ with intraoperative, label-free, multimodal optical spectroscopy,”
Cancer Res., 77
(14), 3942
–3950
(2017). https://doi.org/10.1158/0008-5472.CAN-17-0668 CNREA8 0008-5472 Google Scholar
S. Webb,
“Deep learning for biology,”
Nature, 554
(7693), 555
–557
(2018). https://doi.org/10.1038/d41586-018-02174-z Google Scholar
J. A. M. Sidey-Gibbons and C. J. Sidey-Gibbons,
“Machine learning in medicine: a practical introduction,”
BMC Med. Res. Methodol., 19 64
(2019). https://doi.org/10.1186/s12874-019-0681-4 Google Scholar
J. Zhao et al.,
“Automated autofluorescence background subtraction algorithm for biomedical Raman spectroscopy,”
Appl. Spectrosc., 61 1225
–1232
(2007). https://doi.org/10.1366/000370207782597003 APSPA4 0003-7028 Google Scholar
M. Pinto et al.,
“Integration of a Raman spectroscopy system to a robotic-assisted surgical system for real-time tissue characterization during radical prostatectomy procedures,”
J. Biomed. Opt., 24
(2), 025001
(2019). https://doi.org/10.1117/1.JBO.24.2.025001 JBOPFO 1083-3668 Google Scholar
R. Bro and A. K. Smilde,
“Principal component analysis,”
Anal. Methods, 6 2812
–2831
(2014). https://doi.org/10.1039/C3AY41907J AMNEGX 1759-9679 Google Scholar
D.-Y. Tzeng and R. S. Berns,
“A review of principal component analysis and its applications to color technology,”
COLOR Res. Appl., 30 84
–98
(2005). https://doi.org/10.1002/col.20086 CREADU 0361-2317 Google Scholar
Q. Mai,
“A review of discriminant analysis in high dimensions,”
WIREs Comp. Stats., 5
(3), 190
–197
(2013). https://doi.org/10.1002/wics.1257 Google Scholar
N. V. Chernomyrdin et al.,
“Principle component analysis and linear discriminant analysis of multi-spectral autofluorescence imaging data for differentiating basal cell carcinoma and healthy skin,”
Proc. SPIE, 9976 99760B
(2016). https://doi.org/10.1117/12.2237607 Google Scholar
N. H. Othman et al.,
“Linear discriminant analysis for detection of salivary NS1 from SERS spectra,”
in IEEE TENCON 2017,
2876
–2879
(2017). https://doi.org/10.1109/TENCON.2017.8228352 Google Scholar
W. Zhu, N. Zeng and N. Wang,
“Sensitivity, specificity, accuracy, associated confidence interval and ROC analysis with practical SAS® implementations,”
in NESUG 201,
1
–9
(02010). Google Scholar
F. Li and H. He,
“Assessing the accuracy of diagnostic tests,”
Shanghai Arch. Psychiatry, 30
(3), 207
–212
(2018). https://doi.org/10.11919/j.issn.1002-0829.218052 Google Scholar
S. Feng et al.,
“Surface-enhanced Raman spectroscopy of saliva proteins for the noninvasive differentiation of benign and malignant breast tumors,”
Int. J. Nanomed., 10 537
–547
(2015). https://doi.org/10.2147/IJN.S71811 Google Scholar
N. S. Choutai et al.,
“Use of Raman spectroscopy to decrease time for identifying the species of Candida growth in cultures,”
Avicenna J. Med. Biotechnol., 7 45
–48
(2015). Google Scholar
C. De Luca, K. Dholakia and M. Mazilu,
“Modulated Raman spectroscopy for enhanced cancer diagnosis at the cellular level,”
Sensors, 15
(6), 13680
–13704
(2015). https://doi.org/10.3390/s150613680 SNSRES 0746-9462 Google Scholar
G. B. Jung et al.,
“Evaluation of antibiotic effects on Pseudomonas aeruginosa biofilm using Raman spectroscopy and multivariate analysis,”
Biomed. Opt. Express, 5 3238
–3251
(2014). https://doi.org/10.1364/BOE.5.003238 BOEICL 2156-7085 Google Scholar
P. Seow et al.,
“Quantification and visualization of lipid landscape in glioma using in and opposed-phase imaging,”
Neuroimage Clin., 20 531
–536
(2018). https://doi.org/10.1016/j.nicl.2018.08.003 Google Scholar
O. Uckermann et al.,
“Optical analysis of glioma: Fourier-transform infrared spectroscopy reveals the IDH1 mutation status,”
Clin. Cancer Res., 24
(11), 2530
–2538
(2018). https://doi.org/10.1158/1078-0432.CCR-17-1795 Google Scholar
W. G. Zijlstraa and A. Buursma,
“Spectrophotometry of hemoglobin: absorption spectra of bovine oxyhemoglobin, deoxyhemoglobin, carboxyhemoglobin, and methemoglobin,”
Comp. Biochem. Physiol. B: Biochem. Mol. Biol., 118
(4), 743
–749
(1997). https://doi.org/10.1016/S0305-0491(97)00230-7 CBPBB8 0305-0491 Google Scholar
J. E. Phelps et al.,
“Rapid ratiometric determination of hemoglobin concentration using UV-VIS diffuse reflectance at isosbestic wavelengths,”
Opt. Express, 18
(18), 18779
–18792
(2010). https://doi.org/10.1364/OE.18.018779 OPEXFF 1094-4087 Google Scholar
D. M. Gilkes, G. L. Semenza and D. Wirtz,
“Hypoxia and the extracellular matrix: drivers of tumour metastasis,”
Nat. Rev. Cancer, 14
(6), 430
–439
(2014). https://doi.org/10.1038/nrc3726 NRCAC4 1474-175X Google Scholar
M. B. Wallace et al.,
“Reflectance spectroscopy,”
Gastrointest. Endosc. Clin. N. Am., 19
(2), 233
–242
(2009). https://doi.org/10.1016/j.giec.2009.02.008 GECNED 1052-5157 Google Scholar
G. Zonios, J. Bykowski and N. Kollias,
“Skin melanin, hemoglobin, and light scattering properties can be quantitatively assessed in vivo using diffuse reflectance spectroscopy,”
J. Investig. Dermatol., 117
(6), 1452
–1457
(2001). https://doi.org/10.1046/j.0022-202x.2001.01577.x Google Scholar
S. C. Kanick et al.,
“Measurement of the reduced scattering coefficient of turbid media using single fiber reflectance spectroscopy: fiber diameter and phase function dependence,”
Biomed. Opt. Express, 2
(6), 1687
–1702
(2011). https://doi.org/10.1364/BOE.2.001687 BOEICL 2156-7085 Google Scholar
D. Boruah and P. Deb,
“Utility of nuclear morphometry in predicting grades of diffusely infiltrating gliomas,”
ISRN Oncol., 2013 760653
(2013). https://doi.org/10.1155/2013/760653 Google Scholar
M. Jermyn et al.,
“Intraoperative brain cancer detection with Raman spectroscopy in humans,”
Sci. Transl. Med., 7
(274), 274ra19
(2015). https://doi.org/10.1126/scitranslmed.aaa2384 STMCBQ 1946-6234 Google Scholar
M. Kirsch et al.,
“Raman spectroscopic imaging for in vivo detection of cerebral brain metastases,”
Anal. Bioanal. Chem., 398 1707
–1713
(2010). https://doi.org/10.1007/s00216-010-4116-7 ABCNBP 1618-2642 Google Scholar
P. Kumar,
“Raman spectroscopy as a promising noninvasive tool in brain cancer detection,”
J. Innov. Opt. Heal. Sci., 10
(5), 1730012
(2017). https://doi.org/10.1142/S1793545817300129 Google Scholar
J. Zhang et al.,
“Accuracy of Raman spectroscopy in differentiating brain tumor from normal brain tissue,”
Oncotarget, 8
(22), 36824
–36831
(2017). https://doi.org/10.18632/oncotarget.15975 Google Scholar
R. Galli et al.,
“Rapid label-free analysis of brain tumor biopsies by near infrared Raman and fluorescence spectroscopy—a study of 209 patients,”
Front. Oncol., 9 1165
(2019). https://doi.org/10.3389/fonc.2019.01165 FRTOA7 0071-9676 Google Scholar
S. C. Gebhart, W. C. Lin and A. Mahadevan-Jansen,
“In vitro determination of normal and neoplastic human brain tissue optical properties using inverse adding-doubling,”
Phys. Med. Biol., 51 2011
–2027
(2006). https://doi.org/10.1088/0031-9155/51/8/004 PHMBA7 0031-9155 Google Scholar
W. C. Lin et al.,
“Diffuse reflectance spectroscopy for in vivo pediatric brain tumor detection,”
J. Biomed. Opt., 15
(6), 061709
(2010). https://doi.org/10.1117/1.3505012 JBOPFO 1083-3668 Google Scholar
J. L. Lagarto et al,
“Real‐time fiber‐based fluorescence lifetime imaging with synchronous external illumination: a new path for clinical translation,”
J. Biophotonics, 13
(3), e201960119
(2019). https://doi.org/10.1002/jbio.201960119 Google Scholar
J. Yoon et al.,
“A clinically translatable hyperspectral endoscopy (HySE) system for imaging the gastrointestinal tract,”
Nat. Commun., 10 1902
(2019). https://doi.org/10.1038/s41467-019-09484-4 NCAOBW 2041-1723 Google Scholar
E. Baria et al.,
“In vivo multimodal fibre-probe spectroscopy for glioblastoma detection in mouse model,”
Proc. SPIE, 11073 110731N
(2019). https://doi.org/10.1117/12.2527124 PSISDG 0277-786X Google Scholar
E. Baria et al.,
“In vivo detection of glioblastoma through multimodal fibre-probe spectroscopy,”
Proc. SPIE, 11234 112341I
(2020). https://doi.org/10.1117/12.2546124 PSISDG 0277-786X Google Scholar
BiographyEnrico Baria is a postdoctoral researcher at the Department of Physics of the University of Florence. He received his bachelor’s degree in physics in 2011 and his master’s degree in physics and astrophysics in 2014 from the University of Florence. Finally, in 2018, he defended his PhD thesis at the European Laboratory for Non-linear Spectroscopy in Florence. His research activities focus on non-linear/Raman imaging and multimodal spectroscopy for tissue diagnostics. He is a member of SPIE. Enrico Pracucci is a PhD in molecular biophysics, currently working at Scuola Normale Superiore (Pisa, Italy) as postdoctoral fellow. He received his bachelor degree in molecular biology in 2011 and his master degree in Neurobiology in 2013, both from the University of Pisa. He received his PhD in molecular biophysics in 2019 from Scuola Normale Superiore, with a thesis titled “Unraveling alterations of excitation/inhibition balance in in vivo models of epilepsy and genetic autism”. Vinoshene Pillai is a PhD student in biophysical sciences at Scuola Normale Superiore, Pisa. She received her bachelor’s degree in chemistry in 2011 from the National University of Malaysia (UKM) and her master’s degree in organic chemistry in 2015 from the collaboration of two universities, National University of Malaysia (UKM), and National Science University of Malaysia (USM). Currently, she is finishing her doctoral thesis in intravital two-photon imaging of in-vivo glioblastoma models. Francesco S. Pavone is full professor at the Department of Physics, University of Florence, and the head of the biophotonics group at the European Laboratory for Nonlinear Spectroscopy (LENS). He obtained his degree in physics in 1989 at the University of Florence. In 1993, he obtained his PhD in optics at the National Institute of Optics (Florence). Currently, his research group is working on various topics in the field of biophotonics: single molecule biophysics, microscopy imaging-spectroscopy techniques, biomedical imaging, laser manipulation of bio-samples, neural and cardiac imaging. Gian M. Ratto is senior scientist at the Institute of Nanoscience (NANO) and professor at the Scuola Normale Superiore in Pisa. He graduated in physics at the University of Genoa and received his postdoctoral training in Berkeley and in Cambridge. He became tenured scientist at the Institute of Neuroscience in Pisa. In 2008, he moved to the physics laboratory of Scuola Normale in 2008, where he heads the in vivo imaging laboratory. The lab is interested in the cellular mechanisms at the basis of synaptic plasticity in physiological and pathological conditions. Riccardo Cicchi is a researcher at National Institute of Optics – National Research Council and he is associated to the European Laboratory for Non-linear Spectroscopy (LENS). He also serves as teacher in the course on Molecular Biotechnology at the University of Florence. His research includes multimodal microscopy imaging and bio-mechanical characterisation of tissues using multiphoton and inelastic scattering microscopies and fibre-probe spectroscopic methods for optical biopsy, clinical diagnostics and surgical guidance. He is a member of SPIE. |